7
SPECIFICATIONS
Titrator
Range
Titratable alkalinity (low range)
mg/L
: 15.0 - 500.0 mg/L as CaCO
3
meq/L : 0.3 - 10.0 meq/L
as CaCO
3
Titratable alkalinity (high range)
mg/L
: 400 - 4000 mg/L as CaCO
3
meq/L : 8 - 80 meq/L
as CaCO
3
Resolution
Titratable alkalinity (low range):
0.1 mg/L
0.1 meq/L
Titratable alkalinity (high range):
1mg/L
1meq/L
Accuracy
5% of reading
Titration method Acid-base titration (total alkalinity / strong alkalinity)
Principle
End point titration: 4.50 pH / 8.30 pH
Pump debit
0.5 mL/min
Stirring speed
600 rpm
Log data
Up to 50 samples
pH meter
pH meter
-2.0 to 16.0 pH / -2.00 to 16.00 pH
pH Resolution:
0.1 pH / 0.01 pH
pH Accuracy:
± 0.01 pH
pH Calibration:
1, 2 or 3 calibration points;
3 available buffers (4.01; 8.30; 10.01)
Temperature
manual or automatic from
compensation:
-20 to 120 °C (-4 to 248 °F)
Log data
Up to 50 samples
Temperature
Range
-20.0 to 120.0 °C (-4.0 to 248.0 °F)
Resolution
0.1 °C
Accuracy
±0.4 °C without probe error
Electrode
HI 1131B
(included)
Temperature Probe
HI 7662-M
(included)
Environment
0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Power supply
12 Vdc power adapter
Dimensions
208 × 214 × 163 mm (8.2×8.4×6.4”) (with beaker)
Weight
2200 g (77 oz.)