background image

23

22

Wrong slope

The calibration cannot be confirmed.

The evaluated slope is less than the lowest accepted value (80% of
default slope).

The evaluated slope is more than the highest accepted value (110 % of
default slope).

Wrong buffer

The calibration cannot be confirmed.

The pH reading is not close to the selected buffer. Select another buffer
using the 

ARROW 

keys or change the buffer.

Electrode Dirty/Broken

“ alternatively with “

Buffer Contaminated

“.

The calibration cannot be confirmed.

The offset of the electrode is not in the accepted range. Check if the
electrode is broken or clean it following the CLEANING PROCEDURE (see
page 74). Check the quality of the buffer. If necessary, change the
buffer.

FOUR, THREE or TWO-POINT CALIBRATION

• Proceed as described in “FIVE-POINT CALIBRATION” section.

• Press 

ESC 

after the appropriate accepted calibration point. The

instruments will return to calibration screen and will memorize the
calibration data.

SINGLE-POINT CALIBRATION

There are two selectable options for a single-point calibration:

Replace 

and 

Offset

.

This option is configured in SETUP mode under the parameter First Point
Mode. In both cases the meter will alter the present calibration data in
the instrument. If 

“Replace”

 is selected a new calibration point will be

added to the existing data, and the slope is calculated. The slopes
between current buffer and nearest lower and higher buffers will be
reevaluated.

If the 

“Offset”

 option is selected, an electrode offset correction is

performed to all buffer data keeping the existing slopes unchanged.

• Proceed as described in “FIVE-POINT CALIBRATION” section.
• Press 

ESC 

after the first calibration point was confirmed. The instruments

will memorize the single-point calibration data and will return to
calibration screen.

Notes:

• Press 

MTC 

key to toggle between pH buffer selection and the

temperature reading during calibration while temperature
probe is disconnected (

MTC

 mode).

• The displayed arrow is moving to the temperature value.

Use 

ARROW 

keys in order to change the temperature.

ERROR SCREENS
Wrong temperature

The calibration cannot be confirmed.

If the temperature is less than 0 °C or greater than 100 °C during
calibration, the calibration cannot be confirmed.

Summary of Contents for HI 3512

Page 1: ...1 Two Channels pH ORP ISE EC TDS NaCl Resistivity Temperature Bench Meter HI 3512 Instruction Manual w w w h a n n a i n s t c o m...

Page 2: ...th the accessories supplied Dear Customer Thank you for choosing a Hanna Instruments product Please read this instruction manual carefully before using this instrument Thismanualwillprovideyouwiththen...

Page 3: ...ange warning User selectable Calibration Timeout to remind when a new calibration is necessary Extended temperature range on EC channel from 20 to 120 C 4 to 248 F using temperature sensor inside EC p...

Page 4: ...m g n i d a e r f o 5 0 s n o i t n e l a v i d g n i d a e r f o 1 F 4 0 C 2 0 r o r r e e b o r p g n i d u l c x e e g n a r t e s f f o V m l e R V m 0 0 0 2 n o i t a r b i l a C H p n o i t a r...

Page 5: ...e r g r e v e h c i h w t i g i d 1 r o m p p 5 0 0 r o r r e e b o r p g n i d u l c x e y t i n i l a S e g n a R 0 0 0 4 o t 0 0 l C a N n o i t u l o s e R 1 0 y c a r u c c A r o r r e e b o r p...

Page 6: ...temperature value Note When in MTC mode the user can press and hold an ARROW key and the instrument will start incrementing decrementing the temperature value The instrument keeps measuring and the di...

Page 7: ...rrent temperature reading The ATC or MTC message is turned off because ppm readings are not temperature compensated In order to take accurate ISE measurements make sure that the appropriate ISE electr...

Page 8: ...leandonthe salinitycoefficient TheEC ResistivityandTDSscalesareauto ranging Themeterautomatically sets the scale with the highest possible resolution By pressing Lock the auto ranging feature is disab...

Page 9: ...e lists the temperature dependence of the HANNA calibration buffers C F 0 3 0 7 I H 0 3 0 8 I H m c S 1 3 0 7 I H 1 3 0 8 I H m c S 3 3 0 7 I H 3 3 0 8 I H m c S 4 3 0 7 I H 4 3 0 8 I H m c S 5 3 0 7...

Page 10: ...the value within a 1 00 pH window and then press Accept Press ESC to leave custom buffers value unchanged Press Confirm ForaccuratepHmeasurements itisrecommendedtoperformamultipoint calibration At lea...

Page 11: ...FIVE POINT CALIBRATION Immerse the pH electrode and the temperature probe approximately 3 cm 1 into a buffer solution of your choice pH 1 68 4 01 6 86 7 01 9 18 10 01 12 45 or a custom buffer andstir...

Page 12: ...for a single point calibration Replace and Offset This option is configured in SETUP mode under the parameter First Point Mode In both cases the meter will alter the present calibration data in the in...

Page 13: ...ustom buffer was set in SETUP menu it can be selected for calibration by pressing the ARROW keys The Custom function key will be displayed Press Custom if you want to adjust the buffer to its value at...

Page 14: ...2 1 4 5 8 6 9 9 6 3 9 8 5 7 9 1 2 1 1 5 7 7 6 1 6 7 1 4 1 4 6 8 6 0 0 7 1 9 8 4 7 9 0 1 1 1 0 8 6 7 1 7 7 1 6 1 4 7 8 6 1 0 7 9 8 8 4 7 9 0 0 1 1 5 8 5 8 1 8 7 1 7 1 4 7 8 6 2 0 7 7 8 8 4 7 9 1 9 0 1...

Page 15: ...10 100 1000 10000 It is advised to bracket expected ion concentration with your standards For fluoride electrode the 2 ppm standard is also available ForaccurateISEmeasurementsitisrecommendedtoperform...

Page 16: ...tag will blink on the LCD until the reading is stable When the reading is stable and close to the selected standard the Confirm function key is displayed Press Confirm to confirm calibration The instr...

Page 17: ...t least once a week After testing aggressive chemicals When calibration alarm time out is expired CAL DUE blinks if feature is enabled in SETUP If Out of cal range message appears on the LCD during EC...

Page 18: ...er few seconds the reading is stable If necessary press the ARROW keys to select a different standard value The tag will blink on the LCD until the reading is stable When the reading is stable and clo...

Page 19: ...lways shows the CAL DUE warning blinking When an abnormal condition in the RTC Real Time Clock is detected the instrument forces the expired calibration status LAST pH CALIBRATION DATA The last pH cal...

Page 20: ...not displayed press MENU key The instrument will display the ISE calibration information calibration date time slope calibration status and electrode type LAST EC CALIBRATION DATA The last pH calibrat...

Page 21: ...rmat DD MM YYYY YYYY MM DD MM DD YYYY YYYY MM DD YYYY MM DD MonDD YYYY DD Mon YYYY YYYY Mon DD Language Messagedisplay Uptofour English language languages Temperature Cor F C unit AutoEnd SelectAutoEn...

Page 22: ...odify Use ARROW keys to select date format then press Accept Press ESC to leave without changing GENERALPARAMETERSCREENS Backlight Focus on the Backlight item PressModify Use keys to change the intens...

Page 23: ...the calibration can be confirmed A long beep alert means that the pressed key is not active or a wrong condition is detected while in calibration AutoEnd stability Focus on AutoEnd Stability item Pres...

Page 24: ...item PressModify Use ARROW keys to select the desired communication baud Press Accept to confirm or ESC to exit Meter information Focus on the Meter Information item Press Select The meter information...

Page 25: ...probe item Press Custom in order to set the parameters for a custom probe PressStandardinordertoselectoneprobefromthestandardprobeslist If Custom is pressed UseARROWkeystofocusontheparametertobechange...

Page 26: ...will be displayed to alert that the ISE range must be calibrated If a new probe was selected or custom probe parameter are changed the ISE range must be calibrated Focus on Charge Slope item Use ARRO...

Page 27: ...automatically compensate conductivity using the set temperature coefficient with the temperature manually selected by the user Select ATC in order to automatically compensate conductivity using the se...

Page 28: ...e records in order to free log space LOTLOGGING Select the desired interval in SETUP as a timed interval or AutoEnd waits for a stable value before logging Press the StartLog function key in order to...

Page 29: ...he600recordsspaceor100lotsarereached the AutoLog space is full message will be displayed Enter View Log Data mode and delete lots in order to free space VIEW LOGGED DATA Press the Recall function key...

Page 30: ...ts are factory calibrated for mV EC and temperature Hanna s temperature probes are interchangeable and no temperature calibration is needed when they are replaced If the temperature or ORP measurement...

Page 31: ...xt scheme command prefix command CR where command prefix is the 16 ASCII character command is the command code Note Either small or capital letters can be used SIMPLE COMMANDS KF1 Is equivalent to pre...

Page 32: ...t represents always Primaryreading correspondingtotheselectedrange 12 ASCII chars including sign and decimal point and exponent CHRxx Change the instrument range according with the parameter value xx...

Page 33: ...on mode 2 chars 00 NoTC 01 MTC 02 ATC the temperature compensation coefficient with The unit of the second reading 1 char it is sent onlyforResistivity TDSandSalinityrangesonly It representstheunitoft...

Page 34: ...er 3 chars the number of records in lot 3 chars sign and decimal point 6 chars Salinity calibration data if available which contains thenumberofcalibratedstandards 1charalways1 the salinity factor wit...

Page 35: ...rd The number of the record 3 chars The reading status 1 char R O U The unit of the reading 2 chars it is sent only for ISE EC Resistivity TDS Salinity log only The calculated reading with sign decima...

Page 36: ...20 25 C 68 77 F Concentration pH Error 13 00 13 50 14 00 12 50 13 00 13 50 14 00 0 10 0 14 0 20 0 10 0 18 0 29 0 40 TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS Notes Err3 is sent if the Log on dem...

Page 37: ...or HI 8082 for double junction electrodes Follow the Preparation Procedure on page 72 before taking measurements Note NEVERSTORETHEELECTRODEINDISTILLEDORDEIONIZEDWATER PERIODIC MAINTENANCE Inspect th...

Page 38: ...rinse the electrode thoroughly with distilled water refill the reference chamber withfreshelectrolyte notnecessaryforgel filledelectrodes andsoakthe electrode in HI 70300 or HI 80300 Storage Solution...

Page 39: ...C l e v e l e t y l o r t c e l e w o L s e d o r t c e l e e l b a l l i f e r y l n o y l r e p o r p t o n e b o r p C E d e t c e n n o c e e s e d o r t c e l e e h t n a e l C e v o b a e t y l...

Page 40: ...mm DIA 0 12 HI 1083 120 mm 4 7 12 mm 0 5 S VERSION HI 1053 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 S VERSION HI 1043 HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x30 mL fordoublejunctionelectrode...

Page 41: ...S gel filled combinationgoldORPelectrode Use general purpose ConsulttheHannaGeneralCatalogforacompleteandwideselectionofelectrodes 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 FC 220 120 mm 4 7 12 mm 0 5 FC 21...

Page 42: ...to modify the design construction or appearance of its products without advance notice EXTENSIONCABLEFORSCREW TYPEELECTRODES SCREWTO BNC ADAPTER HI 7855 1 Extension cable 1 m 3 3 long HI 7855 3 Extens...

Page 43: ...na Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst com Lo...

Reviews: