14
CFM
• Submerse the temperature probe in the second vessel as near as
possible to the
reference thermometer. Allow a few seconds for the
probe to stabilize.
• Use the
ARROW
keys to set the reading on the secondary LCD to
that of the hot water.
• When the reading is stable and close to the selected calibration
point, “
READY
” tag will appear and “
CFM
” tag will blink.
• Press
CFM
to confirm. The instrument returns to measurement mode.
Note
: If the reading is not close to the selected calibration point,
“
WRONG
” tag will blink. Change the temperature probe and
restart calibration.