LOGGING MODE WITH PRINTING
This function is suggested when an immediate report of the measure-
ment is required in addition to the recording of the data into
memory.
Press the ALT and the LOG key simulta-
neously to enter the logging mode. The log
number and page number will appear for a
few seconds on the display to indicate the
correct operational mode. The printer will
print a complete set of data and the "LOG"
symbol will appear on the bottom left corner
of the display.
If no key is pressed, the display goes blank after about 5 minutes.
During printing, the display shows the time, interval and the “LOG”
symbol.
To reactivate the display press the TEMP key.
The printout provides the following information:
a - Date (DD-MM-YY)
b - A running log number
c - A running sample number (in that
particular log)
d - Printing interval indicator in minutes
e - Time (HH-MM)
f - Temperature value/s
It is always possible to switch from the logging with printing function
to the logging only function. Press the ALT and the PAPER keys at the
same time and the "LOG" symbol will start to blink to indicate that
the data are now stored only into memory.
LOG
INTV
TIME
TEMP
10-01-96
==07== 0005M
0014 *14.25
#1 25.0 °C
#2 27.3 °C
.
.
b
c
d
a
e
f
CH
°C
LOG
UP
PAPER
ALT
Untitled-27
18
18
Notes:
•
It is recommended to use the adapter during logging with
printing mode, especially when many printout are going to be
taken.
•
Before proceeding with logging with printing, make sure there
is enough paper for your measurements. When the paper is
finished the meter will not advise the operator and the
printouts could be lost. If this happens, data will continue to
be stored into memory, and it is always possible to print them
at different time (see above).
•
It is possible to insert a new paper roll during logging session
(see page 26).
•
Once in the logging mode, the inter-
val cannot be changed. Exit the logging
mode first (pressing the ALT and the
LOG keys) before setting the new in-
terval.
•
If the PRINT key is pressed while in
logging mode, a printout is produced
without affecting the running sample
number.
TO STOP LOGGING
Press the ALT and the LOG keys simulta-
neously, this will generate a log exit status
print out.
UP
DOWN
TIME
LOG
ALT
TEMP
UP
DOWN
TIME
LOG
ALT
TEMP
10-01-96
==07== 0005M
0016 *16.25
Log number
Total sample number
LOG EXIT
Date/Time