Hanna Instruments Gro Line HI98318 Instruction Manual Download Page 2

2. Place the probe in calibration solution. The probe tip should be centered in the 

solution and submersed 1.18” (3 cm). The meter will automatically recognize the 

solution.

3.  If  the  solution  is  not  recognized  or  is  out  of  the  accepted  range  “

---- 

Err“  is 

displayed.

4. If the buffer is recognized “REC“ is displayed until the reading is stable and the 

calibration is accepted.

• After  acceptance,  the  “Stor“  message  is  displayed  and  meter  returns  to 

measurement mode.

Clear Calibration

Place meter in calibration mode. Press ON/OFF and “Clr” is displayed. The meter will 

now be at default calibration. 

Quick Calibration

Select calibration type “CAL quic “ from meter setup.

• Enter calibration mode.
• Immerse the probe in the 

HI50036

 calibration solution.

• When  the  standard  values  is  recognized  and  stability  is  reached  the  meter 

automatically accepts the calibration.

• The LCD will display “Stor“ for 1 second and return to normal measurement mode.
• If the standard is not recognized or the slope is out of accepted range “----Err‘ is 

displayed. Change the calibration solution, clean the electrode or press any key 

to exit calibration.

Error message

During user calibration, if the reading is out of the accepted range, the meter will 

display “---- Err“.i.e. 3000 ppm blinking means the limit of the measured value is 

out of range.
If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the 50.0 ºC or 

0.0 ºC temperature value will blink on the LCD in measure mode.

Operational Guide

To turn the meter ON and check the battery status

Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are 

displayed for 1 second, then the percent indication of the remaining battery life is 

displayed for another second. The meter then enters the normal measuring mode.

Note:  Keeping  the  ON/OFF  button  pressed  while  turning  the  meter  on  will 
display all LCD segments as long as the button is pressed.

To enter calibration mode

Press the CAL button. “CAL“ message is displayed.

To enter setup mode

Remove the battery cover and press the Setup button located on the side of the battery.

Meter Setup

While in measurement mode, remove the battery cover. Press the Setup button located 

on the side of the battery in the battery compartment. The meter will enter in setup 

mode. Press the ON/OFF button to move through setup parameters. Change option by 

pressing CAL button. The default settings are: Set t measure unit - °C, AOFF - 8 min, 

CAL - standard and EC unit.

To select the temperature unit (°C/°F)

To select the temperature unit when “SET t” is displayed press the CAL button to 

change between °C or °F.

To select the Auto-Off time

To select the AUTO OFF when “AOFF” is displayed press the CAL button to change 

between 8 min, 60 min or --- (disabled).

To select calibration type

To select calibration type when “CAL“is displayed press the CAL button to choose from 

“CAL STD“ (standard calibration) or “CAL quick “(one-point quick calibration).

To select EC or TDS

To select TDS measurement with 0.5 TDS factor press CAL button until “ppm 500 UNIT“ is 

displayed. To select TDS measurement with 0.7 TDS factor press CAL button until “ppm 

700 UNIT“ is displayed. By pressing CAL button again “mS/cm EC UNIT“ is selected.

To return to measurement mode

Press the ON/OFF button.

To obtain the highest accuracy for measurements it is important to follow these tips:

•  Calibration is only as good as the solution being used. The calibration solution 

values change over time. Fresh solution should be used for each calibration.

•  The probe should be rinsed with purified water each time before placing in 

calibration solution or sample to be tested.

Care and Maintenance

Battery Replacement

Setup 
Button

Warranty

These meters are warranted for a period of one year against defects in workmanship 

and  materials  when  used  for  their  intended  purpose  and  maintained  according  to 

instructions. Probes are warranted for a period of six months. This warranty is limited 

to repair or replacement free of charge. Damage due to accidents, misuse, tampering or 

lack of prescribed maintenance is not covered. If service is required, contact your local 

Hanna office. If under warranty, report the model number, date of purchase, serial 

number and the nature of the problem. If the repair is not covered by the warranty, 

you will be notified of the charges incurred. If the instrument is to be returned to 

Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from 

the Technical Service department and then send it with shipping costs prepaid. When 

shipping any instrument, make sure it is properly packaged for complete protection.

Before  using  Hanna  products,  make  sure  that  they  are  entirely  suitable  for  your 

specific application and for the environment in which they are used. Operation of these 

instruments  may  cause  unacceptable  interferences  to  other  electronic  equipment, 

thus requiring the operator to take all necessary steps to correct such interferences. 

Any  variation  introduced  by  the  user  to  the  supplied  equipment  may  degrade  the 

instrument’s EMC performance. To avoid damages or burns, do not put the instrument 

in  microwave  oven.  For  yours  and  the  instrument  safety  do  not  use  or  store  the 

instrument in hazardous environments.

Recommendations for Users

Solutions

Code

Description

HI5036-023

Quick Cal pH and EC one-point 

calibration solution,  

230 mL bottle

HI50036P

Quick Cal pH and EC one-point

 calibration solution,  

20 mL sachets (25 pcs.)

HI70031P

1413 µS/cm calibration solution,

 

20 mL sachets (25 pcs.)

HI7031L

1413 µS/cm calibration solution,  

500 mL bottle

HI8031L

1413 µS/cm calibration solution, 

500 mL bottle

Accessories

 

 IST98318   07/16-1

Measurement and Calibration

For better accuracy, frequent calibration of the meter is recommended. In addition, 

the instrument must be recalibrated:

• When high accuracy is required.
• At least once a month.

Measurement

Press the ON/OFF button to turn the meter ON. Place the probe to the desired solution. 

The conductivity value, automatically compensated for temperature, will be shown 

on the LCD.

The measured temperature will be shown on the secondary LCD.

Note: Before taking any measurement make sure the meter has been calibrated.

• Press CAL button to enter calibration mode from measurement mode. 
• The meter will enter the calibration mode, displaying “mS/cm 1.41 USE” with 

CAL tag blinking.

1. Pour 2” (5 cm) of standard solution into a container.

• When the calibration procedure is completed, the “CAL“ tag is turned on.
• After  entering  in  the  calibration  mode  and  before  the  point  is  accepted,  it  is 

possible to quit the procedure and return to the last calibration data by pressing 

the CAL button. The LCD displays”----ESC” for 1 second and the meter returns to 

measurement mode.

To change the CR2032 Li-ion battery, turn the battery cover located on the back of 

the meter counterclockwise to unlock. Remove cover and replace with new battery 

+ side facing up.

Note:  Batteries  should  only  be  replaced  in  a  safe  area  using  the  battery 
type specified in this instruction manual. Old batteries should be disposed in 
accordance with local regulations.

Calibration Procedure

The meter features a low battery indicator. When the battery is running low (under 

10%), the battery indicator will blink on the LCD. When the battery is discharged  

“dEAd bAtt”

 will be displayed on the LCD for 2 seconds and the meter will turn off.