Hanna Instruments Gro Line HI981030 Instruction Manual Download Page 2

Specifications

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy 

±0.2 pH @25 °C/77 °F

Calibration

Automatic, one or two-point

Electrode

Built-in probe for specific application

Battery Type

CR2032 Li-ion

Battery Life

Approximately 1000 hours of continuous use

Auto-off

8 minutes, 60 minutes or can be disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

51 x 151 x 21 mm (2 x 5.9 x 0.9“) 

Weight

44 g (1.6 oz.)

Preliminary Examination

Remove the meter from the packing material and examine it carefully to make sure that 
no damage has occurred during shipment. If noticeable damage is evident, contact your 
local Hanna Instruments Office.
Each meter is supplied with:

• pH 4.01 buffer solution liquid sachet (2 pcs.)
• pH 7.01 buffer solution liquid sachet (2 pcs.)
• Cleaning solution for soil deposits
• Cleaning solution for humus deposits
• Electrode storage solution, 13 mL dropper
• Electrode fill solution
• Instruction manual
• Quality certificate

Note: Save all packing material until you are sure that the instrument functions 

correctly. All defective items must be returned in the original packaging with the 

supplied accessories.

Liquid Crystal 
Display

ON/OFF 
Button

Maximum 
Immersion 
Level

Recommended 
Immersion Level

Procedure for Direct Ground Measurement

1)  Verify meter is set up correctly and pH probe is calibrated.
2)  Dig, discarding 5 cm of topsoil.
3)   Perforate the soil (with 

HI721319

 soil auger) to a depth of about 20 cm or more.

4)  If the soil is dry, moisten it with a small amount of tap water.
5)  Wash the electrode with tap water (not distilled).
6)  Insert the electrode pushing it slightly into the soil to ensure proper contact.
7)  Observe the measurement.
8)  Wash the electrode with tap water (not distilled) and (using a finger) gently remove 

any soil remaining on the electrode (avoid using a rag or cloth).

9)  Repeat the procedure in different locations in the field.
10) Consider the average of the measured data.
For best result, it is advisable to measure the pH of a soil solution, using a sample of 
soil and soil preparation solution 

HI7051

; it is better to use this procedure if you have 

to test a stony field in which you risk damaging the electrode.

Probe Features

Removable PVDF reference sleeve

Polyvinylidene fluoride (PVDF) is a durable plastic that is resistant to most chemicals 
and solvents, including sodium hypochlorite. It has high abrasion resistance, mechanical 
strength and resistance to ultraviolet. PVDF is also resistant to fungal growth.

Operation

Press the ON/OFF button to turn the meter on. All tags 
will be displayed.

The meter will go into measurement mode: current 
reading and calibrated buffers are displayed.

Meter Calibration

From measurement 
mode, press and 
hold the ON/OFF 
button until “CAL” 
is displayed.

When “7.01” blinks 
on the display, place 
the tip of the probe 
into a pH 4.01, 7.01 
buffer solution.

When the 

reading is stable, 

the stability 

indicator will 

disappear.“Sto” 

will be displayed 

when the 

calibration 

is saved.

If pH 4.01 buffer 
solution is used 
as the first point

 

the value of the 
buffer is recognized 
and displayed 
with the blinking 
stability indicator.

Meter will 
exit to 
measurement 
mode and the 
calibration 
tag will be 
displayed.

One-Point Calibration with pH 4.01

B

On

e-

Po

in

t

Meter will exit 
to measurement 
mode and the 
calibration 
tag will be 
displayed.

pH 4.01 will then 
blink on the display. 
Ignore it and press 
ON/OFF button.

Tw

o-

Po

in

t

Use pH 4.01 to 
perform a two-
point calibration. 
The value is 
automatically 
recognized and 
displayed with the 
blinking stability 
indicator.

When the reading 
is stable, the 
stability indicator 
will disappear. 
“Sto” will be 
displayed when 
the calibration 
is saved.

Meter will exit 
to measurement 
mode and the 
calibration 
tags will be 
displayed

.

If pH 7.01 buffer solution is used as 
the first point

 the buffer is recognized 

with the blinking stability indicator.
When the reading is stable, the 
stability indicator will disappear 
and pH 7.01 will be calibrated. 

If pH 7.01 is the only calibration point,

 

finish one-point procedure at right. 

If using pH 4.01 as a second point,

 

continue two-point procedure at right.

Meter Overview

Preparation:

The pH electrode is shipped ready for use with a protective cap containing storage solution. Before 

using the meter, 

remove the protective cap 

and inspect the probe to verify the bridge gel has 

not dried out. Refresh the bridge electrolyte if necessary. Following care procedure (ignoring the 

cleaning procedures). 
Alternately, rinse off the probe tip with water and condition the electrode by soaking the tip (bottom 

4 cm (1.5”)) in pH 7.01 buffer solution for several minutes. Perform a probe calibration before using.

•   Do not be alarmed if white crystals appear around the cap. This is normal with pH 

electrodes. Rinse off the probe tip with water before using.

•  Turn the meter on by pressing ON/OFF button.
•  Remove the protective cap. DO NOT SCRATCH THE GLASS. Do not use the probe to bore 

a hole through the soil. Make sure an entry hole is made in the soil and water has been 

added. After placing the probe in the hole, wait for a stable reading.

NEVER IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION LEVEL.

•  For best results, recalibrate periodically.
•  After use, rinse the electrode with water and store it with a few drops of storage solution 

in the protective cap.

•  Reattach the protective cap after each use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.

A

One or Two-Point Calibration with pH 7.01

For one-point calibration using pH 4.01 buffer go to procedure B 

For one or two-point calibration using pH 7.01 buffer go to procedure A

A

B

Open Reference Junction with  
removable sleeve design

Suspended solids can permanently clog the pores 
of a ceramic reference junction. The open junction 
design provides a constant junction potential and 
minimizes blockage by providing an open gel 
interface between the sample and internal Ag/AgCl 
reference. Should soil enter into the junction, the junction can be easily cleaned and 
refreshed with new bridge electrolyte.

pH Probe

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using this instrument. 
For technical support, contact your local Hanna Instruments Office or email 
us at [email protected].
To find your local Hanna Instruments Office or for additional information on Hanna 
Instruments products, visit www.hannainst.com

Battery 

indicator

Unstable 

tag

Measurement 

unit

Digits

Calibration 

tags

"Sto" will be 
displayed when 
the calibration 
is saved.

Reviews: