10

When the buffer value is recognized and the
reading is stable, an acoustic signal (if en-
abled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.

Press CFM to store the first calibration point.

The meter expects a pH 4.01 or 10.01 buffer.

Rinse and immerse the pH electrode and the
temperature probe in pH 4.01 or pH 10.01
buffer (2

nd

 calibration point) and stir gently.

When the buffer value is recognized and the
reading is stable, an acoustic signal (if en-
abled) advises the user, the "pH" symbol stops
flashing, the hourglass indicator disappears
and the "CFM" tag starts blinking to indicate
that the value can be confirmed.

Press CFM to store the second calibration point.

The meter returns to normal mode.

The pH calibration is now complete; "CAL" and the
pH tags corresponding to the buffers used for
calibration are lit on the LCD.

Notes

:

• If the buffer value is not recognized, after 12 seconds the meter will display

blinking dashes together with the "WRONG"
tag. Either the buffer solution is wrong or out
of specification and needs to be replaced or the
electrode is damaged.

• The meter will retain the calibration if the battery is removed.
• To quit calibration and keep previous data: press CAL after entering the

calibration mode and before the first point is accepted.

• To perform a single-point calibration: press CAL after the first point has been

confirmed.

If the temperature probe is not connected and manual temperature compensa-
tion is required, follow the procedure below:

Press RANGE to select the temperature mode.

Rinse the pH electrode and place it into the pH 7.01 buffer, stir briefly and
wait a few minutes to reach thermal equilibrium.

19

pH CALIBRATION SOLUTIONS

HI 70004P

pH 4.01 buffer solution, 20 mL sachet (25 pcs)

HI 7004M

pH 4.01 buffer solution, 230 mL bottle

HI 7004L

pH 4.01 buffer solution, 500 mL bottle

HI 8004L

pH 4.01 buffer solution, 500 mL FDA bottle

HI 70007P

pH 7.01 buffer solution, 20 mL sachet (25 pcs)

HI 7007M

pH 7.01 buffer solution, 230 mL bottle

HI 7007L

pH 7.01 buffer solution, 500 mL bottle

HI 8007L

pH 7.01 buffer solution, 500 mL FDA bottle

HI 70010P

pH 10.01 buffer solution, 20 mL sachet (25 pcs)

HI 7010M

pH 10.01 buffer solution, 230 mL bottle

HI 7010L

pH 10.01 buffer solution, 500 mL bottle

HI 8010L

pH 10.01 buffer solution, 500 mL FDA bottle

STORAGE & CLEANING SOLUTIONS

HI 70300M

Storage solution, 230 mL bottle

HI 80300M

Storage solution, 230 mL FDA bottle

HI 70300L

Storage solution, 500 mL bottle

HI 80300L

Storage solution, 500 mL FDA bottle

HI 70000P

Electrode rinsing solution, 20 mL sachet (25 pcs.)

HI 7061M

General cleaning solution, 230 mL bottle

HI 8061M

General cleaning solution, 230 mL FDA bottle

HI 7061L

General cleaning solution, 500 mL bottle

HI 8061L

General cleaning solution, 500 mL FDA bottle

HI 7073M

Protein cleaning solution, 230 mL bottle

HI 8073M

Protein cleaning solution, 230 mL FDA bottle

HI 7073L

Protein cleaning solution, 500 mL bottle

HI 8073L

Protein cleaning solution, 230 mL FDA bottle

HI 7074M

Inorganic cleaning solution, 230 mL bottle

HI 7074L

Inorganic cleaning solution, 500 mL bottle

HI 7077M

Oil & Fat cleaning solution, 230 mL bottle

HI 8077M

Oil & Fat cleaning solution, 230 mL FDA bottle

HI 7077L

Oil & Fat cleaning solution, 500 mL bottle

HI 8077L

Oil & Fat cleaning solution, 500 mL FDA bottle

REFILLING ELECTROLYTE SOLUTIONS

HI 7071

3.5 M KCl+AgCl electrolyte solution, 4 x 30 mL bottle, for single
junction electrodes

HI 8071

3.5 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for
single junction electrodes

HI 7072

1 M KNO

3

 electrolyte solution, 4 x 30 mL bottle

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

Summary of Contents for HI 8424

Page 1: ...Instruction Manual HI 8424 Portable pH mV C Meter MAN8424W R2 09 10 PrintedinEUROPE ROMANIA Hanna Instruments Inc T Local Sales and Customer Service Office ...

Page 2: ...ru ment If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instruments are to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid When s...

Page 3: ...an set the following parameters by entering the setup mode through the keyboard enabling or disabling the auto off feature enabling or disabling the acoustic signal selecting the temperature unit C or F GENERAL DESCRIPTION GENERAL DESCRIPTION GENERAL DESCRIPTION GENERAL DESCRIPTION GENERAL DESCRIPTION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION P...

Page 4: ...0S 5 Cheese FC 200B FC200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1230B HI 1210S 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 1331B HI1310S 11 Foodindustrygeneraluse FC 911B FC 100B 12 Fruit FC 200B FC 220B FC 200S 13 Fruitjuices organic FC 210B 14 Galvanizing waste solution HI 1130B HI1110S 15 Highpuritywater HI 1053B HI1050S 16 Horticulture HI 1053B FC 200B H...

Page 5: ...pC pocket size thermometer 50 0 to 150 0 C HI 98502 ChecktempF pocket size thermometer 58 0 to 302 0 F HI 710015 Shockproof rubber boot blue HI 710016 Shockproof rubber boot orange HI 710022 Spare protective case HI 76405 Electrode holder HI 7662 Temperature probe with 1 m 3 3 screened cable HI 8427 pH ORP electrode simulator with 1 m 3 3 coaxial cable and BNC connector HI 931001 pH ORP electrode ...

Page 6: ...display will show the pH value automatically compensated for temperature Notes In order to take accurate pH measurements make sure that the instrument has been calibrated before use see page 9 If measurements are taken in different samples successively it is recom mended to rinse the electrode thoroughly to avoid cross contamination After cleaning it is recommended to rinse the electrode with some...

Page 7: ...case at 25 C ORP MEASUREMENTS Oxidation Reduction Potential ORP measurements provide a quantification of the oxidizing or reducing power of the sample tested Connect the ORP electrode optional to the BNC connector To enter the mV mode turn the instrument ON and press the RANGE key until the display changes to mV Submerge the ORP electrode tip at least 4 cm 1 into the sample to be tested and allow ...

Page 8: ... HOLD again to return to normal mode If enabled keypresses are followed with an acoustic signal A lower note indicates that the key is not currently active To save battery life the meter is provided with an auto off feature which turns the instrument off after 20 minutes of non use This feature can be disabled by the user see Menu selection for details 21 120 mm 4 7 5mm 0 2 5mm DIA 0 2 S VERSION H...

Page 9: ...submerged approximately 4 cm 1 into the solution The temperature probe should be located close to the pH electrode Press RANGE to display pH measurement Press CAL to enter the calibration mode The buffer value 25 C 77 F and the pH symbol will blink on the display The meter expects a pH 7 01 buffer 20 HI 7082 3 5 M KCl electrolyte solution 4 x 30 mL bottle for double junction electrodes HI 8082 3 5...

Page 10: ...m 19 pH CALIBRATION SOLUTIONS HI 70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI 7004M pH 4 01 buffer solution 230 mL bottle HI 7004L pH 4 01 buffer solution 500 mL bottle HI 8004L pH 4 01 buffer solution 500 mL FDA bottle HI 70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI 7007M pH 7 01 buffer solution 230 mL bottle HI 7007L pH 7 01 buffer solution 500 mL bottle HI 8007L pH 7 01 buffer...

Page 11: ...ter immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution to be tested before taking any measurements STORAGE PROCEDURE To minimize clogging and ensure a quick response time the glass bulb and the junction should always be kept moist When not in use store it with a few drops ...

Page 12: ...UP to move to next step At this point it is possible to set the temperature unit by selecting Unit C or Unit F Press CAL SELECT to toggle the selection and CFM SETUP to exit the menu selection mode and return to normal measurement mode MENU SELECTION MENU SELECTION MENU SELECTION MENU SELECTION MENU SELECTION 17 ELECTRODE CONDITIONING MAINTENANCE ELECTRODE CONDITIONING MAINTENANCE ELECTRODE CONDIT...

Page 13: ... Dirty pH electrode Soak the electrode tip excessive drift in HI 7061 solution for 30 minutes Reading fluctuates Clogged dirty junction Soak the electrode tip up and down noise or low electrolyte level in warm HI 7082 refillable electrodes solution for one hour then rinse it with distilled water refill with fresh electrolyte if necessary Blinking full scale Reading is out of range value Blinking C...

Page 14: ... MESSAGES TROUBLESHOOTING LCD MESSAGES TROUBLESHOOTING LCD MESSAGES TROUBLESHOOTING LCD MESSAGES TROUBLESHOOTING TAGS SYMBOLS pH mV C F Measurement unit of the selected mode ATC Indicates Automatic Temperature Compensation in pH or temperature mode MTC Indicates Manual Temperature Compensation in pH or temperature mode HOLD Blinks when in Hold mode Reading frozen on LCD The user can scroll through...