background image

HI9350011

Portable K-Type Fast Response Thermometer for 

Foodcare Products

SPECIFICATIONS

 RANGE*

 

  -50.0 to 199.9 ºC ; 200 to 300 ºC

    

  -58.0 to 399.9 ºF ; 400 to 572 ºF

  METER ACCURACY

    ±0.4 ºC (-50.0 to 300 ºC)

  @ 23.0 °C ±5 °C    ±0.7 ºF (-58.0 to 572 ºF) 

  PROBE ACCURACY 

  ±1.6 ºC (-50.0 to 300 ºC) 

 

FC766C1

 

  ±2.9 ºF (-58.0 to 572 ºF)

 

*The measurement range may be limited by probe type, and applies to the probe shaft.

OPERATIONAL GUIDE

 

To switch ON and OFF: 

press 

ON/OFF

 

( )

.

 

To take measurements: 

connect a suitable temperature probe to the meter. 

At start-up the thermometer displays all LCD segments for a few seconds, 
followed by the percentage indication of the remaining battery life. It then 
enters CAL Check mode followed by measurement mode.

  If a temperature probe is plugged in, the meter displays the measured 

temperature.

  If reading is slightly beyond the measurement range, the meter flashes the 

closest full-scale value.

  To switch the meter 

OFF

, press the 

ON/OFF

 key.

 

 For auto-off feature: 

  To save battery life, each model is provided with an auto-off feature, which 

switches the instrument off after a certain period of non use.

  The auto-off feature can be set in 3 modes from 

SETUP MENU

: 8 minutes, 

60 minutes or disabled.

• 

Calibration Check mode:

  The meter is provided with an automatic CAL Check feature at start-up, which 

allows the user to check the instrument calibration status. If the meter is 
calibrated, the display will stabilize and show “

0.0

 

°C

” 

(±0.3 °C)

 for a 

few seconds.

Warning messages:

• “

NO PROBE

 message

: when the probe is not connected or the signal is 

interrupted.

• “

Err

 message

: if during Calibration Check the meter displays more than  

±0.3 °C. Contact your local Hanna Instruments Office for recalibration.

• 

DAMAGED PROBE message

: when the readings are very far from the 

measurement range, the probe may be damaged. The display shows 
blinking dashes and “

NO PROBE

” message is displayed.

Quick Reference Guide