background image

21

5 mL x 2

• Remove the cap and fill the

Demineralizer Bottle with tap
water.

• Replace the cap and shake

gently for at least 2 minutes.

• Open the upper part of the

Demineralizer Bottle cap and squirt
gently the demineralized water into
the cylinder, up to the 100 mL mark.
Note: The ion exchange resin

contained in the Demineralizer Bottle
is provided with an indicator
substance. The indicator will change
from green to blue when the resin
has been exhausted and the resin
has to be replaced.

• Pour the solution in the large 170mL

beaker, replace the cap and invert
several times to mix.

• If the solution contains some turbidity

or color, add a powder packet of
active carbon and follow the
procedure described for 

Irrigation

Water

.

NUTRIENTS SOLUTION (HR):

• Use the 5 mL syringe twice to add to

the graduated cylinder exactly 10
mL of sample.
Note:

 

To measure exactly 5 mL of

40

Phosphorus LR

x 10

• Wait for a few seconds and the

display will show "-0.0-". Now
the meter is zeroed and ready for
measurement.

• Remove the cuvet.

• Add 10 drops of HI 93706A

Molybdate reagent.

• Add the content of one packet of

HI 93706B Phosphorus Reagent
B (Amino Acid) to the cuvet.
Replace the cap and shake gently
until dissolution is complete.

• Reinsert the cuvet into the

instrument.

• Press TIMER and the display will show the countdown prior to the

measurement or, alternatively, wait for 5 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.

• The instrument directly displays concentration in mg/L of phosphorus

(P) on the Liquid Crystal Display.

• To convert the reading to mg/L of phosphate (PO

4

3–

), multiply by

a factor of 3.07.

INTERFERENCES

Calcium above 10000 mg/L as CaCO

3

Chloride above 150000 mg/L as Cl

Ferrous iron above 100 mg/L
Magnesium above 40000 mg/L as CaCO

3

Sulfide

Summary of Contents for HI 83225

Page 1: ...Instruction Manual HI 83225 Grow Master for Nutrient Analyses This Instrument is in Compliance with the CE Directives w w w h a n n a i n s t c o m MAN83225R1 04 05 w w w h a n n a i n s t c o m...

Page 2: ...RUS MR 41 PHOSPHORUS HR 43 POTASSIUM LR 45 POTASSIUM MR 47 POTASSIUM HR 49 POTASSIUM UHR 50 SULFATE 51 INTERFACE WITH PC 53 STANDARD METHODS 55 BATTERY REPLACEMENT 55 ACCESSORIES 56 WARRANTY 57 CE DEC...

Page 3: ...83215 Nutrient Analyses HI 83216 Pool Spa Applic HI 83218 Environmental Applic HI 83225 Nutrient Analyses HI 83226 Pool Spa Applic All meters use an exclusive positive locking system to ensure that th...

Page 4: ...ce of the substance as the other factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reage...

Page 5: ...ly identi cal to provide the same measurement conditions Whenever possible use the same cuvet for both It is also necessary that the surface of the cuvet is clean and not scratched This is to avoid me...

Page 6: ...to 100 mg L Turbidimetric Potassium HR 12 20 to 200 mg L Turbidimetric Calcium 13 0 to 400 mg L Oxalate Magnesium HR 14 0 to 150 mg L Calmagite Sulfate 15 0 to 150 mg L Turbidimetric Besides to the 1...

Page 7: ...that the meter is in a ready state and zeroing can be performed This indicates that the batteries are dead and must be replaced GUIDE TO DISPLAY CODES 54 PUP Program Up PDN Program Down PTM Turn Test...

Page 8: ...l able from your Hanna Dealer Make sure that your meter is switched off and plug the connectors one into the meter RS 232C socket the other into the serial port of your PC Note Cables other than HI 92...

Page 9: ...n sample reading There is too much light for the sample measurement Please check if the right sample cuvet is inserted Wait for a few seconds and the display will show 0 0 Now the meter is zeroed and...

Page 10: ...ent is added or accuracy will be lost In order to avoid reagent leaking and to obtain more accurate measurements it is recommended to close the cuvet first with the supplied HDPE plastic stopper and t...

Page 11: ...rement Remove the cuvet and add the content of one packet of HI 93750B reagent Re place the cap and mix gently for one minute by turning slowly the cuvet up side down Reinsert the cuvet into the instr...

Page 12: ...Range 20 to 200 mg L Resolution 5 mg L Accuracy 30 mg L 7 of reading Light Source Tungsten lamp with narrow band interference filter 610 nm Method Adaptation of the Turbidimetric Tetraphenylborate me...

Page 13: ...UHR UHR Ultra High Range ABBREVIATIONS 48 Press ZERO and SIP will appear on the display Wait for a few seconds and the display will show 0 0 Now the meter is zeroed and ready for measurement Remove t...

Page 14: ...ps to absorb other nutrients in particular phosphorus assists a bigger production for both size and number of fruits SHORTAGE slower growth EFFECTS smaller leaves yellowing of leaves smaller fruits pr...

Page 15: ...used by fungi increases the cuticular thickness regulates the cellular turgidity helps to regulate the osmotic processes and increases the resistance to dryness regulates the stomatic opening and clos...

Page 16: ...SHORTAGE young leafs become distorted EFFECTS bud development is inhibited root tips often die plant growth stunted with short internodes EXCESS EFFECTS no consistent visual symptoms usually associate...

Page 17: ...g LasCl Ferrousironabove100mg L Magnesiumabove40000mg LasCaCO3 Sulfide x 10 Wait for a few seconds and the display will show 0 0 Now the meter is zeroed and ready for measurement Remove the cuvet Add...

Page 18: ...ation or irrational liquid manure spreading The analysis of irrigation waters allows us to find out which are the substances present in major or minor quantity and to organize an advantageous fertiliz...

Page 19: ...TION WATER LR Measure with the graduated cylinder about 100 mL of sample If the solution contains some turbidity or color pour it in the large 170mL beaker and add a powder packet of active carbon SAM...

Page 20: ...mL of sample 41 SPECIFICATIONS Range 0 0 to 50 0 mg L Resolution 0 5 mg L Accuracy 2 5 mg L 4 of reading Light Source Tungsten lamp with narrow band interference filter 525 nm Method Adaptation of th...

Page 21: ...ce to add to the graduated cylinder exactly 10 mL of sample Note To measure exactly 5 mL of 40 Phosphorus LR x 10 Wait for a few seconds and the display will show 0 0 Now the meter is zeroed and ready...

Page 22: ...uracy 0 5 mg L 4 of reading Light Source Tungsten lamp with narrow band interference filter 525 nm Method Adaptation of the Standard Methods for the Examination of Water and Wastewater 18th edition Am...

Page 23: ...LOW RANGE 23 Ammonia MR 38 Nitrate HR Remove the cuvet and add the content of one packet of HI 93728 reagent Replace the cap and imme diately shake vigorously for exactly 10 seconds by mov ing the cu...

Page 24: ...the meter is zeroed and ready for measurement Ammonia LR x 4 37 SPECIFICATIONS Range 0 to 300 mg L Resolution 5 mg L Accuracy 10 mg L 8 of reading Light Source Tungsten lamp with narrow band interfere...

Page 25: ...onia MR AMMONIA MEDIUM RANGE 36 Nitrate MR Remove the cuvet and add the content of one packet of HI 93728 reagent Replace the cap and imme diately shake vigorously for exactly 10 seconds by mov ing th...

Page 26: ...w 0 0 Now the meter is zeroed and ready for measurement x 4 35 SPECIFICATIONS Range 0 to 150 mg L Resolution 2 0 mg L Accuracy 5 mg L 8 of reading Light Source Tungsten lamp with narrow band interfere...

Page 27: ...AMMONIA HIGH RANGE 34 Remove the cuvet and add the content of one packet of HI 93728 reagent Replace the cap and imme diately shake vigorously for exactly 10 seconds by mov ing the cuvet up and down C...

Page 28: ...e meter is zeroed and ready for measurement x 4 Ammonia HR 33 SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 5 mg L Accuracy 1 0 mg L 8 of reading Light Source Tungsten lamp with narrow band inter...

Page 29: ...10 mL mark with the HI 93752A 0 Ca Buffer reagent CALCIUM Calcium 29 3 mL of sample x4 3 mL probable level of liquid taken up by syringe Press ZERO and SIP will blink on the display Wait for a few se...

Page 30: ...er Alkalinity as CaCO3 above 1000 mg L higher Magnesium Mg2 above 400 mg L higher 1 mL Calcium 30 SPECIFICATIONS Range 0 to 150 mg L Resolution 5 mg L Accuracy 5 mg L 3 of reading Light Source Tungste...

Reviews: