background image

10

IX.   

IX.   

IX.   

IX.   

IX.   Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Connect sensors to pH/mV/ISE meter

Place meter in mV mode.

Place 100 mL of DIW into a beaker with stir bar.

Place electrodes into prepared sample.

Add 1 mL of a standard (0.1 M or 1000 ppm stan-
dard) to beaker. Record the mV value when stable.

Add an additional 10 mL of standard to the solution.
Record the mV when reading has stabilized.  This
value should be less than the previous noted (more
negative).

Determine the difference between the two mV values.
An acceptable value for this slope is -56 ± 4 mV.

X.   

X.   

X.   

X.   

X.   Corrective action

Corrective action

Corrective action

Corrective action

Corrective action

Verify protective cap has been removed (HI 4002).

Verify plastic film has been removed from inner stem
(HI 4102).

Verify electrodes are connected properly to meter and
meter is powered.

Verify dilute standards are freshly made and stored.
Remake solutions if appropriate.

If the sensor slope just misses the suggested slope
window, soaking the sensor in a dilute standard may
solve the problem. (

<

10

-3 

M bromide or <80 ppm

standard).

A scratched sensing surface can be polished with HI
4000-70 polishing strip.  Cut off approximately 1
inch of the micro-abrasive strip. Wet the frosted side
with deionized water 

 

and place against damaged

membrane of the electrode.  Place your thumb against
the shiny backing  and slowly rotate back and forth
while applying gentle pressure.  Continue polishing
until you are satisfied with the surface.   If dark depos-
its appear on polishing strip move the paper slightly
and continue polishing.

Summary of Contents for HI 4102

Page 1: ...1 HI 4002 HI 4002 HI 4002 HI 4002 HI 4002 HI 4102 HI 4102 HI 4102 HI 4102 HI 4102 Bromide Ion Selective Electrode Instruction Manual Half cell Combination BROMIDE HI 4002 HI 4102 BROMIDE COMBINATION...

Page 2: ...ions in aqueous solutions The HI 4002 is a solid state half cell sensor that requires a separate reference The HI 4102 is a combination ion selective electrode II II II II II Specifications Specificat...

Page 3: ...rolytic circuit The HI4102incorporates areferenceelectrode ThesilverBro mide pellet is practically insoluble in the test solutions being measured and produces a potential change due to changesinthesam...

Page 4: ...I 4102 Design elements of the HI 4002 and HI 4102 Design elements of the HI 4002 and HI 4102 electrodes electrodes electrodes electrodes electrodes BROMIDE HI 4002 Cap Sensing Membrane Sensor Handle U...

Page 5: ...romide Measurements for Bromide Measurements for Bromide Measurements for Bromide Measurements for Bromide Measurements 0 1 M Bromide Standard 500 mL HI4002 01 ISA 500 mL HI4000 00 ForMolarsolutions U...

Page 6: ...ample from magnetic stirrer by placing cork or other insulative sheet between beaker and stirrer plate Calibration standards and sample solutions should be stirred at the same rate using identical siz...

Page 7: ...e electrode with the fill hole open During normal use fill solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthis...

Page 8: ...m seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Page 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Page 10: ...rrective action Corrective action Verify protective cap has been removed HI 4002 Verify plastic film has been removed from inner stem HI4102 Verify electrodes are connected properly to meter and meter...

Page 11: ...the unknowns Morecalibrationstandardsarerequiredinnon linear regions Unknowns are read directly At very low levels of bromide special precautions must be employed for reproducible measurements Water...

Page 12: ...measurements 3 Follow sectionVII GeneralGuidelinestooptimizetest set up 4 Duringcalibrationitisbesttostartwithlowerconcen tration samples first Wait for a stable measurement before recording values Sl...

Page 13: ...cause a mV change of at least 30 mV This method is preprogrammed in the Hanna HI 4222 pH ISE mV meter which simplifies the method greatly Example Bromideiondeterminationwithknownaddition 1 A 50 mL sam...

Page 14: ...Titration Titration Titration Titration Titration XIV S XIV S XIV S XIV S XIV Storage and Care of the HI 4002 and torage and Care of the HI 4002 and torage and Care of the HI 4002 and torage and Care...

Page 15: ...ersion T Conversion T Conversion T Conversion Tables ables ables ables ables For Br For Br For Br For Br For Br Multiply by Multiply by Multiply by Multiply by Multiply by Moles L M to ppm mg L 7 99x1...

Page 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Reviews: