background image

7

The LCD will show the pH measurement and the temperature of the

sample.

In order to take more accurate pH measurements, make sure that the

instrument is calibrated (see page 9 for details).
The glass bulb and the junction on the pH electrode should always be

moist, never allow it to dry out.
If several measurements are taken successively in different samples,

rinse the electrode thoroughly with deionized or tap water and a

small amount of the sample to be measured.

The pH reading is directly affected by temperature. In order for the meter to

measure the pH accurately, temperature must be compensated for. If the

sample temperature is different from the temperature at which the pH

electrode was kept, allow a few minutes for thermal equilibrium.

To use the meter's Automatic Temperature Compensation feature,

submerge the temperature probe into the sample as close to the

electrode as possible and wait for a few minutes.

If manual temperature compensation is desired the temperature
probe must be disconnected from the instrument

.

The display will show the default temperature of 25 °C, or the last

temperature set with the "°C" (or "°F") indicator blinking.

The temperature can now be adjusted with the UP and DOWN arrow keys.

ORP MEASUREMENTS (HI 9125 only)

To perform ORP measurements, connect an optional ORP electrode

(see "Accessories" section) to the meter and turn it ON.

If necessary, enter the "mV" mode by pressing RANGE until the

display changes to mV.

22

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

ACCESSORIES

pH CALIBRATION SOLUTIONS

HI 70004P

pH 4.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 70007P

pH 7.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 70010P

pH 10.01 Buffer Solution, 20 mL sachet, 25 pcs

HI 7004L

pH 4.01 Buffer Solution, 500 mL bottle

HI 7004M

pH 4.01 Buffer Solution, 230 mL bottle

HI 7006L

pH 6.86 Buffer Solution, 500 mL bottle

HI 7006M

pH 6.86 Buffer Solution, 230 mL bottle

HI 7007M

pH 7.01 Buffer Solution, 500 mL bottle

HI 7007M

pH 7.01 Buffer Solution, 230 mL bottle

HI 7009L

pH 9.18 Buffer Solution, 500 mL bottle

HI 7009M

pH 9.18 Buffer Solution, 230 mL bottle

HI 7010L

pH 10.01 Buffer Solution, 500 mL bottle

HI 7010M

pH 10.01 Buffer Solution, 230 mL bottle

ELECTRODE STORAGE SOLUTION

HI 70300L

Storage Solution, 460 mL bottle

HI 70300M

Storage Solution, 230 mL bottle

ELECTRODE CLEANING SOLUTIONS

HI 70000P

Electrode Rinse Solution, 20 mL sachet, 25 pcs

HI 7061L

General Cleaning Solution, 460 mL bottle

HI 7061M

General Cleaning Solution, 230 mL bottle

HI 7073L

Protein Cleaning Solution, 460 mL bottle

HI 7073M

Protein Cleaning Solution, 230 mL bottle

HI 7074L

Inorganic Cleaning Solution, 460 mL bottle

HI 7074M

Inorganic Cleaning Solution, 230 mL bottle

HI 7077L

Oil & Fat Cleaning Solution, 460 mL bottle

HI 7077M

Oil & Fat Cleaning Solution, 230 mL bottle

REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs)

HI 7071

3.5M KCl+AgCl Electrolyte for single junction electrodes

HI 7072

1M KNO

3

 Electrolyte

HI 7082

3.5M KCl Electrolyte for double junction electrodes

HI 8093

1M KCl+AgCl Electrolyte

ORP PRETREATMENT SOLUTIONS

HI 7091L

Reducing Pretreatment Solution, 460 mL bottle

HI 7091M

Reducing Pretreatment Solution, 230 mL bottle

HI 7092L

Oxidizing Pretreatment Solution, 460 mL bottle

HI 7092M

Oxidizing Pretreatment Solution, 230 mL bottle

ORP SOLUTIONS

HI 7020L

Test Solution 200-275 mV, 500 mL bottle

HI 7020M

Test Solution 200-275 mV, 230 mL bottle

HI 7021L

Test Solution 240 mV, 500 mL bottle

HI 7021M

Test Solution 240 mV, 230 mL bottle

HI 7022L

Test Solution 470 mV, 500 mL bottle

HI 7022M

Test Solution 470 mV, 230 mL bottle

Summary of Contents for HI 9125

Page 1: ...4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfri...

Page 2: ...n shipping any instrument make sure it is properly packed for complete protection TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS WARRANTY 2 PRELIMINARY EXAMI...

Page 3: ...ith 1 m 3 3 cable pH 4 01 pH 7 01 buffer solutions 20 mL sachet 100 mL plastic beaker 4 x 1 2 AAA rechargeable batteries inside the instrument HI 710044 inductive recharger with power adapter Instruct...

Page 4: ...CRIPTION FUNCTIONAL DESCRIPTION 25 120 mm 4 7 12 mm 0 5 FC 100 75 mm 2 95 6 mm 0 25 FC 200 120 mm 4 7 12 mm 0 5 FC 210 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 FC 220 110 mm 4 3 12 mm 0 5 FC 911 FC 100B Pl...

Page 5: ...e c n a d e p m I t u p n I 0 1 2 1 s m h o e f i L e p y T y r e t t a B e z i s A A A V 2 1 x 4 s e i r e t t a b e l b a e g r a h c e r e s u s u o u n i t n o c f o s r u o h 0 0 2 x o r p p a f...

Page 6: ...f after 20 minutes with no button pressed To disable this feature see Setup Menu section on page 13 pH MEASUREMENTS To take a pH measurement remove the electrode protective cap and submerge the electr...

Page 7: ...er Solution 20 mL sachet 25 pcs HI 7004L pH 4 01 Buffer Solution 500 mL bottle HI 7004M pH 4 01 Buffer Solution 230 mL bottle HI 7006L pH 6 86 Buffer Solution 500 mL bottle HI 7006M pH 6 86 Buffer Sol...

Page 8: ...e h C r o t c e n n o c s w o h s y a l p s i D e l a c s l l u f g n i k n i l b e u l a v f o t u o g n i d a e R e g n a r s i e d o r t c e l e e h t e r u s e k a M d e t c e n n o c s i e l p m...

Page 9: ...with distilled water immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution...

Page 10: ...o the solution place the temperature probe as close as possible to the electrode and stir gently Select the second buffer value on the secondary display by pressing the UP and DOWN arrow keys 19 E E E...

Page 11: ...irst point see above press the CAL key to end the calibration procedure 18 Typical Electrode Life Ambient Temperature 1 3 years 90 C Less than 4 months 120 C Less than 1 month Alkaline Error High conc...

Page 12: ...8 9 6 9 9 8 9 7 9 0 6 0 4 1 9 0 4 4 8 6 8 9 6 7 9 8 7 7 9 5 6 9 4 1 1 1 4 4 8 6 9 9 6 5 9 8 6 7 9 0 7 8 5 1 2 1 4 5 8 6 9 9 6 3 9 8 5 7 9 5 7 7 6 1 4 1 4 6 8 6 0 0 7 1 9 8 4 7 9 0 8 6 7 1 6 1 4 7 8 6...

Page 13: ...NE ARE USED IN ALK POSITION PUT THE SWITCH WARNING Ni MH ALK Ni MH ALK Replace the rechargeable batteries only if necessary To replace the rechargeable batteries follow the next steps Turn OFF the ins...

Page 14: ...tion m m m m mV CALIBRATION HI 9125 V CALIBRATION HI 9125 V CALIBRATION HI 9125 V CALIBRATION HI 9125 V CALIBRATION HI 9125 only only only only only TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEM...

Reviews: