• The outer junction sleeve can be removed and cleaned with purified water. 

Once cleaned a small amount of supplied gel electrolyte should be added to 

refresh the junction and improve the pH measurement.

Note: 

Never immerse the tester over the maximum immersion level.

Refilling the Electrode

• To remove the electrode sleeve, rotate it carefully and slide it off the probe body, 

keeping it parallel to the pH electrode.

Note: 

Handle the probe with care, the 

pH electrode stem is made of glass.

• Rinse off any traces of electrolyte  

gel from the sleeve. Soak the tip in 

HI700601

 General purpose cleaning 

solution or 

HI700630

 Meat, grease 

and fats acid cleaning solution for 

20 minutes. Rinse with distilled or 

deionized.

• Refill the reference well with 

HI9071

 

Gelled bridge electrolyte.

• Replace the sleeve making sure that the black O ring is 

fixed inside the electrode.

• Rinse off excess gel with distilled or deionized water.

• Shake the probe down as you would do with a clinical 

thermometer to eliminate any trapped air bubbles.

• Soak the probe in 

HI70300

 Electrode storage solution for a 

minimum of 30 minutes before calibrating.

Accessories

pH Buffer Solution

HI70004P

pH 4.01 buffer solution, 20 mL sachet (25 pcs.)

HI70007P

pH 7.01 buffer solution, 20 mL sachet (25 pcs.)

HI77400P

pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each)

Electrode Cleaning Solution

HI700601P

General purpose cleaning solution, 20 mL sachet (25 pcs.)

HI700630P

Meat, grease and fats acid cleaning solution, 20 mL sachet (25 pcs.)

Electrode Storage Solution

HI70300L

Electrode storage solution, 500 mL

HI70300M

Electrode storage solution, 230 mL

HI9072

Electrode storage solution, 13 mL dropper

Electrode Fill Solution

HI9071

Gelled bridge electrolyte

Auto-Off

From measurement mode, press and hold the ON/OFF 

button. The LCD will show "OFF", "CAL" followed by 

"d08" (default setting, 8 minutes). Release the ON/OFF 

button. A single press on the ON/OFF button will change 

the auto-off timer to "d60" (60 minutes). To disable the 

auto-off feature press the ON/OFF button again, LCD will 

show "d--". Press and hold to exit.

Clear Calibration

To clear the user calibration and restore the tester 

to factory default. From calibration mode, press and 

hold the ON/OFF button, the LCD will show "CLr".

“Err” Message

In calibration mode, if the probe is in the correct 

buffer solution and the "Err" message is displayed, the 

probe should be cleaned. Soak the probe in cleaning 

solution for 20 minutes. Rinse with water and hydrate 

electrode in storage solution for a minimum of 30 

minutes before calibrating.

Battery Indicator

When the battery level is low, the tag on the LCD will 

blink. When the battery is depleted the "Erb" message 

is displayed and the tester is powered off.

Care & Maintenance

General Information

Please read the information below, to ensure the highest possible accuracy:

• Fresh buffer should be used for each calibration, once the sachets are opened 

the buffer value can change over time.

• For improved accuracy a two-point calibration is recommended.

• If the electrode is slow or sluggish, soak it in cleaning solution for 20 

minutes. Rinse with water and hydrate the electrode in storage solution for a 

minimum of 30 minutes before calibrating.

• If measurements are taken successively, rinse the probe thoroughly in 

distilled or deionized water to eliminate cross-contamination.

• When not in use, add a few drops of storage solution to the protective cap 

to keep the glass tip and the junction hydrated. If storage solution is not 

available, pH 4.01 or pH 7.01 buffer can be used. Never store the probe in 

distilled or deionized water.

Battery Replacement

To change the CR2032 lithium-ion battery, turn 

the battery cover located on the back of the tester 

counterclockwise to unlock. Remove cover and 

replace the battery with positive (+) side facing out.

Note: 

Only use the battery type specified in 

the manual. Old batteries should be disposed 

in accordance with local regulations.

Warranty

The 

HI981036

 is warranted for a period of one year against defects in 

workmanship and materials when used for its intended purpose and maintained 

according to instructions. This warranty is limited to repair or replacement free 

of charge. Damage due to accidents, misuse, tampering or lack of prescribed 

maintenance is not covered. If service is required, contact your local Hanna 

Instruments Office. If under warranty, report the model number, date of 

purchase, serial number and the nature of the problem. If the repair is not 

covered by the warranty, you will be notified of the charges incurred. If the 

instrument is to be returned to Hanna Instruments, first obtain a Returned 

Goods Authorization (RGA) number from the Technical Service department and 

then send it with shipping costs prepaid. When shipping any instrument, make 

sure it is properly packaged for complete protection.

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific 

application and for the environment in which it is used. Any variation introduced by 

the user to the supplied equipment may degrade the tester’s performance. For yours 

and the tester’s safety do not use or store the meter in hazardous environments.

Certification

All Hanna Instruments conform to the 

CE European Directives

.

Disposal of Electrical & Electronic Equipment

. The product 

should not be treated as household waste. Instead hand it over to 

the appropriate collection point for the recycling of electrical and 

electronic equipment which will conserve natural resources.

Disposal of waste batteries.

 This product contains batteries, do 

not dispose of them with other household waste. Hand them over 

to the appropriate collection point for recycling.

Ensuring proper product and battery disposal prevents potential negative 

consequences for the environment and human health. For more information, 

contact your city, your local household waste disposal service, the place of 

purchase or go to 

www.hannainst.com

.

Hanna Instruments reserves the right to modify the design, construction, or 

appearance of its products without advance notice.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent 

of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

IST981036 10/19

RoHS

compliant

HI981036

Meat pH Tester

INSTRUCTION MANUAL

Summary of Contents for Foodcare HI981036

Page 1: ...HI981036 Meat pH Tester INSTRUCTION MANUAL ...

Page 2: ... hygiene and Hazard Analysis Critical Control Point HACCP regulations It features a single button operation system and is easy to use It has a compact and waterproof casing and automatic pH calibration at one or two points All readings are automatically compensated for temperature variations with a built in temperature sensor The specialized pH probe has an outer junction sleeve that can be remove...

Page 3: ...cy 0 05 pH Calibration Automatic one or two point Temperature compensation Automatic 0 to 50 C Electrode Built in probe for specific application Battery type CR2032 Lithium ion included Battery life Approximately 800 hours of continuous use Auto off 8 minutes 60 minutes or disabled Environment 0 to 50 C 32 to 122 F RH 95 max Dimensions 51 x 148 x 21 mm 2 x 5 8 x 0 9 Weight 45 g 1 58 oz Functional ...

Page 4: ...ure B Note It is recommended to calibrate the electrode with buffers at the temperature it will be used at A B A One or Two Point Calibration with pH 7 01 When 7 01 is displayed place the tip of the electrode in pH 7 01 buffer When the reading is stable the stability icon will disappear When 4 01 is displayed follow the procedure to the right for a one or two point calibration One Point Press the ...

Page 5: ...icator When the battery level is low the tag on the LCD will blink When the battery is depleted the Erb message is displayed and the tester is powered off Care Maintenance General Information Please read the information below to ensure the highest possible accuracy Fresh buffer should be used for each calibration once the sachets are opened the buffer value can change over time For improved accura...

Page 6: ...Gelled bridge electrolyte Replace the sleeve making sure that the black O ring is fixed inside the electrode Rinse off excess gel with distilled or deionized water Shake the probe down as you would do with a clinical thermometer to eliminate any trapped air bubbles SoaktheprobeinHI70300Electrodestoragesolutionfora minimum of 30 minutes before calibrating Accessories pH Buffer Solution HI70004P pH ...

Page 7: ...on Recommendations for Users Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by theusertothesuppliedequipmentmaydegradethetester sperformance Foryours and the tester s safety do not use or store the meter in hazardous environments Certification All Hanna Instruments conform to the CE Euro...

Reviews: