• If using pH 7.01, after acceptance of the buffer press CAL button to exit. “Stor”
message is displayed and meter returns to pH measurement mode.
• If using pH 4.01 or 10.01 buffer the “Stor“ message is displayed and meter returns
to pH measurement mode.
• The “CAL“ tag will be shown in measurement mode with the calibrated buffer
tag next to it.
Thank You
Thank you for choosing a Hanna Instruments product. Please read this
instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products, visit
www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments Office or
e-mail us at [email protected]
Find your local Hanna Instruments Office on www.hannainst.com
Preliminary Examination
Remove the meter from the packing material and examine it carefully to make sure that
no damage has occurred during shipment. If noticeable damage is evident, contact your
local Hanna Instruments Office.
Each meter is supplied with:
• CR2032 battery (1 pc.)
• Storage / Protection sleeve
• Instruction manual
• Quality Certificate
Note: Save all packing material until you are sure that the instrument functions
correctly. All defective items must be returned in the original packaging with the
supplied accessories.
Operation
pH electrode
Liquid Crystal
Display
CAL button
ON/OFF
button
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
Operational guide
To turn the meter ON
Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are dis-
played for 1 second, then the percent indication of the remaining battery life is displayed
for another second. The meter then enters the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on will display all
LCD segments as long as the button is pressed.
To enter calibration mode
Press the CAL button. “CAL“ message is displayed.
To enter setup mode
While in measurement mode, remove the battery cover and press the button inside the
battery compartment.
Range
0.00 to 14.00 pH
0.0 to 50.0 °C (32.0 to 122.0 °F)
Resolution
0.01 pH / 0.1 °C / 0.1 °F
Accuracy
(@25 °C/77 °F)
±0.10 pH / ±0.5 °C / ±1.0 °F
Calibration
automatic, one or two-points
(pH 4.01, 7.01, 10.01)
Quick Calibration
one point calibration using Quick Cal pH and EC
one-point calibration solution
HI50036
Temperature
Compensation
automatic, 0 to 50 °C
Battery Type
CR2032 3V Li-ion (1 pc.)
Battery Life
approximately 800 hours of continuous use
Environment
0 to 50 °C (32 to 122 °F);
RH 100% max
Dimensions
160 x 40 x 17 mm (6.3 x 1.6 x 0.7“)
Weight
(without battery)
65 g (2.3 oz.)
Specifications
pH Measurement and Calibration
• Make sure the meter has been calibrated before use.
• If the electrode is dry, soak it in
HI70300
storage solution for 30 minutes to reactivate
it.
• Submerge the electrode in the sample to be tested while stirring it gently. Wait until
the Stability Indicator on the LCD disappears.
• The LCD displays the pH value (automatically compensated for temperature) on the
primary LCD, while the secondary LCD displays the sample temperature.
• If measurements are taken in different samples successively, rinse the electrode tip
thoroughly to eliminate cross-contamination. After cleaning, rinse the electrode tip
with some deionized water and some of the sample to be measured.
pH calibration
• Enter calibration mode while in pH measurement mode.
• Place the electrode into the first calibration buffer. If performing a two-point
calibration, use pH 7.01 buffer first.
• The meter will enter the calibration mode, displaying “pH 7.01 USE”.
Single-point calibration
• Place the electrode in buffer pH 7.01, 4.01 or 10.01 buffer. The meter will
automatically recognize the buffer value.
• If the buffer is recognized “REC“ is displayed until the reading is stable and the
calibration is accepted.
• If the buffer is not recognized, the pH electrode is not immersed in solution, or the
reading is out of the accepted range “---- Err“ is displayed.
Preparation
The pH electrode is shipped dry. Before using the pH electrode, remove the protective
cap and condition the electrode by soaking the tip (bottom 3 cm (1.18”)) in
HI70300
storage solution or in pH 7.01 buffer solution for several hours. Then follow the
calibration procedure:
• Do not be alarmed if white crystals appear around the cap. This is normal with pH
electrodes and they dissolve when rinsed with water.
• Turn the pH electrode on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the electrode in the sample
to be tested.
• Stir gently and wait for the stability tag to disappear.
• For best results, recalibrate periodically.
• After use, rinse the electrode with water and store it with a few drops of HI70300
storage solution or pH 4.01 or pH 7.01 buffer solution in the protective cap.
• Always put back the protective cap after each use.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
• Large differences in pH readings (±1.0 pH) could result from lack of calibration
or dry electrode.
Note: The electrode tip should be rinsed with purified water (reverse osmosis, distilled,
or deionized) before and after placing in any solution (buffer, storage or sample).
• Electrode cleaning solution sachet (1 pc.)
• Quick Cal pH and EC One-point calibration
solution sachet (3 pcs.)
The default settings are: “Set t” measure unit - °C, “AOFF” - 8 min and “CAL” standard.
After the last setting, it will exit Setup.
To select the temperature unit (°C/°F)
To select the measurement unit when “Set t” is displayed press the CAL button to
change between °C or °F.
To select the Auto-Off time
To select AOFF TIME unit when “AOFF” is displayed press the CAL button to change
between 8 min, 60 min or --- (disabled).
To select calibration type
To select calibration type when “CAL” is displayed press CAL button to change between
Standard and Quick calibration.
To return to measurement mode
Press ON/OFF button to exit the menu.
Setup
Button
Meter Setup
While in measurement mode, remove the battery cover. Press the Setup button located
on the side of the battery in the battery compartment. The meter will enter in setup
mode. Press the ON/OFF button to move through setup parameters. Press the CAL
button to change the settings.
Quick Calibration
• Enter calibration mode while in measurement mode.
• Place the electrode in calibration solution
HI50036
.
• The “CAL quic” message is displayed.
• If the buffer is recognised, the meter will wait for stable measurement and when
the calibration is completed will display “Stor“ and exit calibration.
• If the buffer is not recognised “----Err“ is displayed. Check the buffer, clean the
electrode or press CAL button to exit calibration.
• After entering in the calibration mode and before the point is accepted, it is
possible to quit the procedure and return to the last calibration data by pressing
the CAL button. The LCD displays “----ESC” for 1 second and the meter returns to
measurement mode.
Before using Hanna Instruments products, make sure that they are entirely suitable
for your specific application and for the environment in which they are used. Avoid
touching the electrode area. Any variation introduced by the user to the supplied
equipment may degrade the instrument’s performance. Do not use or store the
instrument in hazardous environments.
Recommendations for Users