background image

HI8711 allows the selection of two set points with two independent outputs for acid and alkaline 
dosages. 

Auto-diagnostics to verify offset and slope calibration and electrode contamination/deterioration

Fail Safe Alarm System protects against problems caused by power supply failure or signal 
interruption 

Selectable manual ON, OFF and automatic dosing control for alkaline and acid dosing with 
indicator. 

Fail Safe Alarm System

Hanna’s exclusive Fail Safe Alarm System protects against problems caused by power supply 
failure or signal interruption, which are typical risks in industrial environments. An alarm 
condition is indicated by a red LED, located directly on the front panel of the controller. 

Self-Diagnostic Test

Built-in autodiagnostic functions enable the user to check and troubleshoot any malfunctions. 
The functions are made via front panel keys to isolate the cause of malfunction whether it is due
to pH electrode contamination, internal offset circuit or the amplifier circuit. 

Overtime Dosing Protection

The HI8711 incorporates adjustable overtime dosing protection from 5 to 60 minutes. If dosing 
exceeds selected time, the alarm will be triggered and the dosing contact will deactivate. This 
feature can be activated or deactivated. 

Matching Pin Connection

Supports potential matching pin connection to avoid typical problems caused by grounding loop
current, such as progressive damage of the electrode, fluctuating measurements, and poor 
process regulation. 

Output and Input Connections

Each model accepts either a direct input from a pH or ORP electrode or from a transmitter 
through 4-20 mA input. The instrument also provides ±5V power output and input terminals for 
amplified electrodes. Moreover, you can choose the output configuration for connecting a 
recorder or a PLC, between 0-20 or 4-20 mA. 

Robust Design

A removable, transparent splash-proof cover protects the front panel. 

pH

Summary of Contents for HI 8510

Page 1: ... panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination internal offset circuit or the amplifier circuit Overtime Dosing Protection The HI8711 incorporates adjustable overtime dosing protection from 5 to 60 minutes If dosing exceeds selected time the alarm will be triggered and the dosing contact will deactivate This feature can be activated or deactivated M...

Page 2: ...ontrol OFF AUTO ON with selection switch Over Dosing Control adjustable from 5 min to 60 min with knob or disable by wire strap on rear panel Backlight continuous on Voltage Output 5 Vcc 150 mA max load for amplified electrodes Enclosure flame retardant ABS body and front panel transparent splash proof front cover Panel Cutout 141 x 69 mm 5 6 x 2 7 Recorder Output 0 20 mA or 4 20 mA isolated Set P...

Page 3: ...720 HI 931500 HI931501 HI 932500 Panel Mounted ORP pH Indicators and Controllers SLOPE DOSAGE OXID REDUC AL COARSE SET FINE AL MEA SURE SET 0 mV TEST 500 mV TEST HI 8720E ORP mV http www hannainst com These Instruments are in Compliance with the CE Directives ...

Page 4: ...unctional Description HI 8710 12 Functional Description HI 8711 16 Functional Description HI 8720 20 Functional Description HI 931500 24 Functional Description HI 931501 26 Functional Description HI 932500 29 Specifications HI 8510 and HI 8512 31 Specifications HI 8710 and HI 8711 32 Specifications HI 8720 and HI 931500 33 Specifications HI 931501 and HI 932500 34 Initial Preparation 35 Operationa...

Page 5: ...icity of use in a wide range of industrial process applications The models are designed with DIN standard panel mount with membrane keypads on the front panel a large LCD display and autodiagnostic functions not HI 931500 HI 931501 and HI 932500 Connections to the power supply contacts and recorders are made through screw termi nals on the rear panel Two models are available for HI 8510 HI 8711 HI...

Page 6: ...ng the ORP indicators in conjunction with a 4 20 mA output ORP transmitter HI 8615 or HI 8615L with LCD display will assure you of a strong interference free signal at distances up to 300 m 1000 ft All instruments are supplied with a plastic transparent front cover and mounting brack ets electrode and mains cable excluded ...

