Preliminary Examination

Remove the tester and accessories from the packing material and examine 

it carefully. If you require any further information, please contact Hanna 

Instruments technical support team at [email protected].

Each 

HI981038

 is delivered in a cardboard box and is supplied with:

• 

HI70004

 pH 4.01 buffer solution, 20 mL sachet (2 pcs.)

• 

HI70007 

pH 7.01 buffer solution, 20 mL sachet (2 pcs.)

• 

HI700684

 Electrode cleaning and disinfection solution for bread and dough 

deposits, 20 mL sachet (2 pcs.)

• 

HI9072

 Electrode storage solution, 13 mL dropper

• CR2032 3V Lithium-ion battery

• Instrument quality certificate

• Instruction manual

Note: 

Save all packing material until you are sure that the tester works 

correctly. Any damaged or defective item must be returned in its original 

packing material with the supplied accessories.

General Description & Intended Use

The 

HI981038

 Bread & Dough pH Tester is designed to measure the pH during baking 

processes and ensure it meets the food-hygiene and Hazard Analysis Critical Control 

Point (HACCP) regulations. 

It features a single button operation system and is easy to use. It has a compact and 

waterproof casing and automatic pH calibration at one or two points. 

All readings are automatically compensated for temperature variations with a built-in 

temperature sensor. The conical shaped electrode is ideal to pierce solid and semisolid 

samples. The open junction prevents samples from clogging the junction.

Probe Features

Conical Tip

 

The conical tip allows for direct measurement in whey 

and semisolid samples.

Specialized Glass Formulation

Specialized low temperature (LT) pH glass ensures fast stabilization and accurate 

results at lower temperatures.

Open Junction Reference

 

The open junction design consists of a solid gel interface between the dough 

sample and internal Ag/AgCl reference. This interface prevents silver from 

Dear Customer, 

Thank you for choosing a Hanna Instruments product. Please read this 

instruction manual carefully before using the tester. 

For more information about Hanna Instruments and our products, visit  

www.hannainst.com or e-mail us at [email protected].

For technical support, contact your local Hanna Instruments Office or 

e-mail us at [email protected].

entering the sample and makes it impermeable to clogging, resulting in a fast 

response and stable reading.

Polyvinylidene Fluoride (PVDF) Body

 

PVDF is a food-grade plastic that is resistant to most chemicals and solvents, 

including sodium hypochlorite, it is also resistant to fungal growth. It has high 

abrasion resistance, mechanical strength, and resistance to ultraviolet. 

Specifications

Range

0.00 to 12.00 pH

Resolution

0.01 pH

Accuracy 

±0.05 pH

Calibration

Automatic, one or two-point

Temperature compensation

Automatic, 0 to 50 °C 

Electrode

Built-in probe for specific application

Battery type

CR2032 Lithium-ion (included)

Battery life

Approximately 800 hours of continuous use

Auto-off

8 minutes, 60 minutes or disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

50 x 129 x 21 mm (2 x 5.1 x 0.9“) 

Weight

42 g (1.48 oz.)

Functional Description & LCD Display

Liquid Crystal Display (LCD)

ON/OFF button

Maximum immersion level

Recommended immersion level

pH probe

Calibration tags

Preparation

• Remove the protective cap. Do not be alarmed if salt deposits are present. 

Rinse the probe with water and blot dry.

• If the glass and / or junction are dry, soak the electrode (bottom 4 cm / 

1.5 “) in in storage solution for a minimum of 30 minutes. Rinse with water 

and blot dry.

• Calibrate the electrode before using. For best results is recommended to 

recalibrate periodically.

Storage

• To ensure a quick response, the glass tip and junction should be kept moist.

• Replace the protective cap with a few drops of storage solution when not in 

use. Do not store the electrode in distilled or deionized water.

Operation

Press the ON/OFF button to turn the tester 

on. The tester displays all LCD segments 

for a few seconds. The tester will enter 

measurement mode, the current reading 

and calibrated buffers will be shown.

Low battery 

indicator

Stability 

indicator

Measurement

unit

Calibration 

tags

Calibration

From measurement mode, press and hold the ON/OFF button until “CAL” is 
displayed.

 

 

For one or two-point calibration using pH 7.01 buffer solution, follow 
procedure A.

