Hanna Instruments HI 709 Instructions Download Page 2

Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “

Add

”, “

C.1

” with “

Press

blinking, the meter is ready.

Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.

Press the button. When the display shows

Add

”, “

C.2

” with “

Press

” blinking the

meter is zeroed.

Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of 

HI 709A-0

 reagent. Replace the

cap and shake gently for 2 minutes.

Unscrew the cap and add the content of one
packet of 

HI 709B-0

 reagent. Replace the

cap and invert gently for 2 minutes. Place
the cuvette back into the meter.

Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 1 minute and
30 seconds and press the button.

The instrument displays the concentration in
ppm of manganese. The meter automatically
turns off after 10 minutes.

10 mL

REAGENT SETS
HI 709-25

Reagent for 25 Manganese High Range Tests

OTHER ACCESSORIES
HI 709-11

Manganese HR Certified Standard Kit

HI 731318

Cloth for wiping cuvettes (4 pcs.)

HI 731321

Glass cuvettes (4 pcs.)

HI 731225

Black Cuvette Cap for Checker HC (4 pcs.)

HI 740028

1.5V AAA batteries (4 pcs.)

HI 93703-50

Cuvettes cleaning solution (230 mL)

Recommendations for Users

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find a Hanna Office in your area,

visit our web site:

w w w . h a n n a i n s t . c o m

Hanna Instruments 

reserves the right to modify the design, construction or

appearance of its products without advance notice.

Battery management

IST709   07/15

Measurement procedure:

Accessories:

Tips for an accurate measurement

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry

outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with

HI 731318

 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To obtain

accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.

• Do not let the reacted sample stand for too long after reagent is added, as

accuracy will be affected.

• After the reading it is important to immediately discard the sample, otherwise

the glass might become permanently stained.

To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead
the instrument will display “

bAd

” then ”

bAt

” for 1 second and then turns off.

To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.

• Remove the battery from its location and replace it with a new one, inserting the

negative end first.

• Insert the battery cover and replace the screw with a screwdriver.

Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.