Hanna Instruments HI 146-00 Instruction Manual Download Page 2

SPECIFICATIONS

HI 146-00

Range

-50.0 to 150.0°C

Resolution

 0.1°C

Accuracy

±0.3°C (-20 to 90°C), ±0.5°C (outside)

Typical EMC Deviation

±0.3°C

HI 146-01

Range

-58.0 to 302°F

Resolution

0.1°F (-58.0 to 199.9°F), 1°F(200 to 302°F)

Accuracy

±0.5°F (-4 to 194°F), ±1°F (outside)

Typical EMC Deviation

±0.5°F

Common Specifications

Probe

 Stainless steel probe with 2 m (6.6’) cable

Battery Type

1 x 1.5V AA alkaline

Battery Life

approx. 5 years

Casing

IP54

Environment

0 to 50°C (32 to 122°F); RH max 95%

Dimensions

86 x 110 x 43 mm (3.4 x 4.3 x 1.7’’)

Weight

150 g (5.3 oz.)

OPERATION

• Dip the probe in the sample.

An immersion level of 4 cm (1 ½") is
recommended.

• Do NOT immerse the probe handle.
• Do NOT wrench the probe or put heavy

objects on the cable. If the cable is
damaged, the meter

 

needs to be re-

placed.

• Wait for a few seconds for the display

to stabilize.

• Variations in readings or fading dis-

play can indicate a weak battery.

CALIBRATION CHECK SWITCH

Remove the back cover and move the internal switch to the
“TEST” position.
The display should show 0.0±0.3°C (32.0°F±0.5°F).

OPERATIONAL GUIDE

Contact your dealer for

 

recalibration, if necessary.

To return to the measurement mode, move back the switch

 

to “MEAS” position. Replace the back cover and tighten the

 

screws.

BATTERY REPLACEMENT

When the display fades, remove the back cover and replace
the 1.5V AA battery with a new one, while paying atten-
tion to its polarity. Close the meter.
Battery should only be replaced in a safe area using the
battery type specified in this instruction manual.

WARRANTY

HI146 is warranted for two years 

against defects in workmanship and materials when used

for its intended purpose and maintained according to instructions.

 The probe is warranted

for a period of six months.

This warranty is limited to repair or replacement free of charge. Damages due to accident,
misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact the dealer from whom you purchased the instrument. If under
warranty, report the model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service department and then send it with
shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for
complete protection.

I

ST146R2  05/05

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 