For one-point calibration using pH 4.01 buffer solution, follow 
procedure B.

Note: 

It is recommended to calibrate the electrode with buffers at the 

temperature it will be used at.

A

B

A

One or Two-Point Calibration with pH 7.01

When "7.01" is displayed, 

place the tip of the electrode 

in pH 7.01 buffer.

When the reading is stable, the 

stability icon will disappear.

When "4.01" is displayed, follow 

the procedure to the right for a 

one or two-point calibration.

One

-P

oint

Press the ON/OFF 

button to save 

the one-point 

calibration.

"Sto" will be 

displayed when 

the calibration is 

saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tag will be 

displayed.

Two

-P

oint

Place the tip of 

the electrode in 

pH 4.01 buffer. 

The buffer is 

automatically 

recognized, the 

stability indicator 

will blink.

Wait until the 

measurement is 

stable and the 

stability indicator 

disappears. "Sto" 

will be displayed 

when the calibration 

is saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tags will be 

displayed.

One-Point Calibration with pH 4.01

B

When "7.01" is 

displayed, place the 

tip of the electrode 

in pH 4.01 buffer. 

The buffer value 

will be recognized 

automatically and 

"4.01" will be 

displayed.

When the 

reading is stable, 

the stability 

indicator will 

disappear.“Sto” 

will be displayed 

when the 

calibration is 

saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tag will be 

displayed.

Summary of Contents for Foodcare HI981038

Page 1: ...HI981038 Bread Dough pH Tester INSTRUCTION MANUAL ...

Page 2: ...rol Point HACCP regulations It features a single button operation system and is easy to use It has a compact and waterproof casing and automatic pH calibration at one or two points All readings are automatically compensated for temperature variations with a built in temperature sensor The conical shaped electrode is ideal to pierce solid and semisolid samples The open junction prevents samples fro...

Page 3: ...ge 0 00 to 12 00 pH Resolution 0 01 pH Accuracy 0 05 pH Calibration Automatic one or two point Temperature compensation Automatic 0 to 50 C Electrode Built in probe for specific application Battery type CR2032 Lithium ion included Battery life Approximately 800 hours of continuous use Auto off 8 minutes 60 minutes or disabled Environment 0 to 50 C 32 to 122 F RH 95 max Dimensions 50 x 129 x 21 mm ...

Page 4: ...mmended to recalibrate periodically Storage To ensure a quick response the glass tip and junction should be kept moist Replace the protective cap with a few drops of storage solution when not in use Do not store the electrode in distilled or deionized water Operation Press the ON OFF button to turn the tester on The tester displays all LCD segments for a few seconds The tester will enter measureme...

Page 5: ...L is displayed For one or two point calibration using pH 7 01 buffer solution follow procedure A For one point calibration using pH 4 01 buffer solution follow procedure B Note It is recommended to calibrate the electrode with buffers at the temperature it will be used at A B ...

Page 6: ... electrode in pH 4 01 buffer The buffer is automatically recognized the stability indicator will blink Wait until the measurement is stable and the stability indicator disappears Sto will be displayed when the calibration is saved The tester will return to measurement mode and the calibration tags will be displayed One Point Calibration with pH 4 01 B When 7 01 is displayed place the tip of the el...

Page 7: ...ation and restore the tester to factory default From calibration mode press and hold the ON OFF button the LCD will show CLr Err Message In calibration mode if the probe is in the correct buffer solution and the Err message is displayed the probe should be cleaned Soak the probe in cleaning solution for 20 minutes Rinse with water and hydrate electrode in storage solution for a minimum of 30 minut...

Page 8: ...e probe in distilled or deionized water Note Never immerse the tester over the maximum immersion level Battery Replacement To change the CR2032 lithium ion battery turn the battery cover located on the back of the tester counterclockwise to unlock Remove cover and replacethebatterywithpositive sidefacingout Note Only use the battery type specified in the manual Old batteries should be disposed in ...

Page 9: ...on introduced by the user to the supplied equipment may degrade the tester s performance For yours and the tester s safety do not use or store the tester in hazardous environments Certification All Hanna Instruments conform to the CE European Directives Disposal of Electrical Electronic Equipment The product should not be treated as household waste Instead hand it over to the appropriate collectio...

Reviews: