Turn Checktemp® ON with 

ON/OFF

 button: all LCD 

segments are displayed for 1 second, and CalCheck® 

starts automatically to verify that Checktemp® 

of “

CAL

“ and “

-0-

“ means that Checktemp® is 

measuring temperature, remember that:

·

The temperature sensor is located at the tip 

of the stainless steel probe.

 

·

It is necessary to wait for a few seconds for 

the display to stabilize.

Liquid Crystal Display

ON/OFF button

Maximum Immersion Level

Temperature Probe

4 cm

1.5”

Battery

Setup  
Button

Recommended 
Immersion Level

Range

-50.0 to 150.0 ºC / -58.0 to 302 ºF

Resolution

0.1 ºC (-50.0 to 150.0 ºC)  
0.1 ºF (-58.0 to 199.9 ºF) / 1 ºF (above 200 ºF)

Accuracy

 

±0.2 ºC (-30.0 to 120.0 ºC) 
±0.3 ºC (outside: -50.0 to -30.0 ºC  
and 120.0 to 150.0 ºC) 
±0.5 ºF (-22.0 to 199.9 ºF)  
±1 ºF (outside: -58.0 to -22.0 ºF  
and 200 to 302 ºF)

Environment

  -30 to 50 ºC (-22 to 122 ºF) 

Battery Type

1 x CR2032

Battery Life

approx. 2000 hours of continuous use

Auto Off

8 min (default), 60 min or OFF

Dimensions

50 x 185 x 21 mm (2 x 7.3 x 0.9“)

Weight

50 g (1.8 oz.)

All rights are reserved. Reproduction in whole or in part is 

prohibited without the written consent of the copyright owner, 

Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

IST98501     

09/14

Warranty

Checktemp® is guaranteed for a period of one year after date of purchase 

against defects in workmanship and materials when used for their intended 

purpose and maintained according to instructions. This warranty is limited 

to repair or replacement free of charge. Damage due to accidents, misuse, 

tampering or lack of prescribed maintenance is not covered. If service is 

required, contact your local Hanna Office. If under warranty, report the model 

number, date of purchase, serial number and the nature of the problem. If 

the repair is not covered by the warranty, you will be notified of the charges 

incurred. If the instrument is to be returned to Hanna Instruments, first 

obtain a Returned Goods Authorization number from the Technical Service 

department and then send it with shipping costs prepaid. When shipping any 

instrument, make sure it is properly packed for complete protection.

Recommendations for Users

Before using Hanna Products, make sure that they are entirely suitable for 

your specific application and for the environment in which they are used.

Operation of these instruments may cause unacceptable interferences to 

other electronic equipment, thus requiring the operator to take all necessary 

steps to correct such interferences. Batteries should only be replaced in a 

safe area using the battery type specified in this instruction manual. Any 

variation introduced by the user to the supplied equipment may degrade the 

instrument’s EMC performance. To avoid damages or burns, do not put the 

instrument in microwave oven. For yours and the instrument safety do not 

use or store the instrument in hazardous environments.

Meter Setup

Enter Setup to change the temperature unit (ºC or 

Setup MODE, press the 

SET

 button located inside 

the battery compartment once for temperature 

·

Use the 

ON/OFF

 button to switch between 

ºC

” or “

ºF

”. Press 

SET

 once to save and continue 

to auto shut off settings.

·

Use the 

ON/OFF

 button to cycle through the 

following Auto Shut Off options: 8 minutes 

(“

d08

”, default value), 60 minutes (“

d60

”), 

OFF (“

d--

”). Press the 

SET

 button once to save 

and continue to normal operation.

Battery Replacement

Replace the battery when the battery symbol is 

blinking or when Checktemp® does not switch 

on. If the meter has been submerged in a liquid, 

make sure it has been completely dried. Check the 

polarity of the battery, open the compartment, and 

replace the battery. Close battery cover.

Note:

Batteries should only be replaced in a safe area using the battery 

type specified in this instruction manual. Old batteries should be disposed in 

accordance with local regulations.

Troubleshooting

Errors that may occur: A message of “

CAL

” and 

Err

” means an internal error has occurred and 

Checktemp® is not guaranteed to be within its 

and the display shows “---” the temperature 

connection to the probe has been broken. If a 

message of “

Err

” occurs then a calibration data 

read error has occurred. A message of “

Er1

” is 

caused by an unexpected device reset. A message 

of “

Er2

” occurs if internally a temperature 

measurement malfunctions. After these messages 

Checktemp® OFF and then ON with the 

ON/OFF

 

button to reset. If the error message continues to 

Certification

EN 13485:2001 - Suitability: Storage;

Environment: E;  Accuracy Class: 0.5; IP65