Extractable Cloth Junction
Allows for clearing any clogging from
solids that would result in slow response
and unstable readings. Pull 3 mm
(1/8 “) to expose a fresh new surface
of the junction resulting in fast response
time and reading stability.
Titanium Body
Rugged and resilient, the titanium body of the probe works as an electronic shield
protecting against interferences from electrical noise or humidity. The titanium
body is virtually unbreakable, offering protection from accidental breakage.
Specifications
Range
0.00 to 12.00 pH
Resolution
0.01 pH
Accuracy
±0.05 pH
Calibration
Automatic, one or two-point
Temperature compensation Automatic, 0 to 80 °C
Electrode
Built-in probe for specific application
Battery type
CR2032 Lithium-ion (included)
Battery life
Approximately 800 hours of continuous use
Auto-off
8 minutes, 60 minutes or disabled
Environment
0 to 50 °C (32 to 122 °F); RH 95% max
Dimensions
51 x 165 x 21 mm (2 x 6.5 x 0.9“)
Weight
60 g (2.11 oz.)
Functional Description & LCD Display
Liquid Crystal Display (LCD)
Calibration tags
ON/OFF button
Maximum immersion level
pH sensitive glass
pH probe
Cloth
junction
pH sensitive glass
Dear Customer,
Thank you for choosing a Hanna Instruments product. Please read this
instruction manual carefully before using the tester.
For more information about Hanna Instruments and our products, visit
www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments Office or
e-mail us at [email protected].
Preliminary Examination
Remove the tester and accessories from the packing material and examine
it carefully. If you require any further information, please contact Hanna
Instruments technical support team at [email protected].
Each
HI981031
is delivered in a cardboard box and is supplied with:
•
HI70004
pH 4.01 buffer solution, 20 mL sachet (2 pcs.)
•
HI70007
pH 7.01 buffer solution, 20 mL sachet (2 pcs.)
•
HI700682
Cleaning solution for brewing deposits (2 pcs.)
•
HI9072
Electrode storage solution, 13 mL dropper
• CR2032 3V Lithium-ion battery
• Instrument quality certificate
• Instruction manual
Note:
Save all packing material until you are sure that the tester works
correctly. Any damaged or defective item must be returned in its original
packing material with the supplied accessories.
General Description & Intended Use
The
HI981031
Beer pH Tester is designed to measure the pH during the
brewing process. It features a single button operation system and is easy to
use. It has a compact and waterproof casing and automatic pH calibration at
one or two points. All readings are automatically compensated for temperature
variations with a built-in temperature sensor. The flat tip of the pH electrode
allows easy cleaning of the pH sensing surface as solids from mash and cooled
wart collect on the surface. The pH electrode is equipped with extractable cloth
junction. Pulling out a small portion of the junction increases response time and
extends electrode life.
Probe Features
Flat Glass Tip
The flat tip provides optimal surface contact and allows easy cleaning of the
sensing surface as solids collect on the surface. The pH electrode is designed to
measure the pH of samples with temperature up to 80 °C (176 °F).
Preparation
• Remove the protective cap. Do not be alarmed if salt deposits are present.
Rinse the probe with water and blot dry.
• If the glass and / or junction are dry, soak the electrode (bottom 4 cm /
1.5“) in storage solution for a minimum of 30 minutes. Rinse with water
and blot dry.
• Calibrate the electrode before using. For best results is recommended to
recalibrate periodically.
Storage
• To ensure a quick response, the glass tip and junction should be kept moist.
• Replace the protective cap with a few drops of storage solution when not in
use. Do not store the electrode in distilled or deionized water.
Operation
Press the ON/OFF button to turn the tester
on. The tester displays all LCD segments
for a few seconds. The tester will enter
measurement mode, the current reading
and calibrated buffers will be shown.
Low battery
indicator
Stability
indicator
Measurement
unit
Calibration
tags
Calibration
From measurement mode, press and hold the ON/OFF button until “CAL” is
displayed.
For one or two-point calibration using pH 7.01 buffer solution, follow
procedure A.
For one-point calibration using pH 4.01 buffer solution, follow
procedure B.
Note:
It is recommended to calibrate the electrode with buffers at the
temperature it will be used at.
A
B
A
One or Two-Point Calibration with pH 7.01
When "7.01" is displayed, place
the tip of the electrode in pH 7.01
buffer.
When the reading is stable, the
stability icon will disappear.
When "4.01" is displayed, follow
the procedure to the right for a
one or two-point calibration.
One
-P
oint
Press the ON/OFF
button to save
the one-point
calibration.
"Sto" will be
displayed when
the calibration is
saved.
The tester
will return to
measurement
mode and the
calibration
tag will be
displayed.
Two
-P
oint
Place the tip of
the electrode in
pH 4.01 buffer.
The buffer is
automatically
recognized, the
stability indicator
will blink.
Wait until the
measurement is
stable and the
stability indicator
disappears. "Sto"
will be displayed
when the calibration
is saved.
The tester
will return to
measurement
mode and the
calibration
tags will be
displayed.
One-Point Calibration with pH 4.01
B
When "7.01" is
displayed, place the
tip of the electrode
in pH 4.01 buffer.
The buffer value
will be recognized
automatically and
"4.01" will be
displayed.
When the
reading is stable,
the stability
indicator will
disappear.“Sto”
will be displayed
when the
calibration is
saved.
The tester
will return to
measurement
mode and the
calibration
tag will be
displayed.