Hanna Instruments HI 93750 Instruction Manual Download Page 2

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

OPERATIONAL GUIDE

MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “- - -”, it is

ready.

• Fill the cuvet with 10 mL of unre-

acted sample up to the mark.

• Add six drops of HI 93750A-0, replace the cap and

swirl the solution.

• Place the cuvet into the holder

and ensure that the notch on
the cap is positioned securely
into the groove.

• Press ZERO and “SIP” will appear on the display.

• Wait for a few seconds and the

display will show “-0.0-”. Now the
meter is zeroed and ready for measurement.

• Remove the cuvet.

ON

OFF

ZERO

• Add one packet of HI 93750B-0 reagent. Replace the

cap and mix very gently for one minute by turning the
cuvet upside down.

• Reinsert the cuvet into the instru-

ment.

• Press READ TIMED and the display will show the count-

down prior to the measurement or, alternatively, wait
for 2 minutes and press READ DIRECT. In both cases
“SIP” will appear during measurement.

• The instrument directly displays concentration in mg/L

(ppm) of potassium (K).

INTERFERENCES

The following ions do not interfere below the concentrations
reported:
Ammonium

    10 ppm

Calcium

10000 ppm as CaCO

3

Chloride

12000 ppm

Magnesium

 8000 ppm as CaCO

3

Sodium

 8000 ppm

Color or suspended matter in large amounts will cause
interference. Therefore the turbidity should be removed by
prior filtration.
For best results samples should be tested at room tempera-
ture (20°C or 68°F). Lower temperatures will result in
higher readings.

The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the

zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.

• Do not let the test sample stand too long after reagent

is added or accuracy will be lost.

• Whenever the cuvet is placed into the measurement cell,

it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with 

HI 731318

 or a lint-free cloth

prior to insertion.

• It is important that the sample does not contain any

debris. This would corrupt the readings.

• It is possible to take multiple readings in a row, but it

is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.

• It is important to discard the sample immediately after

the reading is taken because the glass might become
permanently stained.

• Shaking the cuvet can generate bubbles in the sample,

causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.

TIPS FOR AN ACCURATE

TIPS FOR AN ACCURATE

TIPS FOR AN ACCURATE

TIPS FOR AN ACCURATE

TIPS FOR AN ACCURATE

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

BATTERY REPLACEMENT

Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.

9V

SLIDE

OFF

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environ-
ment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

REAGENT SETS

HI 93750-01

Reagents for 100 tests

OTHER ACCESSORIES

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731321

Glass cuvets (4 pcs)

HI 731325

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL).

CE DECLARATION

CE DECLARATION

CE DECLARATION

CE DECLARATION

CE DECLARATION

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

OF CONFORMITY

IST93750

09/99

PRINTED IN PORTUGAL

READ

TIMED

Reviews: