Hanna Instruments DiST 2 Instruction Manual Download Page 2

OPERATION

• Remove the protective cap.
• Turn the meter ON sliding the

switch on the top.

• Immerse into solution up to

the max immersion level with-
out touching the bottom of the
beaker.

• Stir gently and wait until the

display stabilizes. Readings are
automatically compensated for
temperature variations.

• The value on the display has

to be multiplied by a factor of
10 for DiST1 & DiST3 and
by a factor of 100 for DiST2
DiST4 (as shown on the
upper left corner of the LCD) to give
direct reading in ppm or µS/cm.

CALIBRATION

• Immerse the tester up to the maximum

immersion level in the appropriate cali-
bration solution (see Specifications).

• Allow the reading to stabilize and using

a small screwdriver adjust the calibra-
tion trimmer to match the solution value.
E.g. calibrating DiST1 with HI70032P
solution, turn the trimmer to read "138"
(x10 = 1380 ppm).

BATTERY REPLACEMENT

When the meter cannot be switched on or
the display fades, pull out
the battery compartment and
replace all four 1.5V batter-
ies, while paying attention
to their polarity.
Batteries should only be re-
placed in a safe area using
the battery type specified in
this instruction manual.

Recommendations for users
Before using these products, make sure that they are
entirely suitable for the environment in which they are
used. Operation of these instruments in residential areas
could cause interference to radio and TV equipment.
The pins at the end of the sensor are sensitive to
electrostatic discharges. Avoid touching this metal band
at all times. Any variation introduced by the user to the
supplied equipment may degrade the instrument's EMC
performance. To avoid electrical shock, do not use this
instrument when voltages at the measurement surface
exceed 24 Vac or 60 Vdc. To avoid damage or burns, do
not perform any measurement in microwave ovens.

WARRANTY
These meters are guaranteed for six months against de-
fects in workmanship and materials when used for their
intended purpose and maintained according to instruc-
tions. This warranty is limited to repair or replacement free
of charge. Damages due to accident, misuse, tampering
or lack of prescribed maintenance are not covered. If
service is required, contact the dealer from whom you
purchased the instrument. If under warranty, report the
model number, date of purchase, serial number and the
nature of the failure. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from
the Customer Service department and then send it with
shipment costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protec-
tion.

Reviews: