Hanna Instruments HI 93730 Instruction Manual Download Page 2

ON

OFF

MEASUREMENT PROCEDURE

• Turn the meter on by

pressing ON/OFF.

• When the LCD displays

"- - -", it is ready.

• Fill one cuvet with 10 mL

of sample up to 1.5 cm
(¾") below the rim and
replace the cap.

• Place the cuvet into the

holder and ensure that the
notch on the cap is
positioned securely into
the groove.

• Press ZERO and "SIP" will appear on the display.

• Wait for a few seconds and

the display will show "-0.0-".
Now the meter is zeroed and
ready for measurement.

• Fill one graduated mixing

cylinder up to the 25 mL mark
with the sample.

• Add the content of one packet of HI 93730A-0 Reagent

A, close the cylinder and invert several times until the
powder is completely dissolved.

ZERO

READ

TIMED

12

3

6

9

5’

READ

DIRECT

The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the

zeroing and the measuring phases, it is necessary to close
the cuvet to prevent any contamination.

• Do not let the test sample stand too long after reagent is

added or accuracy will be lost.

• Whenever the cuvet is placed into the measurement cell,

it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with 

HI 731318

 or a lint-free cloth

prior to insertion.

• It is important that the sample does not contain any

debris. This would corrupt the readings.

• It is possible to take multiple readings in a row, but it is

recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.

• It is important to discard the sample immediately after

the reading is taken because the glass might become
permanently stained.

• Shaking the cuvet can generate bubbles in the sample,

causing higher readings. To obtain accurate measurements,
remove such bubbles by swirling or by gently tapping the
vial.

• All the reaction times reported in this manual are referred

to 20°C (68°F). As a general rule of thumb, they should
be doubled at 10°C (50°F) and halved at 30°C (86°F).

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment
in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences
to radio and TV equipments, requiring the operator to take all necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.

REAGENT SETS

HI 93730-01

Reagents for 100 tests

HI 93730-03

Reagents for 300 tests

OTHER ACCESSORIES

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731321

Glass cuvets (4 pcs)

HI 731325

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL).

Operational Guide

Tips for an Accurate Measurement

CE Declaration of Conformity

Accessories

Read the relevant SDS sheets before performing the test.

Safety Data Sheets are available at: www.hannainst.com

Safety Data Sheets

10 mL

25 ml

• Add the content of one packet of HI 93730B-0 Reagent

B to the cylinder, close it and invert several times until the
powder is completely dissolved.

• Add the content of one packet of HI 93730C-0 Reagent

C to the cylinder, close and shake it vigorously.

• Fill an empty cuvet with 10 mL of the reacted sample up

to 1.5 cm (¾") below the rim, replace the cap and insert
the cuvet into the instrument.

• Press READ TIMED and the display will show the countdown

prior to the measurement, or, alternatively wait for 5
minutes and press READ DIRECT. In both cases "SIP" will
appear during measurement.

• The instrument directly displays concentration in mg/L of

molybdenum on the Liquid Crystal Display.

INTERFERENCES

Interference may be caused by:
Aluminum above 50 mg/L

Chromium above 1000 mg/L

Copper above 10 mg/L

Iron above 50 mg/L

Nickel above 50 mg/L

Nitrite, as NO

2

–––––

Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the
buffering capacity of the reagents.

Reviews: