
15
The
HI 93742
meter measures the iron and manganese content in
water, wastewater and seawater (iron only) in the following ranges:
Iron
0 to 400 µg/L (ppb)
Manganese 0 to 300 µg/L (ppb).
SPECIFICATIONS
Range
Fe
0 to 400 µg/L
Mn
0 to 300 µg/L
Resolution
1 µg/L
Accuracy
Fe
±10 µg/L ±8% of reading
Mn
±2 µg/L ±3% of reading
Typical EMC
±1 µg/L
Deviation
Light Source
Light Emitting Diode @ 555 nm
Method
Adaptation of the TPTZ method for iron and
PAN method for manganese. The reaction be-
tween iron or manganese and the reagents
respectively causes a blue or an orange tint in
the sample.
REQUIRED REAGENT
Code
Description
Quantity
Iron
HI 93746-0
TPTZ reagent
2 packets
HI 93742
Iron LR/Manganese LR
HI 93742 -
HI 93742 -
HI 93742 -
HI 93742 -
HI 93742 - Iron Low Range and
Iron Low Range and
Iron Low Range and
Iron Low Range and
Iron Low Range and
Manganese Low Range
Manganese Low Range
Manganese Low Range
Manganese Low Range
Manganese Low Range