
23
Otherwise:
• For
HI 2214
the instrument will display the
pH
or
relative mV
value
on the primary LCD and the last stored record number on the
secondary LCD, along with “
LOG
” tag.
• For
HI 2215
the instrument will display the lot number on the
primary LCD, the number of records on the secondary LCD, “
LOG
”
tag and “
CFM
” blinking. If samples were logged on demand “
MAn
”
will be displayed on the primary LCD and the number of samples
logged on the secondary LCD (see example below: manual log,
15 samples logged).
Press the
ARROW
keys to select a different lot.
All the existing lots are displayed.
Press
CFM
to view record information.
The instrument will display the
pH
or
relative mV
value on the primary
LCD and the last stored record number on the secondary LCD, along with
“
LOG
” tag.
or
Use the
ARROW
keys to scroll through the records.
To view additional information, press
RANGE
:
• The
mV
value is displayed on the primary LCD and the temperature
value on the secondary LCD.
26
SETUP
Setup mode allows viewing and modifying the following parameters:
• Expired Calibration Alarm
• Log Interval (
HI 2215
only)
• First Custom Buffer
• Second Custom Buffer
• One-point Calibration Behavior
• pH Resolution
• Current Time (hour & minute)
• Current Date (year, month & day)
• Beep Status
• Instrument Id
• Temperature Unit
To enter SETUP mode, press
SETUP
while the instrument is in
measurement mode.
Select a parameter with the
ARROW
keys.
Press
CAL
to change a parameter value. The selected parameter
will
start blinking.
Press
RANGE
to toggle between displayed parameters.
Press the
ARROW
keys to increase or decrease the displayed value.
Press
CFM
to save the modified value or
CAL
to escape without saving.
LOG INTERVAL
(HI 2215 only)
Press
CAL
when log interval is displayed. The log interval and “
CFM
” will
start blinking (“
MAn
” for log on demand, “
StAb
” log on stability,
interval in seconds or minutes).
Press the
ARROW
keys to change the custom buffer value.
Press
CFM
to confirm the selection.
Press
CAL
to escape without saving
EXPIRED CALIBRATION ALARM
Press
CAL
when the calibration time-out is displayed. Calibration time-out
(“
OFF
” or “
1
” to “
7
” days) and “
CFM
” will start blinking.
Press the
ARROW
keys to change the calibration time-out value.
Press
CFM
to save the modified calibration time-out value.
Press
CAL
to escape without saving.