1. PRELIMINARY EXAMINATION
Remove the tester and accessories from packaging and examine them carefully. For further assistance, please contact your
local Hanna Instruments office or email us at [email protected].
Each
HI9810382
is supplied with a starter kit consisting of:
• pH 4.01 buffer solution, 20 mL sachet (2 pcs.)
• pH 7.01 buffer solution, 20 mL sachet (2 pcs.)
• Electrode cleaning and disinfection solution for bread and dough deposits, 20 mL sachet (2 pcs.)
• Electrode storage solution, 13 mL dropper bottle (1 pc.)
• 3V Lithium battery – CR2032
• Instrument quality certificate and Instruction manual
Note:
Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be
returned in its original packing material with the supplied accessories.
2. SPECIFICATIONS
Range
pH
0.00 to 12.00 pH
mV *
pH/mV conversion
Temperature** 0.0 to 60.0 °C (32.0 to 140.0 °F)
Resolution
pH
0.01 or 0.1 pH
mV *
0.1 or 1 mV
Temperature
0.1 °C; 0.1 °F
Accuracy
pH
±0.05 pH
Temperature
±0.5 °C; ±0.9 °F
Calibration
Up to three points or four points *
Automatic buffer recognition with Standard buffers
Hanna (pH 1.68 *, 4.01, 7.01, 10.01) or NIST (pH 1.68 *, 4.01, 6.86, 9.18)
Temperature compensation Automatic (ATC) or Manual (MTC) *
Electrode
Body material
Polyvinylidene Fluoride (PVDF)
Glass
Low Temperature (LT)
Junction Open
Reference cell
Double, Ag/AgCl
Electrolyte Viscolene
Tip / Shape
Conic, Ø 6 x 10 mm (Ø 0.23 x 0.39”)
Outer diameter
13 mm (0.51“)
Length
75 mm (2.95”)
Battery type
3V Lithium – CR2032
Battery life
Approximately 1000 hours (500 hours with Bluetooth enabled)
Environment
0 to 50 °C (32 to 122 °F)
Casing
IP65 ingress protection
Dimensions / Weight
51 x 146 x 21 mm (2.0 x 5.7 x 0.8”) / 45 g (1.6 oz.)
* Available with Hanna Lab App
** Measuring outside the recommended operating temperature range may damage the gel electrolyte and void product warranty.
Note:
The tester can display measurements from
–
2.00 to 16.00 pH. Measurements outside of the pH range will flash. In this
case assess the integrity of the tester and the type of measured sample.