background image

9

8

3

4

5

6

7

8

9

10

11

2

1

°C

CH

HI 955

302

PT100 - LOGGING THERMOMETER
DUAL CHANNEL

PRINT

ALT

TEMP

PAPER

TIME

LOG

UP

DOWN

1.

Connector(s) for Pt100 probe

2.

Power adapter plug

3.

Liquid Crystal Display

4.

PAPER key, to feed paper

5.

PRINT key, to print on-demand (current time
and temperatures)

6.

ALT key, to activate alternate function key

7.

TEMP key, to read temperature and reactivate
the meter when is in stand-by

8.

UP key, to scan data or to set up

9.

DOWN key, to scan data or to set up

10. TIME key, to display current time and printing

interval

11. LOG key, to enter/exit the logging mode

SPECIFICATIONS

HI 955301 AND HI 955302

FUNCTIONAL DESCRIPTION

HI 955301 AND HI 955302

HI 955301

HI 955302

Range

-200.0 to 850.0°C

Resolution

0.1°C or 1°C, selectable

Accuracy

±0.1°C ±0.1% of reading

(for 1 year, excluding probe error)

Typical EMC Deviation

±0.5°C

Channels

1

2

Probe

HI 768 series (not included)

Printer

low-power impact,

14 characters per line using

38 mm plain paper (HI 710034)

Printing Interval

selectable at

1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes

PC Connection

HI  9200 interface through

serial port and using HI 92000 software

Power Supply

4 x 1.5V AA batteries;

approx. 500 hours of use

(with 60' printing interval);

auto-off after  minutes of non-use;

or 12 Vdc input

Environment

0 to 50°C (32 to 122°F);

RH max 95%

Dimensions

220 x 82 x 66 mm

(8.7 x 3.2 x 2.6")

Weight

550 g (1.2 lb.)

Summary of Contents for HI 955201

Page 1: ...1 Instruction Manual HI 955201 HI 955202 HI 955301 HI 955302 1 or 2 Channel Printing and Logging Pt 100 Thermometers w w w h a n n a i n s t c o m...

Page 2: ...Declaration of Conformity 35 TABLE OF CONTENTS WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and m...

Page 3: ...apabilities and an infrared data transfer system In addition to supplying the user with instant printouts of date time and temperature they also provide data log ging facilities that store data for tr...

Page 4: ...de 3 4 5 6 7 8 9 10 11 2 1 C CH HI 955202 PT100 PRINTING THERMOMETER DUAL CHANNEL ON OFF INTV TEMP PAPER PRINT RECORD UP DOWN HI 955201 HI 955202 Range 200 0 to 850 0 C Resolution 0 1 C or 1 C selecta...

Page 5: ...e SPECIFICATIONS HI 955301 AND HI 955302 FUNCTIONAL DESCRIPTION HI 955301 AND HI 955302 HI 955301 HI 955302 Range 200 0 to 850 0 C Resolution 0 1 C or 1 C selectable Accuracy 0 1 C 0 1 of reading for...

Page 6: ...955201 and HI 955301 are provided with a single probe connection while HI 955202 and HI 955302 feature two probe connectors Looking at the display side channel 1 is the first connector on the top lef...

Page 7: ...d TIME keys to leave this mode TIME DATE UP TIME ALT TEMP TIME INTV TIME SETTING DATE TIME PRINTING INTERVAL for HI 955301 and HI 955302 only Turn the instrument on by pressing the TEMP key Press the...

Page 8: ...luggedin the meter will print only the value of the connected probe If the second probe is not connected during the re cording mode the data from the second probe will not be printed PRINT LOG INTV ON...

Page 9: ...will appear on the display next to the channel number not utilized and orrecorded When the meter is in record ing mode the LOG indica tion is displayed on the sec ondary display To save battery life...

Page 10: ...log d Printing interval indicator in minutes e Time HH MM f Temperature value s g Channel HI 955302 only UP DOWN TIME LOG ALT TEMP LOG LOG UP DOWN TIME LOG ALT TEMP LOG PRINT To continue logging with...

Page 11: ...ime and the LOG symbol will start blinking to indicate that data are now only stored into memory Notes It is recommended to use the external power adapter during logging with printing mode especially...

Page 12: ...LT TEMP INTV TIME INTV TIME LOG UP PAPER ALT PRINT Press the DOWN key to revert back to sam pling time and move back through samples To exit the recall mode press LOG Note This mode will not alter dat...

Page 13: ...ations on the secondary display mean that the meter has entered the log mode and has 15 pages of available memory remaining After the initial printout the time and printing logging in terval will reap...

Page 14: ...nd the transmitter Data transmission from the instrument to the PC is handled with the HI 92000 Windows compatible software by HANNA instruments HI 92000 offers a variety of features to help you in al...

Page 15: ...instrument through the 12 Vdc adapter and then proceed with battery replacement Only after new batteries have been installed the adapter can be un plugged without losing stored data HI 92000 allows yo...

Page 16: ...r caught under the cover and prevented printer fromadvancingpaperfeed Replacethebatterycoverandtightenthescrews 1 5V 1 5V 1 5V 1 5V SCREW POTS RESET PAPER PAPER PRINTER MAINTENANCE CHANGING THE INK CA...

Page 17: ...to have all ther mometers recalibrated at least once a year For an accurate annual recalibration contact the nearest HANNA Service Center 1 5V 1 5V 1 5V 1 5V SCREW POTS RESET CH C LOBAT The LO BAT in...

Page 18: ...instruments when voltages at the measurement surface exceed 24 Vac or 60 Vdc To avoid damages or burns do not perform any measure ment in microwave ovens 4 WIRE PT100 PROBES HI768A Air gas probe 1 m 3...

Page 19: ...03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2...