Page 7: ...se dimensions show the cutout size for the installation Side view of the panel mounted unit Adjustable location brackets supplied with the meter allow the indicator to slide into the cutout and will hold the unit securely in place 190 mm 7 50 is the minimum amount of room required to install the indicator with the cables connected 190mm MIN 7 50 135mm 5 31 0 25 4mm 0 01 0 160 144mm 5 67 ADJUSTABLE...

Page 8: ... 8510E pH Keypad SENSORTEST To display the mV response of the electrode in order to verify its working condition pH 7 TEST Toverifytheinternalcircuitofthemeterinterms ofOffsetcompensation pH 4 TEST To verify the amplifier circuit of the meter Trimmers O ForOffsetcalibration SLOPE For Slope calibration ...

Page 9: ...he power supply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Recorderoutput 4 ConnectionsforPT100temperaturesensor 5 BNC socket for pH electrode FUSE FUSE ELECTRODE INPUT PT100 mA OUT 220 110 L 1 2 3 4 5 REARPANELHI8510E ...

Page 10: ...8 Unplug the instrument from the power supply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Recorderoutput 4 Connections to the transmitter REARPANELHI8510T ...

Page 11: ...HI 8512 ORP INDICATOR Keypad 0 mV TEST To verify the instrument cali bration at point 0 500 mV TEST To verify the slope at point 500 mV Trimmers SLOPE For Slope calibration FRONTPANEL SLOPE 0 mV TEST 500 mV TEST mV HI 8512E ORP ...

Page 12: ...g the instrument from the power supply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Recorderoutput 4 BNC socket for ORP electrode REARPANELHI8512E FUSE FUSE ELECTRODE INPUT 220 110 L 1 2 3 4 mA OUT ...

Page 13: ...11 Unplug the instrument from the power supply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Recorderoutput 4 Connections to the transmitter REARPANELHI8512T ...

Page 14: ...EST To display the mV response of the electrode in order to verify its working condition AL To display and set the toler ance of the alarm points pH 7 TEST To verify the internal circuit of the meter in terms of Offset compensation pH 4 TEST To verify the amplifier circuit of the meter FRONTPANEL SLOPE ACID BASE AL 0 COARSE SET FINE SENSOR TEST MEA SURE AL SET pH 7 TEST pH 4 TEST pH HI 8710E pH ...

Page 15: ...alibration AL To set the tolerance of the alarm SET COARSE To coarsely adjust the set point SET FINE To finely adjust the set point Leds ACID To show the acid dosage is active BASE To show the basic dosage is active AL blinking To show the alarm is active ...

Page 16: ...ic dosage selection 4 Reductant or oxidant dosage consent 5 Connections for dosing pump 6 Alarm contacts 7 Recorderoutput 8 ConnectionsforPT100temperaturesensor 9 BNC socket for pH electrode REARPANELHI8710E FUSE FUSE ELECTRODE INPUT PT100 mA OUT OPEN ACID SHORT BASE SET CONSENT ALARM 220 110 L 1 2 3 7 8 9 4 6 5 ...

Page 17: ...upply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Acid Basic dosage selection 4 Reductant or oxidant dosage consent 5 Connections for dosing pump 6 Alarm contacts 7 Recorderoutput 8 Connections to the transmitter REARPANELHI8710T ...

Page 18: ...RE To set HI 8711 on mea surement mode and to en able the diagnostic tests SENSOR TEST To display the mV re sponse of the electrode to verify its working condi tion Al To display and to set the tolerance of the alarm points FRONT PANEL SLOPE AL COARSE ACID SET FINE COARSE FINE BASE SET HI 8711E SENSOR TEST MEA SURE ACID SET AL BASE SET pH 7 TEST pH 4 TEST pH HI 8711E pH ...

Page 19: ...ation AL To set the tolerance of the alarm ACID SET COARSE To coarsely adjust the acid set point ACID SET FINE To finely adjust the acid set point BASE SET COARSE To coarsely adjust the basic set point BASE SET FINE To finely adjust the ba sic set point Leds ACID SET Blinking To show the acid dos age is active BASE SET Blinking To show the basic dosage is active AL Blinking To show the alarm is ac...

Page 20: ...r supply 3 Connections for dosing pump for acid 4 Connections for dosing pump for base 5 Alarm contacts 6 Recorder output 7 Connections for PT100 temperature sensor 8 BNC socket for pH electrode REAR PANEL HI 8711E FUSE FUSE ELECTRODE INPUT PT100 mA OUT BASE ACID ALARM 220 110 L 1 2 6 7 8 3 5 4 ...

Page 21: ... power supply before replacing the fuse 1 Fuse Holder 2 Power supply 3 Connections for dosing pump for acid 4 Connections for dosing pump for base 5 Alarm contacts 6 Recorder output 7 Connections to the transmitter REAR PANEL HI 8711T ...

Page 22: ...measurement mode and to enable the diag nostic tests Al To display and to set the toler ance of the alarm points 0 mV TEST To verify the instrument calibra tion at point 0 500 mV TEST To verify the slope at point 500 mV FRONTPANEL SLOPE DOSAGE OXID REDUC AL COARSE SET FINE AL MEA SURE SET 0 mV TEST 500 mV TEST mV HI 8720E ORP ...

Page 23: ...nd set the tolerance of the alarm points SET COARSE To coarsely adjust the set point SET FINE To finely adjust the set point Leds OXID To show the oxidant dosage is active REDUC To show the reductant dosage is active AL blinking To show the alarm is active ...

Page 24: ...wer supply 3 Oxidant Reductant dosage selection 4 Oxidant or reductant dosage consent 5 Connections for dosing pump 6 Alarm contacts 7 Recorderoutput 8 BNC socket for ORP electrode REARPANELHI8720E FUSE FUSE ELECTRODE INPUT CONSENT mA OUT OPEN OXD SHORT RDX SET ALARM 220 110 L 1 2 3 7 4 8 6 5 ...

Page 25: ...ly before replacing the fuse 1 Fuse Holder 2 Power supply 3 Oxidant Reductant dosage selection 4 Oxidant or Reductant dosage consent 5 Connections for dosing pump 6 Alarm contacts 7 Recorderoutput 8 Connections to the transmitter REARPANELHI8720T ...

Page 26: ...PE SET COARSE FINE O SET MEA SURE Keypad SET To set the working point of pH dosage MEASURE TosetHI931500onmeasure mentmode Trimmers O ForOffsetcalibration SLOPE For Slope calibration SET COARSE Tocoarselyadjustthesetpoint SET FINE To finely adjust the set point Leds ACID To show the acid dosage is active BASE To show the basic dosage is active ...

Page 27: ... 3 Reductant or oxidant dosage consent 4 Acid Basic dosage selection 5 Connections for Dosing Pumps 6 Recorderoutput 7 ConnectionsforPT100temperaturesensor 8 BNC socket for pH electrode REARPANELHI931500 FUSE FUSE ELECTRODE INPUT PT100 mA OUT OPEN ACID SHORT BASE SET CONSENT 220 110 L 1 2 3 5 4 6 7 8 ...

Page 28: ... SET BASE SET FINE COARSE COARSE FINE O ACID SET BASE SET MEA SURE pH Keypad ACID SET To set the working point of acid dosage BASE SET To set the working point of basic dosage MEASURE To set HI 931501 on measure ment mode Trimmers O For Offset calibration SLOPE For Slope calibration ACID SET COARSE To coarsely adjust the acid set point ...

Page 29: ...ET COARSE To coarsely adjust the basic set point BASE SET FINE To finely adjust the ba sic set point Leds ACID SET Blinking To show the acid dos age is active BASE SET Blinking To show the basic dosage is active MEASURE To show the meter is on mea surement mode ...

Page 30: ...6 1 Fuse Holder 2 Power supply 3 Connections for dosing pump for acid 4 Connections for dosing pump for base 5 Recorderoutput 6 ConnectionsforPT100temperaturesensor 7 BNC socket for pH electrode REARPANELHI931501 FUSE FUSE ELECTRODE INPUT PT100 mA OUT BASE ACID 220 110 L 3 4 1 2 ...

Page 31: ...ure mentmode Trimmers SLOPE For Slope calibration SET COARSE To coarsely adjust the ORP set point SET FINE To finely adjust the ORP set point Leds OXID Toshowtheoxidantdosageis active REDUC Toshowthereductantdosage is active FRONTPANEL HI 932500 ORP Process Controller DOSAGE SLOPE OXID mV REDUC SET FINE COARSE MEA SURE SET ...

Page 32: ... Holder 2 Power supply 3 Oxidant or reductant dosage consent 4 Oxidant Reductant dosage selection 5 Connections for dosing pump 6 Recorderoutput 7 BNC socket for ORP electrode REARPANELHI932500 FUSE FUSE ELECTRODE INPUT CONSENT mA OUT OPEN OXD SHORT RDX SET 220 110 L 1 2 4 6 3 7 5 ...

Page 33: ... to 20 mA or 4 to 20 mA isolated Power Supply 110 115 V or 220 240 V 50 60 Hz Environment 10 to 50 C 14 to 122 F Weight 1 kg 2 2 lb SPECIFICATIONS HI 8510 HI8512E HI 8512T Range 1000 to 1000 mV Resolution 1 mV Accuracy 5 mV 0 5 Typical EMC 6 mV 6 mV Deviation 0 2 mA 0 2 mA Installation Cat II Input 1012 Ohm 4 to 20 mA Calibration Slope 90 to 110 slope trimmer Readout 4 digit LCD plus graphic symbo...

Page 34: ...5 V or 220 240 V 50 60 Hz Environment 10 to 50 C 14 to 122 F Weight 1 kg 2 2 lb SPECIFICATIONS HI 8710 HI8710E HI 8710T Range 0 00 to 14 00 pH Resolution 0 01 pH Accuracy 0 02 pH 0 5 Typical EMC 0 1 pH 0 1 pH Deviation 0 2 mA 0 2 mA Installation Cat II Input 1012 Ohm 4 to 20 mA Calibration Offset 2 pH O trimmer Slope 80 to 110 Slope trimmer Temperature Fixed or automatic with PT100 Compensation fr...

Page 35: ...r Supply 110 115 V or 220 240 V 50 60 Hz Environment 10 to 50 C 14 to 122 F Weight 1 kg 2 2 lb SPECIFICATIONS HI 931500 HI8720E HI 8720T Range 1000 to 1000 mV Resolution 1 mV Accuracy 5 mV 0 5 Typical EMC 6 mV 6 mV Deviation 0 2 mA 0 2 mA Installation Cat II Input 1012 Ohm 4 to 20 mA Calibration Slope 90 to 110 Slope trimmer Set Point Relay One Isolated 2 A max 240 V resistive load 1 000 000 strok...

Page 36: ...istive load 1 000 000 strokes Power Supply 110 115 V or 220 240 V 50 60 Hz Environment 10 to 50 C 14 to 122 F Weight 1 kg 2 2 lb SPECIFICATIONS HI 931501 SPECIFICATIONS HI 932500 HI 932500 Range 1000 to 1000 mV Resolution 1 mV Accuracy 5 mV Typical EMC 6 mV Deviation 0 2 mA Installation Cat II Input 1012 Ohm Calibration Slope 90 to 110 Slope trimmer Readout 4 digit LCD plus graphic symbols Recorde...

Page 37: ...TRANSMITTER paying par ticular attention to the indi cated polarity Recorder Terminals these contacts are the output ter minals for connection to a recorder The output is from 0 to 20 mA 4 to 20 mA as indicated and is proportional to the measured pH or ORP value PT 100 Terminals these contacts connect the PT 100 temperature sensor for au tomatic temperature com pensation of pH measure ment If temp...

Page 38: ... oxidants e g in cyanide oxidation make an open circuit be tween oxidation reduction selection terminals see rear panel descriptions 3 at page20 21and 4atpage 28 If you plan to dose reduc tants e g in hexavalent chromium reduction make a short circuit across the above mentioned terminals Set Contacts HI 8710 HI 8720 HI 931500 and HI 932500 only these con tacts maximum 2A 220 V are for connection t...

Page 39: ...oth meters are con nected together to link the ORP and pH controllers so that ORP dosage will occur only if the actual pH value is correct This is to avoid overdosage of oxidants or re ductants which may lead to undesirable pollution The Consent contacts can be left open if HI 8710 or HI 931500 are used indepen dently as pH controllers only The Consent should be shorted if HI 8720 or HI 932500 are...

Page 40: ...s see page 40 42 SET POINTS HI 8710 HI 8720 HI 931500 and HI 932500 only To set the working point of pH or ORP dos age press the SET key The display will show the set value Use a small screwdriver to adjust the trim mers COARSE and FINE until the de sired set value is displayed HI 8711 and HI 931501 only To set the working point of acid dosage press the ACID SET key The display will show the set v...

Page 41: ... set value for base dosage Use a small screwdriver to adjust the trim mers BASE SET COARSE and FINE until the desired base set value is displayed ALARMS HI 8710 HI 8711 and HI 8720 only To set the alarm press the AL key and the display will show the set tolerance for the alarm pH COARSE ACID SET FINE ACID SET BASE SET ACID SET BASE SET COARSE FINE BASE SET pH ACID SET BASE SET COARSE FINE BASE SET...

Page 42: ... 1 5 pH is chosen the instru ment gives an alarm every time the measured pH value is higher than 9 5 pH or lower than 5 5 pH In HI 8720 if the set value is 300 mV and the Alarm of 100 mV is chosen the instrument gives an alarm every time the measured ORP value is higher than 400 mV or lower than 200 mV MEASUREMENTS After setting the pH or ORP value and the eventual alarm immerse the electrode in t...

Page 43: ...be lighted HI 8710 HI 931500 and HI 931501 only When oxidants are dosed the OXID LED will be lighted and when reductants are being dosed the REDUC LED will be lighted HI 8720 HI 932500 only SLOPE DOSAGE ACID BASE AL COARSE SET FINE MEA SURE SET SLOPE DOSAGE OXID REDUC AL COARSE SET FINE MEA SURE SET ...

Page 44: ...hould be lo cated as close to the pH electrode as possible Shake briefly and wait one minute before adjusting the O trimmer to display pH 7 01 on the LCD if the temperature of the buffer solution is at 25 C 77 F If the temperature of the buffer solution is not 25 C 77 F refer to the chart on page 43 for the appropriate buffer value to adjust to the noted temperature Rinse the electrode and tem per...

Page 45: ...n the LCD if the temperature of the buffer solution is at 25 C if not refer to the chart on page 43 for appropriate buffer value for the corresponding temperature The calibration is now complete and the in strument is ready for use Note If the meter is used in conjunction with a PT 100 temperature sensor immerse it into the buffers during the calibration procedure DOSAGE ACID BASE SLOPE SET COARSE...

Page 46: ...06 on the display If the buffer s temperature is 50 C calibrate to read 4 06 6 98 9 82 on the display TEMP pH VALUES C F 4 01 6 86 7 01 9 18 10 01 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 32 41 50 59 68 77 86 95 104 113 122 131 140 149 158 4 01 4 00 4 00 4 00 4 00 4 01 4 02 4 03 4 04 4 05 4 06 4 07 4 09 4 11 4 12 6 98 6 95 6 92 6 90 6 88 6 86 6 85 6 84 6 84 6 83 6 83 6 84 6 84 6 85 6 85 7 13 7 1...

Page 47: ...unctioning of the instrument or electrode Press the MEASURE button before pressing any of the fol lowing keys A Sensor Test Immerse the electrode in pH 7 01 buffer solution HI 7007 press the SEN SOR TEST key and the display shows the mV re sponse of the electrode If the electrode is in good working condi tion the value should be within 30 mV A value between 30 mV and 60 mV or be tween 60 and 30 mV...

Page 48: ...between 7 pH 1pH This will verify the internal circuit of the meter in terms of the offset compen sation C Amplifier Circuit Test Press the pH4 TEST key and the dis play should show a value between 3 30 pH and 4 30 pH This will verify the ampli fier circuit of the meter pH pH 7 TEST pH 4 TEST pH ...

Page 49: ...e of malfunction Press the MEASURE key be fore proceeding with the follow ing tests HI 8720 only A 0 mV Test Press the 0 mV TEST key and the display should show a value of 0 mV 10 mV This will verify in strument calibration at point 0 B 500 mV Test Press the 500 mV TEST key and the display should show a value of 500 mV 20 mV This will verify the slope at point 500 mV COARSE SET FINE MEA SURE 0 mV ...

Page 50: ... states A Light stays on The mode is displayed on the LCD but is not active E g alarm set point is displayed but the alarm contact is open B Light Blinking 25 On 75 Off The mode is not dis played on the LCD but the mode is active E g alarm contact is closed but the alarm set point is not displayed C Light Blinking 75 On 25 Off The mode is active and being displayed D Light Off The function is neit...

Page 51: ...leasing elec trons without causing a chemical reaction with the elements with which it comes into contact The electrodes most usually available for this purpose have gold or platinum surfaces gold possesses a higher resistance than platinum in conditions of strong oxidation while plati num is preferred for the measurements of oxidizing solutions containing halides and for more general uses When a ...

Page 52: ...er than the value in the Table below an oxidizing pre treatment is necessary other wise a reducing pre treatment is necessary pH mV pH mV pH mV pH mV pH mV 0 990 1 920 2 860 3 800 4 740 5 680 6 640 7 580 8 520 9 460 10 400 11 340 12 280 13 220 14 160 Reducing pre treatment immerse the elec trode for some minutes in HI 7091 Oxidizing pre treatment immerse the elec trode for some minutes in HI 7092 ...

Page 53: ...measure the response the obtained value should be within 200 and 275 mV After this functional test it is suggested to wash the electrode thoroughly with water and proceed to the oxidizing or reducing pre treatment before taking measurements When not in use the electrode tip should be kept moist and far from any type of mechani cal stress which might cause damage For this reason the use of the prot...

Page 54: ...r may have formed inside the glass bulb The electrode Reference Filling Hole Reference Filling Hole Fill Hole Screw Sensitive Wire Reference Wire Reference Junction Glass Bulb Glass Bulb Reference Junction Reference Junction Reference Junction Platinum or Gold tip Platinum or Gold tip Reference Wire Reference Wire Sensitive Wire Glass Body pH Electrode Plastic Body pH Electrode Plastic Body ORP El...

Page 55: ... electrode does not respond to pH changes the battery is run down and the electrode should be replaced MEASUREMENT Rinse the electrode tip with distilled water Immerse the tip 4 cm 1 in the sample and stir gently for approx 30 seconds For a faster response and to avoid cross contamination of the samples rinse the elec trode tip with a few drops of the solution to be tested before taking measuremen...

Page 56: ...e electrode Rinse off any salt depos its with water For refillable electrodes Refill the electrode with fresh electrolyte HI7071 for single junction or HI7082 for double junction electrodes Allow the electrode to stand upright for 1 hour Follow the Storage Procedure above CLEANING PROCEDURE General Soak in Hanna HI7061 General Cleaning Solution for approxi mately 1 hour Removal of films dirt or de...

Page 57: ...e Loss of shielding due to low electrolyte level in refillable electrodes only refill with HI7071 for single junction or HI7082 for double junction electrodes Dry Membrane Junction Soak in Stor age Solution HI70300 for at least 1 hour Drifting Soak the electrode tip in warm Hanna Solution HI7082 for one hour and rinse tip with distilled water Refill with fresh HI7071 for single junction electrodes...

Page 58: ...56 and then follow the Cleaning Procedure above For ORP Electrodes polish the metal tip with a light abrasive paper paying atten tion not to scratch the surface and wash thoroughly with water ...

Page 59: ...intain proper operating conditions The conventional electrodes should be used in indoor applications only and should not use a cable longer than 10 m 33 MEDIUM DISTANCE INDOOR OUTDOOR INSTALLATION When an outdoor installation is required it is necessary to install a transmitter to obtain accurate readings at distances from 10 to 50 m 33 165 Since the introduction of AmpHel these dis tances are no ...

Page 60: ...greater than 50 m 165 it is necessary to install a transmitter between the electrode and the meter Hanna offers a full line of pH ORP transmit ters with or without LCD displays AmpHel isaregisteredTrademarkof HannaInstruments pH METER WITH A pH TRANSMITTER AND CONVENTIONAL ELECTRODE MEASUREMENTS CAN BE TAKEN FROM DISTANCES UP TO 300 METERS 1000 FEET WET ENVIRONMENT RELATIVE HUMIDITY UP TO 100 3m M...

