15
14
Press
READ
. The meter will intermittently display “
SIP
” on the upper
level of the LCD.
After a few seconds the display will show the turbidity value, e.g.
5.34 NTU:
COLORIMETRIC MEASUREMENTS
Turn the meter on. After the initialization routine, the LCD will
display:
Use the and keys to choose the
desired parameter:
F CL
Free Chlorine
t CL
Total Chlorine
CY
Cyanuric Acid
PH
pH
Id
Iodine
Br
Bromine
L FE
Iron
Measuring Chlorine, Cyanuric Acid, pH, Iodine, Bromine
Fill the vial with the sample (blank). The surface of the
vial should be clean and scratch free.
Insert the blank sample into the cuvette holder and ensure that the
notch on the cap is positioned securely into the groove. Press
ZERO
.
+
+
CURRENT TIME/DATE RECALL
To recall current TIME/DATE press and hold the
ALT
and
READ
keys
together. The current time and an intermittent “
TIME
” will be
displayed.
Release the keys.
Press and hold the
ALT
and
READ
keys again and the meter will
show the current date together with an intermittent “
DATE
”.
TURBIDITY MEASUREMENTS
Fill the vial with the sample. The surface of the vial
should be clean and scratch free.
Turn the meter on. After the initialization routine, the
LCD will show:
Use the and keys to set the lower level of the LCD to turbidity
(tr).
Insert the sample into the cuvette holder and
ensure that the notch on the cap is positioned
securely into the groove.