Hanna Instruments Foodcare HI981036 Instruction Manual Download Page 2

Liquid Crystal 
Display
ON/OFF 
Button

Maximum 
Immersion 
Level

Recommended 
Immersion 
Level

pH Probe

If pH 4.01 buffer 
solution is used 
as the first point

 

the value of the 
buffer is recognized 
and displayed 
with the blinking 
stability indicator.

If pH 7.01 buffer solution is used as 
the first point

 the buffer is recognized 

with the blinking stability indicator.
When the reading is stable, the 
stability indicator will disappear 
and pH 7.01 will be calibrated. 

If pH 7.01 is the only calibration point,

 

finish one-point procedure at right. 

If using pH 4.01 as a second point,

 

continue two-point procedure at right.

For one-point calibration using pH 4.01 buffer go to procedure B 

For one or two-point calibration using pH 7.01 buffer go to procedure A

A

B

When the 
reading is stable, 
the stability 
indicator will 
disappear.“Sto” 
will be displayed 
when the 
calibration 
is saved.

Meter will 
exit to 
measurement 
mode and the 
calibration 
tag will be 
displayed.

Meter will exit 
to measurement 
mode and the 
calibration 
tag will be 
displayed.

Tw

o-

Po

in

t

Use pH 4.01 to 
perform a two-
point calibration. 
The value is 
automatically 
recognized and 
displayed with the 
blinking stability 
indicator.

When the reading 
is stable, the 
stability indicator 
will disappear. 
“Sto” will be 
displayed when 
the calibration 
is saved.

Meter will exit 
to measurement 
mode and the 
calibration 
tags will be 
displayed

.

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using this instrument. 
For technical support, contact your local Hanna Instruments Office or 
email us at [email protected].
To find your local Hanna Instruments Office or for additional information on 
Hanna Instruments products, visit www.hannainst.com

Meter Calibration

From measurement 
mode, press and 
hold the ON/OFF 
button until “CAL” 
is displayed.

When “7.01” blinks 
on the display, place 
the tip of the probe 
into a pH 7.01 or 
4.01 buffer solution.

On

e-

Po

in

t

One-Point Calibration with pH 4.01

B

A

One or Two-Point Calibration with pH 7.01

pH 4.01 will then 
blink on the display. 
Ignore it and press 
ON/OFF button.

Operation

Press the ON/OFF button to turn the meter on. All tags 
will be displayed.

The meter will go into measurement mode: current 
reading and calibrated buffers are displayed.

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy 

±0.2 pH @25 °C/77 °F

Calibration

Automatic, one or two-point

Electrode

Built-in probe for specific application

Battery Type

CR2032 Li-ion

Battery Life

Approximately 1000 hours of continuous use

Auto-off

8 minutes, 60 minutes or can be disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

51 x 148 x 21 mm (2 x 5.8 x 0.9“) 

Weight

43 g (1.5 oz.)

Preliminary Examination

Remove the meter from the packing material and examine it carefully to make sure that 

no damage has occurred during shipment. If noticeable damage is evident, contact your 

local Hanna Instruments Office.
Each meter is supplied with:

• pH 4.01 buffer solution liquid sachet (2 pcs.)

• pH 7.01 buffer solution liquid sachet (2 pcs.)

• Meat, grease and fats acid cleaning solution (2 pcs.)

• Electrode storage solution, 13 mL dropper

• Electrode fill solution

• Instruction manual
• Quality certificate

Note: Save all packing material until you are sure that the instrument functions correctly. 
All defective items must be returned in the original packaging with the supplied accessories.

Intended use

The meter has been specially designed for meat processing. In the meat processing 

industry, the monitoring of pH is considered to be of the utmost importance due to its 

effect on the meat's quality factors including water binding capacity and shelf life. Upon 

slaughter, biochemical processes begin to break down the meat. Glycolysis begins post-

mortem, converting glycogen to lactic acid, reducing the pH of the carcass. Depending 

on a number of factors such as type of animal and even breed, this decrease in pH can 

take anywhere from a single hour to many. It is vital to monitor pH during this phase 

as once the lowest pH value is reached, the pH will begin to slowly rise, indicating that 

decomposition has begun.

The pH value of meat influences its’ water binding capacity which directly impacts 

consumer qualities such as tenderness and color. Lower pH values result in a lower 

water-binding capacity and lighter colors. Factors such as these can be important when 

considering how to efficiently produce meat products. For example, when producing dry 

sausages the meat must have a low water binding capacity so that it can dry evenly.

Depending on the type of the final product and the steps required to get there, pH 

values will vary throughout the meat processing industry. It is imperative, regardless 

of the final product, that pH be maintained at a low value to prevent bacterial spoilage 

and comply with food safety regulations. By monitoring pH values throughout the meat 

production process, you can ensure the creation of consistent and safe meat products.

Meter Overview

Preparation:

The pH electrode is shipped with a protective cap containing storage solution. Before 
using the meter, 

remove the protective

 cap and condition the electrode by soaking the 

tip (bottom 4 cm (1.5”)) in pH 7.01 buffer solution for several minutes. Then follow the 
calibration procedure.

•  Do not be alarmed if white crystals appear around the cap. This is normal with pH 

electrodes and they dissolve when rinsed with water.

•  Turn the meter on by pressing ON/OFF button.

NEVER IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION LEVEL.

•  For best results recalibrate periodically.
•  After use, rinse the electrode with water and store it with a few drops of storage 

solution in the protective cap.

•  Reinstall the protective cap after each use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.

Open Reference Junction with removable 

sleeve design

Suspended solids can permanently clog the pores of a 
ceramic reference junction. The open junction design 
provides a constant junction potential and minimizes 
blockage by providing an open gel interface between the 
sample and internal Ag/AgCl reference. Should meat enter 
into the junction, the junction can be easily cleaned and 
refreshed with new bridge electrolyte.

Probe Features

PVDF Body

Polyvinylidene fluoride (PVDF) is a food grade plastic that is resistant to most chemicals 
and solvents, including sodium hypochlorite. It has high abrasion resistance, mechanical 
strength and resistance to ultraviolet. PVDF is also resistant to fungal growth.

Calibration 

tags

Battery 

indicator

Unstable 

tag

Measurement 

unit

Digits

Specifications

“Sto” will be 
displayed when 
the calibration 
is saved.