Hanna Instruments HI 87314 Instruction Manual Download Page 2

IST87314Y  10/10

INITIAL PREPARATION

The meter is supplied complete with a 9V battery. Remove
the battery compartment cover on the back of the meter, install
the battery while observing its polarity.
Connect the desired probe (

HI 76302W

 for EC or 

HI 3316D

for resistivity measurements) to the DIN socket on the top of
the meter. The instrument will automatically recognize which
probe is connected.
Turn the meter ON by pressing the 

ON/OFF

 key.

TAKING CONDUCTIVITY MEASUREMENTS

• Immerse the 

HI 76302W

 probe in the solution, while

making sure that the holes of the sleeve are completely
submerged. Tap the probe lightly on the bottom of the
beaker to remove any air bubbles trapped inside.

• Adjust the conductivity TC knob to the temperature coefficient

value of the solution.

• Select the appropriate conductivity range, by pressing 

EC

RANGE

 key.

Note:

If the display shows “

1

”, the reading is over-range.

Select the next higher range.

• Allow a few minutes for the reading to stabilize, and the

LCD will display the temperature compensated conductivity
reading.

TAKING RESISTIVITY MEASUREMENTS

• Immerse the 

HI 3316D

 probe in the solution, while making

sure that the hole on the probe body is completely
submerged.

• Wait for a few seconds to allow the reading to stabilize.

The resistivity value of the solution will be displayed on the
LCD.

Note:

If the display shows “

1

”, the reading is over-range.

• The reading is automatically corrected for temperature

variations by using the linear compensation method:
R

25

=R

t

(1+

β

(t-25))

where R

t

 is the resistivity at the temperature t, and the

reference temperature is 25ºC.

• The temperature coefficient 

β

 is selectable through a front

knob, from 2 to 7%/°C.

HI 76302W

4-ring conductivity probe with built-in temperature
sensor, DIN connector and 1 m (3.3’) cable

HI 3316D

Resistivity probe with built-in temperature sensor,
DIN connector and 1 m (3.3’) cable

HI 7030M

12880 

µ

S/cm, 230 mL bottle

HI 7031M

1413 

µ

S/cm, 230 mL bottle

HI 7033M

84 

µ

S/cm, 230 mL bottle

HI 7034M

80000 

µ

S/cm, 230 mL bottle

HI 7035M

111800 

µ

S/cm, 230 mL bottle

HI 7039M

5000 

µ

S/cm, 230 mL bottle

HI 7061M

General cleaning solution, 230 mL bottle

HI 710007

Shockproof rubber boot, blue

HI 710008

Shockproof rubber boot, orange

HI 731326

Calibration screwdriver (20 pcs)

HI 76405

Electrode holder

CONDUCTIVITY CALIBRATION

• It is recommended to calibrate the instrument for the

conductivity range at least once a week, or when the probe
is changed.

• For better accuracy, choose a solution with a conductivity

value close to the sample to be measured.

If you are measuring in the mS ranges, calibrate the
meter using the 

HI 7030

 (12.88 mS/cm) or 

HI 7034

(80 mS/cm) conductivity solution.

For the 

µ

S ranges, use 

HI 7031

 (1413 

µ

S/cm) for the

0 to 1999 

µ

S/cm range, or 

HI 7033

 (84 

µ

S/cm) for the

0 to 199.9 

µ

S/cm range.

• Rinse the probe thoroughly with distilled water and, if

possible, use plastic beakers to minimize any EMC
interferences.

• Pour a small quantity of the desired conductivity solution

into a plastic beaker, and immerse the probe while
making sure that the holes of the sleeve are completely
submerged. Tap the probe lightly on the bottom of the
beaker to remove any air bubbles trapped inside the sleeve.

• Adjust the conductivity TC knob to 2%/°C.

• Select the appropriate measurement range: 199.9 

µ

S if

calibrating with 

HI 7033

 solution, 1999 

µ

S for 

HI 7031

,

19.99 mS for 

HI 7030

, or 199.9 mS for 

HI 7034

.

Note:

If the display shows “

1

”, the reading is over-range.

Select the next higher range.

• Allow a few minutes for the reading to stabilize and adjust

the conductivity calibration trimmer located in the battery
compartment, to read the calibration solution value @25°C
(77°F).

• All subsequent measurements will be referenced to 25°C.

• If you need to reference the measurements to 20°C, adjust

the calibration knob to read the calibration solution value
@20°C (68°F). See conductivity vs. temperature chart on
the calibration solution sachet or bottle label.

• Calibration is now complete and the instrument is ready

for use.

RESISTIVITY CALIBRATION

• The meter is factory calibrated for resistivity range, and

usually needs recalibration only after a resistivity probe
replacement. In this case, follow the instructions below.

• Immerse the probe in a solution of known resistivity value,

and wait for the reading to stabilize.

• Adjust the resistivity calibration trimmer located in the

battery compartment, until the resistivity value of the
reference solution is displayed.

OPERATIONAL GUIDE

CALIBRATION

BATTERY REPLACEMENT

ACCESSORIES

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment in
which they are used. Operation of these instruments in residential area could cause unaccept-
able interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences. Any variation introduced by the user to the supplied equipment
may degrade the instruments' EMC performance. To avoid damages or burns, do not
perform any measurement in microwave ovens.

When the battery becomes weak,
the meter displays a blinking
additional decimal point on the
left side of the LCD.
When the low battery indicator appears only a few hours of
battery life is remaining. A low battery level may also result in
unreliable measurements. It is recommended to replace the
battery immediately.
Unscrew the 3 screws on the back of the meter, remove the
battery cover and replace the battery while observing its polarity.
Replacement must only take place in a safe area using an
alkaline 9V battery.

Reviews: