background image

68

69

Zinc

ZINC

SPECIFICATIONS

Range

0.00 to 3.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the 

Standard Methods for the Examination of Water and Wastewater,

18

th

 edition,

 Zincon method. The reaction between zinc and the reagents causes an

orange to a dark violet tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93731

A

-0

Zinc Reagent

1 packet

HI 93731

B

-0

Cyclohexanone

0.5 mL

REAGENT SETS

HI 93731-01

Reagents for 100 tests

HI 93731-03

Reagents for 300 tests

For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the 

Zinc

 method using the procedure described in the

Method Selection

 section (see page 12).

• Fill one graduated mixing cylinder up to the 20 mL mark with

the sample.

• Add the content of one packet of HI 93731A-0 zinc reagent, close

the cylinder and invert several times to mix until completely
dissolved.

• Fill one cuvette with 10 mL of the reacted sample up to the mark.

20 mL

10 mL

Silver

• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Fill a second cuvette up to the mark with 10 mL of the reacted sample.

• Insert the second cuvette into the instrument.

• Press READ to start the reading. The instrument displays

the results in mg/L of silver.

Note

a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,

add 2-3 drops of Dispersing agent (HI 93703-51) to each cuvette and swirl until turbidity is completely
dissolved.

INTERFERENCES

Interference may be caused by:
Al

3+

 above 30 mg/L

Fe

2+

 above 1.5 mg/L

Ca

2+

 above 1000 mg/L as CaCO

3

Fe

3+

 above 10 mg/L

Cd

2+

 above 20 mg/L

K

+

 above 500 mg/L

Cl

-

 above 8000 mg/L

Mn

2+

 above 25 mg/L

Co

2+

 above 1.5 mg/L

Mg

2+

 above 1000 mg/L as CaCO

3

Cr

3+

 above 20 mg/L

Na

+

 above 5000 mg/L

Cr

6+

 above 40 mg/L

Ni

2+

 above 1.5 mg/L

Cu

2+

 above 15 mg/L

Pb

2+

 above 20 mg/L

F

-

 above 20 mg/L

Zn

2+

 above 30 mg/L

10 mL

# 2

# 2

Summary of Contents for HI 83208

Page 1: ...1 Instruction Manual HI 83208 Multiparameter Bench Photometer for Water Conditioning w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Page 2: ...HI 92000 Windows Compatible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrumen...

Page 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Page 4: ...elect method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to access...

Page 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Page 6: ...t blank correction The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stability and storing co...

Page 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Page 8: ...n enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Page 9: ...nteractive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To read all availabl...

Page 10: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 11: ...ette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Ammonia LR Remove the cuvette Add 4 drops o...

Page 12: ...O key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap an...

Page 13: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 14: ...the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement 10ml Press TIMER and the display will show the countdown prior to the measurement or alternatively wa...

Page 15: ...ion section see page 12 Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 wh...

Page 16: ...3729 0 SPADNS reagent to two cuvettes Fill one of the cuvettes with distilled water up to the mark replace the cap and invert several times to mix Fill the other cuvette with sample up to the mark rep...

Page 17: ...r measurement Remove the cuvette and add the content of one packet of HI 93721 0 reagent Replace the cap and shake until dissolution is complete 10 mL Press TIMER and the display will show the countdo...

Page 18: ...ee page 12 Fill one graduated mixing cylinder up to the 25 mL mark with deionized water Add the content of one packet of HI 93736 0 TPTZ reagent close the cylinder and shake vigorously for 30 seconds...

Page 19: ...hen the meter is zeroed and ready for measurement Remove the cuvette Fill another graduated mixing cylinder up to the 25 mL mark with the sample Add the content of one packet of HI 93746 0 TPTZ reagen...

Page 20: ...mg L 3 of reading Typical EMC 0 1 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 525 nm Method Adaptation of the Standard Methods for the Examination of Water and Waste...

Page 21: ...iation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN Method The reaction between manganese and the reagents causes an orange tint in the sample RE...

Page 22: ...ple up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Manganese LR Pres...

Page 23: ...er above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one gra...

Page 24: ...he reaction between nickel and the reagent causes a blue tint in the sample REQUIRED REAGENTS Code Description Quantity HI 93726 0 Powder reagent 1 packet REAGENT SETS HI 93726 01 Reagents for 100 tes...

Page 25: ...ical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between nickel and the reagents causes an orange t...

Page 26: ...cedure described in the Method Selection section see page 12 Using the pipette fill the cuvette with 6 ml of sample up to half of its height and replace the cap Place the cuvette into the holder and c...

Page 27: ...e that a part of the sample spills over This is to make sure that no air bubbles have been trapped inside which could alter the reading Invert several times the bottle The sample becomes orange yellow...

Page 28: ...cuvette and add 5 drops of HI 93710 0 Phenol Red Indicator Replace the cap and mix the solution 10 mL x 5 Dissolved Oxygen Let the sample stand and the flocculent agent will start to settle After app...

Page 29: ...93717B 0 Reagent B 1 packet REAGENT SETS HI 93717 01 Reagents for 100 tests HI 93717 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Phosphate HR method us...

Page 30: ...Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93713 0 reagent Replace the cap and shake gently...

Page 31: ...reacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove...

Page 32: ...te reagent Replace the cap and swirl the solution Wait for 4 minutes add the content of one packet of HI 93705B 0Citricacidreagentandshakeuntilitiscompletely dissolved Wait for 1 minute This is the bl...

Page 33: ...e the cuvette and add the content of one packet of HI 93705C 0 Amino acid reagent and shake until it is completely dissolved Reinsert the cuvette into the instrument Press TIMER and the display will s...

Page 34: ...uracy 0 005 mg L 10 of reading Typical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between silver a...

Page 35: ...reagent close the cylinder and invert several times to mix until completely dissolved Fill one cuvette with 10 mL of the reacted sample up to the mark 20 mL 10 mL Silver Press ZERO key The display wil...

Page 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Page 37: ...CCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chlorine tests powder HI 93701 F 300 free chlo...

Page 38: ...omer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make sure that they are entire...

Page 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Reviews: