SPECIFICATIONS:
RANGE / RESOLUTION
DiST1
(HI98300 & HI98301)
0 to 1999 ppm/1ppm
DiST2
(HI98302)
0.00 to 10.00 ppt / 0.01ppt
DiST3
(HI98303)
0 to 1999 µS/cm / 1µS/cm
DiST4
(HI98304)
0.00 to 19.99 mS/cm / 0.01mS/cm
ACCURACY
(@20°C/68°F)
±2% f.s.
TYPICAL EMC DEVIATION
± 1% f.s.
TDS FACTOR
HI98300
0.65
HI98301 & HI98302
0.5
CALIBRATION SOLUTIONS
HI98300
HI70442 (1500 ppm)
HI98301
HI70032 (1382 ppm)
HI98302
HI70038 (6.44 ppt)
HI98303
HI70031 (1413 µS/cm)
HI98304
HI70039 (5.00 mS/cm)-HI70030 (12.88 mS/cm)
TEMP.COMPENSATION
0 to 50 C (32 to 122ºF)
ENVIRONMENT
0 to 50 C (32 to 122 F);
95% RH not condensing
BATTERY TYPE
4 x 1.5V alkaline (included)
BATTERY LIFE
approx. 200 hours
of continuous use
DIMENSIONS
175 x 41 x 23 mm
(7.9 x 1.8 x 1")
WEIGHT
78 g (2.7 oz.)
DISTNEW-R3 07/01
ELECTRODE REPLACEMENT:
The electrode is built-in inside the plastic casing and
can be easily replaced in the following way:
• Slide off the battery cover.
• Remove the screw on the back of the DiST
DiST
DiST
DiST
DiST located
below the battery compartment and slide off the
whole internal part of the tester.
• Replace the built-in electrode with a new:
HI73301
HI73301
HI73301
HI73301
HI73301 for DiST1
DiST1
DiST1
DiST1
DiST1
HI73302
HI73302
HI73302
HI73302
HI73302 for DiST2
DiST2
DiST2
DiST2
DiST2
HI73303
HI73303
HI73303
HI73303
HI73303 for DiST3
DiST3
DiST3
DiST3
DiST3
HI73304
HI73304
HI73304
HI73304
HI73304 for DiST4
DiST4
DiST4
DiST4
DiST4
• Before reinserting the internal circuit into the
new plastic casing, make sure the O-rings (2)
of the temperature sensor (3) are properly
seated (see drawing).
To avoid bending the connections to the con-
ductivity sensors (1), do not apply excessive
force while inserting the circuit. In case of any
problem, reslide and reinsert it.
Internal circuit:
• Fasten the screw on the back, below the
battery compartment.
• Reinsert the battery compartment cover.
• Recalibrate the DiST
DiST
DiST
DiST
DiST before performing new
measurements.
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
ONE YEAR WARRANTY AND SERVICEABLE:
Covered by one year
one year
one year
one year
one year warranty against defects in work-
manship and materials, these testers are now completely
serviceable. Contact your dealer for further information.
The DiST
DiST
DiST
DiST
DiST is in compliance with the CE directives.
www.tools.in.th
www.neonics.co.th