To obtain the highest accuracy for measurements it is important to follow these tips:

•  Fresh buffer should be used for each calibration. Calibration is only as good as the 

buffer being used. The pH buffer values change over time once the sachets are opened. 

•  The probe should be rinsed with purified water each time before placing in buffer or 

sample to be tested.

•  When the meter is not in use it is important to add several drops of storage solution to 

the protective cap to keep the probe hydrated. If storage solution is not available, pH 

4.01 or pH 7.01 buffer can be used.

•  For improved accuracy it is recommended to calibrate in two buffers.

•  It is important to calibrate and measure samples at the same temperature as there 

is no temperature compensation. A dramatic change in temperature between buffer 

solutions and samples to be tested will give inaccurate readings.

Refilling the Electrode

•  If fouled, remove the sleeve and rinse sleeve and sensor tip with purified water.  

Turn over the probe, remove the sleeve by carefully rotating it and pull straight 

along the axis of the electrode. Use 

care as the pH stem is made of glass. 

Rinse off any traces of electrolyte gel.

•  Soak the sensor tip in HI70061G, 

HI700661P, HI700663P or HI700664P 

cleaning solution for 20 minutes then 

rinse with purified water.

•  Refill the reference well with HI9071 

Gelled Bridge Electrolyte.

•  Replace the reference sleeve: insert and 

push the sleeve onto the electrode. Make 

sure the black O-ring is fixed inside the 

electrode body. Any excess of gel will be 

expelled from the end of the electrode through the open junction. 

•  Rinse any excess gel off with purified water and gently dry off 

body with a soft cloth or tissue.

•  Soak assembled probe in Electrode storage solution for a 

minimum of 30 minutes.

•  Rinse probe with purified water. 

•  Shake the electrode down as you would do with a clinical 

thermometer to eliminate any air bubbles inside the glass bulb. 

•  Calibrate in fresh buffers before using for measurements.

Care and Maintenance

To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter 
counterclockwise to unlock. Remove cover and replace the battery with + side facing up.

Note: Batteries should only be replaced in a safe area using the battery 

type specified in this instruction manual. Old batteries should be disposed in 

accordance with local regulations.

Battery Indicator

The meter features a low battery indicator. When 
the battery is running low, the tag will blink on 
screen. When the battery has been depleted, 
“Erb” will appear on screen and the meter will 
turn off.

Clear Calibration

Place meter in calibration mode. Press and hold 
ON/OFF until “CLr” is displayed. The meter will 
now be at default calibration.

“Err” Message

In calibration mode, if the meter displays an 
“Err” message when in the correct fresh buffer 
solution then the probe should be cleaned.  
Place the probe in the

 

cleaning solution for 

20 minutes. Rinse with purified water and place 
in storage solution for a minimum of 30 minutes 
before calibrating. 

Battery Replacement

Auto-off

From measurement mode, press and hold the  
ON/OFF button. The meter will cycle through 
“OFF,” “CAL,” then current auto-off setting.
The default setting is 8 minutes (“d08”). Press 
ON/OFF button to change. “d60” is auto-off after 
60 minutes, and “d--” disables the auto-off 
feature. Press and hold the button to exit the 
menu.

Electrode Cleaning Solutions

Code

Description

HI70061G

GroLine general purpose cleaning solution, 20 mL sachets (25 pcs.)

HI700661P

General purpose cleaning solution for agriculture, 20 mL sachets (25 pcs.)

HI700663P

Cleaning solution for soil deposits, 20 mL sachets (25 pcs.)

HI700664P

Cleaning solution for humus deposits, 20 mL sachets (25 pcs.)

pH Buffer Solution

Code

Description

HI70004G

GroLine pH 4.01 buffer solution, 20 mL sachets (25 pcs.)

HI70007G

GroLine pH 7.01 buffer solution, 20 mL sachets (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachets (10 pcs., 5 ea.)

Accessories

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific application and for the 
environment in which it is used. Any variation introduced by the user to the supplied equipment may 
degrade the meters’ performance. For yours and the meter’s safety do not use or store the meter in 
hazardous environments.

Warranty

The meter is warranted for a period of one year against defects in workmanship and materials when 
used for their intended purpose and maintained according to instructions. This warranty is limited to 
repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed 
maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under 
warranty, report the model number, date of purchase, serial number and the nature of the problem. If the 
repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to 
be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the 
Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, 
make sure it is properly packaged for complete protection.

Certification

All Hanna Instruments conform to the 

CE European Directives

.

Disposal of Electrical & Electronic Equipment.

 The product should not be treated as household 

waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic 
equipment which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do not dispose of them 

with other household waste. Hand them over to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative consequences for the 
environment and human health. For more information, contact your city, your local household waste 
disposal service, the place of purchase or go to 

www.hannainst.com

.

RoHS

compliant

HI981030 

 

Soil pH Tester

Other Accessories

Code

Description

HI721319

Soil auger

HI7051M

Soil preparation solution, 230 mL bottle

Electrode Storage Solutions

Code

Description

HI70300G

GroLine electrode storage solution, 20 mL sachets (25 pcs.)

HI70300-023

GroLine electrode storage solution, 230 mL bottle

HI9072

Electrode storage solution, 13 mL dropper

Electrode Fill Solution

Code

Description

HI9071

Gelled Bridge Electrolyte