Page 61: ...30 mL HI 7091L Pre Treatment Reducing Solution 460 mL HI 7092M Pre treatment Oxidizing Solution 230 mL HI 7092L Pre Treatment Oxidizing Solution 460 mL ELECTRODE STORAGE SOLUTIONS HI70300M Storage Solution 230 mL HI70300L Storage Solution 460 mL ELECTRODE CLEANING SOLUTIONS HI7061M General Cleaning Sol 230 mL HI7061L General Cleaning Sol 460 mL HI7073M Protein Cleaning Sol 230 mL HI7073L Protein C...

Page 62: ... 3 BNC connector 3 m 9 9 cable doublejunctionplastic body HI 1114S HI 1134B 3 HI 1115S Screwconnector singlejunction glass body HI1135B 3 BNC connector 3 m 9 9 cable single junction glass body HI 1115S HI 1135B 3 PG13 5 THREAD φ 12mm φ 9 5mm 110mm 30mm DIA 9 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm M13 x 1 5 DIA 16 mm 7 mm 25 mm PG13 5 THREAD φ 12mm φ 9 5mm 110mm 30mm DIA 20 5mm 3 4 x 16 ...

Page 63: ...connector 1 m 3 3 cable double junction plastic body built inamplifier HI1912B 5 BNCconnector 5m 16 5 cable double junction plastic body built inamplifier HI2114B 5 BNCconnector 5m 16 5 cable double junction plastic body PG13 5 THREAD φ 12mm 110mm 30mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm DIA 9 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm...

Page 64: ...13 5connector Pt glass body HI 3115S Screw typeconnector side arm Pt glass body HI3135B 3 BNC connector 3 m 9 9 cable side arm Pt glass body HI 3115S HI 3135B 3 DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm M13 x 1 5 DIA 16 mm 7 mm 25 mm PG13 5 THREAD φ 12mm φ 1mm 110mm 30mm DIA 12mm DIA 7 6mm 150mm 25mm DI...

Page 65: ...I4130B 3 BNC connector 3 m 9 9 cable Au glass body HI 4110S HI 4130B 3 HI4932B 5 BNCconnector 5m 16 5 cable Au Ultem body built in ampli fier PG13 5 THREAD φ 12mm 110mm 30mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm M13 x 1 5 DIA 16 mm 7 mm 25 mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm DIA 20 5mm 3 4 x 16 UNF DIA 12mm 110mm 38 5mm M13 x 1 5 DIA 16 mm 7 mm 25 mm Ultem is a registered Trad...

Page 66: ...g Pumps with Flow Rate from 1 5 to 20 LPH HI 7871 LevelControllers HI 7873 HI6050 SubmersibleElectrodeHolders HI 6051 HI 6054 ElectrodeHoldersforIn Line HI 6057 Applications HI778P Screened Coaxial Cable and Connectors for Screw type pH ORP Electrodes HI8427 pHandORPElectrodeSimulator with 1 m 3 3 Coaxial Cable ending in Female BNC Connectors HI7858 1 HI931001 pHandORPElectrodeSimulator with LCD D...

Page 67: ...number and the nature of the failure Obtain a Returned Goods Authorization from the Customer Service department first and then return the instrument with the Authorization included along with shipment costs prepaid If the repair is not covered by the warranty you will be notified of the charge for repair or replacement When shipping any instrument make sure it is properly packaged for complete pro...

Page 68: ... performance To avoid electrical shock do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC To avoid damages or burns do not perform any measurement in microwave ovens DECLARATION OF CONFORMITY We Hanna Instruments Srl V le delle industrie 12 35010 Ronchi di Villafranca PD ITALY herewith certify that the process controllers have been tested and found to be in...

Page 69: ...eatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constantly being added to the library For these and other catalogs handbooks and leaflets contact your dealer or the Hanna Customer Service nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com HANNALITERATURE ...

Page 70: ...68 h t t p w w w h a n n a i n s t c o m MANPROCR2 01 02 ...

Reviews: