Operational Guide

The meter is shipped with the battery inside.

To turn the meter ON and check the battery 
status

Press the ON/OFF button to turn the meter on. 
At  start-up, all the LCD segments are displayed 
for 1 second, “Err CAL“ is displayed when no 
calibration is available, then the percent indication 
of the remaining battery life is displayed for 
another second. The meter then enters the normal 
measuring mode.

Note: Keeping the ON button pressed while 

turning the meter on will display all LCD 

segments as long as the button is pressed, after 

10 seconds the firmware version and model will 

be displayed. Pressing the button 3 times will 

display the serial number (SN--1, SN--2, SN--3). 

To enter measurement mode press the ON/OFF 

button briefly.

 

Thank You

Thank you for choosing a Hanna Instruments 
product. Please read this instruction manual 
carefully before using the instrument. 
For more information about Hanna Instruments 
and our products, visit www.hannainst.com or 
e-mail us at [email protected].
For technical support, contact your local Hanna  
Instruments Office or email us at 

[email protected].
Find your local Hanna Instruments Office on 

www.hannainst.com.

Preliminary Examination

Remove the meter from the packing material and 
examine it carefully to make sure that no damage 
has occurred during shipment. If noticeable damage 
is evident, contact your local Hanna Instruments 
Office.
Each meter is supplied with:

• CR2032 battery (1 pc.)
• Storage / Protection sleeve
• Instruction manual
• Quality Certificate

Note: Save all packing material until you are 

sure that the instrument functions correctly. All 

defective items must be returned in the original 

packaging with the supplied accessories.

General Description

The Hanna Instruments GroLine is a special pocket 
conductivity meter expressly designed to directly 
measure soil conductivity. The meter also measures 
temperature.
Soil conductivity is a simple measurement that 
indicates if salts are present in the sample soil. 
Conductivity can be measured in units of μS/cm, 
mS/cm, or dS/m.
The conductivity is calibrated at one point, in liquid 
conductivity standard.

Hanna Instruments reserves the right to modify the 

design, construction, or appearance of its products 

without advance notice.

Specifications

Range

0 to 4000 μS/cm 
0.00 to 4.00 mS/cm (dS/m) 
0.0 to 50.0 °C (32.0 to 122.0 

°

F)

Resolution

1 μS/cm 
0.01 mS/cm (dS/m) 
0.1 °C (0.1 

°

F)

Accuracy 
(@25 °C/77 °F)

±50 μS/cm (0 to 2000 μS/cm)
±300 μS/cm (2000 to 4000 μS/cm)
±0.05 mS/cm (0.00 to 2.00 mS/cm) 
±0.30 mS/cm (2.00 to 4.00 mS/cm) 
±1.0 °C (±1.5 

°

F)

Temperature 
Compensation

automatic, with temperature coefficient 

β

 

) fixed @ 2%/°C

Calibration

single point, 1.41 mS/cm

Battery Type

CR2032 Li-ion

Battery Life

approximately 100 hours 
of continuous use

Probe

114 mm (4.5“) Stainless 
steel penetration

Environment

0 to 50 °C (32 to 122 °F); 
RH 95% max

Dimensions

50 x 196 x 21 mm 
(2.0 x 7.7 x 0.9“) 

Weight

74 g (2.4 oz.)

All rights are reserved. Reproduction in whole or 

in part is prohibited without the written consent of 

the copyright owner.

Liquid 
Crystal 
Display

Maximum 
Immersion 
Level

Recommended 
Immersion 
Level

ON/OFF
button

Operation

Advanced Functions

  

HOLD mode

The HOLD mode can be selected only if in setup 
the HOLD parameter is set to “YES”. While in 
measurement mode, press the ON/OFF button until 
briefly “HOLD“ appears on the secondary display 
and the reading will be frozen on the LCD. The HOLD 
will toggle with the temperature value until the 
ON/OFF button is pressed briefly again. The 
“MEAS” message appears and the meter returns to 
measure mode.

Hold on STABILITY mode

The STAB mode can be selected only if in setup the 
HOLD parameter is set to “StAb”.
By pressing the ON/OFF button briefly and the 
“WAIT” message appears on screen. The parameters 
are hold on screen after stabilization only; before 
stabilization done the stability indicator is present 
on the LCD.
Press  the ON/OFF button briefly to exit the STAB 
mode and return to measure mode. The “MEAS” 
message appears and the meter returns to measure 
mode.
Taking another stable measurement is possible by 
pressing the ON/OFF button briefly.
(If the “HOLD” message appears, then measured 
stable values (EC and temp) are already retained 
and hold on the display.) 

Before using Hanna Instruments products, make 
sure that they are entirely suitable for your specific 
application and for the environment in which 
they are used. Avoid touching the probe at all 
times. Any variation introduced by the user to the 
supplied equipment may degrade the instrument’s  
performance. Do not use or store the instrument in 
hazardous environments.

Recommendations for Users

To enter calibration mode

Press and hold down the ON/OFF button until “OFF“ 
is replaced by “CAL.“ Release the button.

To enter setup mode

While in measurement mode, remove the battery 
cap and hold the Setup button for 1 second. The 
meter will enter in setup mode.

Meter Setup

 

While in measurement mode, remove the battery 
cover. Press Setup button located on the side of the 
battery in the battery compartment. The meter will 
enter in setup mode. 
Press and hold the ON/OFF button to move through 
setup parameters. Change option by pressing 
ON/OFF button briefly. The default settings are: 
“EC UNIT” - mS/cm, “TEMP” measure unit - °C, 
“AOFF” - 8 min, “HOLD” functionality - disabled 
“no“. After Hold, the meter returns to measurement 

mode.

Summary of Contents for GroLine HI98331

Page 1: ...INSTRUCTION MANUAL HI98331 Direct Soil Conductivity Temperature Meter with built in Stainless Steel Conductivity Probe ...

Page 2: ...fice Each meter is supplied with CR2032 battery 1 pc Storage Protection sleeve Instruction manual Quality Certificate Note Save all packing material until you are sure that the instrument functions correctly All defective items must be returned in the original packaging with the supplied accessories General Description The Hanna Instruments GroLine is a special pocket conductivity meter expressly ...

Page 3: ... button Operation Before using Hanna Instruments products make sure that they are entirely suitable for your specific application and for the environment in which they are used Avoid touching the probe at all times Any variation introduced by the user to the supplied equipment may degrade the instrument s performance Do not use or store the instrument in hazardous environments Recommendations for ...

Page 4: ...model will be displayed Pressing the button 3 times will display the serial number SN 1 SN 2 SN 3 To enter measurement mode press the ON OFF button briefly Specifications Range 0 to 4000 µS cm 0 00 to 4 00 mS cm dS m 0 0 to 50 0 C 32 0 to 122 0 F Resolution 1 µS cm 0 01 mS cm dS m 0 1 C 0 1 F Accuracy 25 C 77 F 50 µS cm 0 to 2000 µS cm 300µS cm 2000to4000µS cm 0 05 mS cm 0 00 to 2 00 mS cm 0 30 mS...

Page 5: ...return to measure mode The MEAS message appears and the meter returns to measure mode Taking another stable measurement is possible by pressing the ON OFF button briefly If the HOLD message appears then measured stable values EC and temp are already retained and hold on the display To enter calibration mode Press and hold down the ON OFF button until OFF is replaced by CAL Release the button To en...

Page 6: ...ton for 1 second to exit the menu or press and hold ON OFF button for 2 seconds after the last parameter was selected For better accuracy frequent calibration of the meter is recommended In addition the instrument must be recalibrated whenever When high accuracy is required At least once a month Measurement PresstheON OFFbuttontoturnthemeterON Place the probe to the desired depth and saturate the ...

Page 7: ...d and meter returns to measurement mode Clear calibration Place meter in calibration mode Press and hold ON OFF until Clr is displayed The meter will now be at default calibration After restarting the meter the Err CAL message will appear for 2 seconds until a new calibration is performed Error Message If the measured temperature is higher than 50 0 ºC or lower than 0 ºC the 50 0 ºC or 0 0 ºC temp...

Page 8: ...partment and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packaged for complete protection HI7031 023 1413 μS cm 1 41 mS cm GroLine Conductivity Standard Solution 230 mL bottle HI7031M 1413 μS cm 1 41 mS cm calibration solution 230 mL HI7031L 1413 μS cm 1 41 mS cm calibration solution 500 mL HI70031P 1413 μS cm 1 41 mS cm calibration solution 20 mL...

Reviews: