Hanna Instruments HI 4007 Instruction Manual Download Page 6

6

A user supplied oxidizing reagent may be used to supress
interferences such as hydroxide, sulfide, bromide, ammo-
nia or cyanide .  Add 10 mL to 100 mL standard or sample.
Permit sample to mix 10 minutes before measurements are
taken. Discard samples and     standards promptly as the
chloride in these will oxidize upon standing making them
unrelible. Do not permit  electrodes to remain in this solu-
tion for long periods of time.  Work in a hood and protect
eyes and skin. Prepare oxidizing reagent by diluting 6.25
mL analytical grade concentrated nitic acid into approxi-
mately 800 mL deionized water. Mix carefully. Dilute to 1
liter.

VII

VII

VII

VII

VII. General Guidelines

General Guidelines

General Guidelines

General Guidelines

General Guidelines

Calibration standards and sample solutions should
have the same ionic strength. ISA, acetate buffer  or
oxidizing reagent should be added to both samples
and standards in the same ratio.

Calibration standards and sample solutions should
be at same temperature.

Thermally insulate beaker with standard or sample
from magnetic stirrer.

Calibration standards and sample solutions should
be stirred at the same rate using identical sized TFE
coated stir bars.

Rinse electrode pair with distilled or deionized water
between samples and gently dry off  using soft dis-
posable absorbent toweling. Do not rub sensor sur-
face.

Presoaking chloride sensor in a dilute standard will
optimize response. Do not use concentration above
10

-3

M.

A scratched, pitted, or tarnished pellet surface can
cause drift, a loss of low level response, or poor repeat-
ability.  Optimum response can be restored by remov-
ing the damaged surface with the microabrasive strip
HI 4000-70.

Summary of Contents for HI 4007

Page 1: ...HI 4007 HI 4007 HI 4007 HI 4007 HI 4007 HI 4107 HI 4107 HI 4107 HI 4107 HI 4107 Chloride Ion Selective Electrode Instruction Manual CHLORIDE HI 4007 HI 410 COMBINATION 7 CHLORIDE Half cell Combinatio...

Page 2: ...utions The HI 4007 is a solid state half cell sensor that requires a separate reference The HI 4107 is a combination ion selective electrode II II II II II Specifications Specifications Specifications...

Page 3: ...asareferenceelectrode incorporated in its design The silver chloride pellet is prac tically insoluble in the test solutions being measured and producesapotentialchangeduetochangesinthesample s ion act...

Page 4: ...4107 Design elements of the HI 4007 and HI 4107 Design elements of the HI 4007 and HI 4107 electrodes electrodes electrodes electrodes electrodes CHLORIDE HI 4007 Cap Sensing Membrane Sensor Handle U...

Page 5: ...lutions Required Solutions Required for Chloride Measurements for Chloride Measurements for Chloride Measurements for Chloride Measurements for Chloride Measurements Select appropriate Hanna Instrumen...

Page 6: ...libration standards and sample solutions should have the same ionic strength ISA acetate buffer or oxidizing reagent should be added to both samples and standards in the same ratio Calibration standar...

Page 7: ...ing normal use fill solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the inter...

Page 8: ...m seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Page 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Page 10: ...action Corrective action Corrective action Corrective action Corrective action Verify protective cap has been removed HI 4007 Verify plastic film has been removed from inner stem HI4107 Verify electr...

Page 11: ...he meter is calibrated with two or more freshly made stan dards that are in the linear measurement range of the unknowns Morecalibrationstandardsarerequiredinnon linear regions Unknowns are read direc...

Page 12: ...ements using the oxidizing reagent must be made 10 minutes after mixing and not permitted react longer 3 Follow sectionVII GeneralGuidelinestooptimizetest set up 4 Duringcalibrationitisbesttostartwith...

Page 13: ...mplifies the method greatly Example Chloride ion determination with known addition 1 A 50 mL sample of unknown Vsample is placed in a clean plastic beaker with a chloride sensor Add 2 mLofacetatepHbuf...

Page 14: ...r See Section VI Titration Titration Titration Titration Titration XIV S XIV S XIV S XIV S XIV Storage and Care of the HI 4007 and torage and Care of the HI 4007 and torage and Care of the HI 4007 and...

Page 15: ...s Conversion tables Conversion tables Conversion tables Conversion tables For Cl For Cl For Cl For Cl For Cl Multiply by Multiply by Multiply by Multiply by Multiply by Moles L M to ppm mg L 3 500x104...

Page 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Reviews: