Hanna Instruments Foodcare HI981036 User Manual Download Page 2

Open Junction Reference

The open junction design consists of a solid gel 

interface between the meat sample and internal 

Ag/AgCl reference. This interface prevents silver from 

entering the sample and makes it impermeable to 

clogging, resulting in fast response and stable 

reading.

Specifications

Range

0.00 to 12.00 pH

Resolution

0.01 pH

Accuracy 

±0.05 pH 

Calibration

Automatic, one or two-point

Temperature compensation Automatic, 0 to 50 °C
Electrode

Built-in probe for specific application

Battery type

CR2032 Lithium-ion (included)

Battery life

Approximately 800 hours of continuous use

Auto-off

8 minutes, 60 minutes or disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

51 x 148 x 21 mm (2 x 5.8 x 0.9“) 

Weight

45 g (1.58 oz.)

Functional Description & LCD Display

Liquid Crystal Display (LCD)

Calibration tags

ON/OFF button

Maximum immersion level

Recommended immersion level

pH probe

Preliminary Examination

Remove the tester and accessories from the packing material and examine 

it carefully. If you require any further information, please contact Hanna 

Instruments technical support team at [email protected].

Each 

HI981036

 is delivered in a cardboard box and is supplied with:

• 

HI70004

 pH 4.01 buffer solution, 20 mL sachet (2 pcs.)

• 

HI70007 

pH 7.01 buffer solution, 20 mL sachet (2 pcs.)

• 

HI700630 

Meat, grease & fats acid cleaning solution, 20 mL sachet (2 pcs.)

• 

HI70300 

Electrode storage solution, 13 mL dropper

• 

HI9071

 Gelled bridge electrolyte

• CR2032 3V Lithium-ion battery

• Instrument quality certificate

• Instruction manual

Note: 

Save all packing material until you are sure that the tester works 

correctly. Any damaged or defective item must be returned in its original 

packing material with the supplied accessories.

General Description & Intended Use

The 

HI981036

 Meat pH Tester is designed to measure the pH during meat 

processing process  and ensure it meets the food-hygiene and Hazard Analysis 

Critical Control Point (HACCP) regulations. 

It features a single button operation system and is easy to use. It has a compact 

and waterproof casing and automatic pH calibration at one or two points. 

All readings are automatically compensated for temperature variations with a 

built-in temperature sensor.

The specialized  pH probe has an outer junction sleeve that can be removed and 

cleaned, extending the life of the tester. The conical tip allows measurements to 

be taken directly in solid or semisolid samples.

Probe Features

Polyvinylidene Fluoride (PVDF) Body

PVDF is a food-grade plastic that is resistant to most chemicals and solvents, 

including sodium hypochlorite, it is also resistant to fungal growth. It has high 
abrasion resistance, mechanical strength, and resistance to ultraviolet.

Dear Customer, 

Thank you for choosing a Hanna Instruments product. Please read this 

instruction manual carefully before using the tester. 

For more information about Hanna Instruments and our products, visit  

www.hannainst.com or e-mail us at [email protected].

For technical support, contact your local Hanna Instruments Office or 

e-mail us at [email protected].

Preparation

• Remove the protective cap. Do not be alarmed if salt deposits are present. 

Rinse the probe with water and blot dry.

• If the glass and / or junction are dry, soak the electrode (bottom 4 cm / 

1.5 “) in storage solution for a minimum of 30 minutes. Rinse with water 

and blot dry.

• Calibrate the electrode before using. For best results is recommended to 

recalibrate periodically.

Storage

• To ensure a quick response, the glass tip and junction should be kept moist.

• Replace the protective cap with a few drops of storage solution when not in 

use. Do not store the electrode in distilled or deionized water.

Operation

Press the ON/OFF button to turn the tester 

on. The tester displays all LCD segments 

for a few seconds. The tester will enter 

measurement mode, the current reading 

and calibrated buffers will be shown.

Low battery 

indicator

Stability 

indicator

Measurement

unit

Calibration 

tags

Calibration

From measurement mode, press and hold the ON/OFF button until “CAL” is 
displayed.

 

 

For one or two-point calibration using pH 7.01 buffer solution, follow 
procedure A.

For one-point calibration using pH 4.01 buffer solution, follow 
procedure B.

Note: 

It is recommended to calibrate the electrode with buffers at the 

temperature it will be used at.

A

B

A

One or Two-Point Calibration with pH 7.01

When "7.01" is displayed, 

place the tip of the electrode 

in pH 7.01 buffer.

When the reading is stable, the 

stability icon will disappear.

When "4.01" is displayed, follow 

the procedure to the right for a 

one or two-point calibration.

One

-P

oint

Press the ON/OFF 

button to save 

the one-point 

calibration.

"Sto" will be 

displayed when 

the calibration is 

saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tag will be 

displayed.

Two

-P

oint

Place the tip of 

the electrode in 

pH 4.01 buffer. 

The buffer is 

automatically 

recognized, the 

stability indicator 

will blink.

Wait until the 

measurement is 

stable and the 

stability indicator 

disappears. "Sto" 

will be displayed 

when the calibration 

is saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tags will be 

displayed.

One-Point Calibration with pH 4.01

B

When "7.01" is 

displayed, place the 

tip of the electrode 

in pH 4.01 buffer. 

The buffer value 

will be recognized 

automatically and 

"4.01" will be 

displayed.

When the 

reading is stable, 

the stability 

indicator will 

disappear.“Sto” 

will be displayed 

when the 

calibration is 

saved.

The tester 

will return to 

measurement 

mode and the 

calibration 

tag will be 

displayed.

Summary of Contents for Foodcare HI981036

Page 1: ...HI981036 Meat pH Tester INSTRUCTION MANUAL ...

Page 2: ... hygiene and Hazard Analysis Critical Control Point HACCP regulations It features a single button operation system and is easy to use It has a compact and waterproof casing and automatic pH calibration at one or two points All readings are automatically compensated for temperature variations with a built in temperature sensor The specialized pH probe has an outer junction sleeve that can be remove...

Page 3: ...cy 0 05 pH Calibration Automatic one or two point Temperature compensation Automatic 0 to 50 C Electrode Built in probe for specific application Battery type CR2032 Lithium ion included Battery life Approximately 800 hours of continuous use Auto off 8 minutes 60 minutes or disabled Environment 0 to 50 C 32 to 122 F RH 95 max Dimensions 51 x 148 x 21 mm 2 x 5 8 x 0 9 Weight 45 g 1 58 oz Functional ...

Page 4: ...ure B Note It is recommended to calibrate the electrode with buffers at the temperature it will be used at A B A One or Two Point Calibration with pH 7 01 When 7 01 is displayed place the tip of the electrode in pH 7 01 buffer When the reading is stable the stability icon will disappear When 4 01 is displayed follow the procedure to the right for a one or two point calibration One Point Press the ...

Page 5: ...icator When the battery level is low the tag on the LCD will blink When the battery is depleted the Erb message is displayed and the tester is powered off Care Maintenance General Information Please read the information below to ensure the highest possible accuracy Fresh buffer should be used for each calibration once the sachets are opened the buffer value can change over time For improved accura...

Page 6: ...Gelled bridge electrolyte Replace the sleeve making sure that the black O ring is fixed inside the electrode Rinse off excess gel with distilled or deionized water Shake the probe down as you would do with a clinical thermometer to eliminate any trapped air bubbles SoaktheprobeinHI70300Electrodestoragesolutionfora minimum of 30 minutes before calibrating Accessories pH Buffer Solution HI70004P pH ...

Page 7: ...on Recommendations for Users Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by theusertothesuppliedequipmentmaydegradethetester sperformance Foryours and the tester s safety do not use or store the meter in hazardous environments Certification All Hanna Instruments conform to the CE Euro...

Reviews: