Hanna Instruments HI 95707C Instruction Manual Download Page 4

4

17

The 

HI 95707 

is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years of

experience as a manufacturer of analytical instruments. It has the advanced optical system based on
a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable
readings. All instruments are factory calibrated and the electronic and optical design minimizes the
need of frequent calibration.
With the powerful CAL CHECK™ 

validation

 function you are able to validate good performance of

your instrument at any time. The validation procedure is extremely user friendly. Just use the exclusive
Hanna ready-made, NIST traceable standards to verify the performance of the instrument and
recalibrate if necessary.
All instruments are splash proof and the lamp and filter units are protected from dust or dirt by a
transparent cup. This makes the instruments fulfill field applications. Display codes aid the user in
routine operation. The meter has an auto-shut off feature that will turn off the instrument  after 10
minutes of non use or after 1 hour if left in the 

calibration mode

.

The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same position

every time it is placed into the measurement cell. It is designed to fit a cuvet with a larger neck
making it easier to add both sample and reagents. The cuvet is made from special optical glass to
obtain best results.
The 

HI 95707

 meter measures the Nitrogen-Nitrite 

Nitrogen-Nitrite 

Nitrogen-Nitrite 

Nitrogen-Nitrite 

Nitrogen-Nitrite (NO

2

-N) content in the 0.000 to 0.600 mg/L

(ppm) range, in drinking, surface and saline water samples and in domestic and industrial wastes. The
method is an adaptation of an EPA approved method.
The reagent is in

 

powder form and is supplied in packets. The amount of reagent is precisely dosed to

ensure the maximum repeatability.

degree Celsius
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter
Environmental Protection Agency (USA)

°C

:

°F

:

mg/L

:

mL

:

EPA:

GENERAL DESCRIPTION

ABBREVIATIONS

or

• Press READ/TIMER and “SIP” will blink during

measurement.

• The instrument will show for three seconds the

CAL CHECK™ standard value.
Then the date of last calibration (e.g.: “01.08”)
appears on the display, or “01.01” if the factory
calibration was selected. In both cases the number
of the month is blinking, ready for date input.

Note

: if the display shows “ERR” the calibration

procedure failed. Verify that the CAL CHECK™
Standards HI 95707-11 Cuvets, 

A

 and

 B

, are

free from fingerprints or dirt and that they are
inserted correctly.

DATE INPUT

• Keep READ/TIMER pressed to scroll to the desired

month number (01-12).

• When the correct month has been set, press ZERO

to confirm. Now the display will show the day
number blinking. Keep READ/TIMER pressed to
scroll to the desired day number (01-31).

Note

: it is possible to change from 

day 

to 

month

and vice versa by pressing ZERO.

• When both the day and month have been selected,

hold CAL CHECK for three seconds to store date
and calibration values. The instrument will show
for one second “Stor”, to confirm that the new

calibration data has been accepted.

• The instrument will return automatically to the

measurement mode by  displaying "707" on the
LCD.

Summary of Contents for HI 95707C

Page 1: ...1 20 MAN95707CR3 06 06 w w w h a n n a i n s t c o m w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95707C Nitrite Low Range ISM...

Page 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Page 3: ...ONRESET Itispossibletorestorefactorycalibration Enter the calibration mode by holding CAL CHECK for three seconds Hold ZERO for 10 seconds The display will show for 2 seconds F CAL and 707 appears The...

Page 4: ...Nitrogen Nitrite Nitrogen Nitrite NO2 N content in the 0 000 to 0 600 mg L ppm range in drinking surface and saline water samples and in domestic and industrial wastes The method is an adaptation of...

Page 5: ...sentative lot of reagent an operator obtained with a single instrument a standard deviation of 0 001 mg L SPECIFICATIONS PRECISION AND ACCURACY CALIBRATION PROCEDURE Turn the meter on by pressing ON O...

Page 6: ...dation corrected value read value 1 0 002 20 T C corrected value read value 1 0 001 68 T F Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter When...

Page 7: ...d reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is ne...

Page 8: ...e 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low Turn the meter on by pressing...

Page 9: ...ion is selected GUIDE TO DISPLAY CODES It is important that the sample does not contain any debris This would corrupt the reading In order to avoid reagent leaking and to obtain more accurate measurem...

Page 10: ...he edges of the packet to form a spout c pour out the content of the packet The flashing digits indicate that 1 the month can be set 2 the day can be set Storage this prompt appears for 1 second at th...

Reviews: