Hanna Instruments Checker HI772 Instruction Manual Download Page 2

Turn the meter on by pressing the button. All 
segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

Fill the cuvette with 10 mL of unreacted sample and 
replace the cap. Place the cuvette into the meter and 
close the meter’s cap.

Press the button. When the display shows “

Add

”, 

C.2

” with “

Press

” blinking the meter is zeroed.

Remove the cuvette from the meter and unscrew 
the cap. Using a 1 mL syringe carefully add exactly 
1.00 mL of 

HI772S

 Reagent to the sample. Replace 

the cap and gently invert 5 times. Place the cuvette 
back into the meter.

Note: Pay attention not to spill reagent otherwise 

full color development may be inhibited.

Press the button. The instrument directly displays 
the concentration of alkalinity in dKH. Alkalinity 
conversion: 
1 dKH 

= 17.86 ppm CaCO

3

 

= 0.358 meq/L

The meter automatically turns off after 10 minutes.

10 mL

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be 

dry outside, and completely free of fingerprints, oil and dirt. Wipe it 
thoroughly with 

HI731318

 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To 

obtain accurate measurements, remove bubbles by swirling or by 
gently tapping the cuvette.

• Do not let the reacted sample stand for too long after reagent is 

added, as accuracy will be affected.

• After the reading it is important to immediately discard the 

sample, otherwise the glass might become permanently stained.

HI772

To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the 
instrument will display “

bAd

” then ”

bAt

” for 1 second and then turns off.

To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•  Turn the instrument off by holding the button until the meter shuts off.
•  Turn the instrument upside down and remove the battery cover with a screwdriver.
•  Remove the battery from its location and replace it with a new one, inserting the negative 

end first.

•  Insert the battery cover and replace the screw with a screwdriver.

IST772   02 /17

Measurement Procedure

HI772S

Tips for an Accurate Measurement

Battery Management

Reagent sets

HI772-26

Reagent for 25 Marine Alkalinity tests, 1 syringe and 1 tip

Other Accessories

HI772-11

Marine Alkalinity Certified Standard Kit

HI731318

Cloth for wiping cuvettes (4 pcs.)

HI731321

Glass cuvetes (4 pcs.)

HI731225

Black Cuvette Cap for Checker HC (4 pcs.)

HI740028P

1.5V AAA batteries (12 pcs.)

HI740142P

1 mL graduated syringe (10 pcs.)

HI93703-50

Cuvette cleaning solution (230 mL)

Accessories

Hanna Instruments reserves the right to modify the design, construction, or appearance of its products 

without advance notice.

Warranty

HI772

 is warranted for a period of one year after date of purchase against defects in 

workmanship and materials when used for their intended purpose and maintained 
according to instructions. This warranty is limited to repair or replacement free of charge. 
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not 
covered. If service is required, contact your local Hanna Instruments Office. If under 
warranty, report the model number, date of purchase, serial number and the nature of the 
problem. If the repair is not covered by the warranty, you will be notified of the charges 
incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Return 
Goods Authorization number from the Technical Service department and then send it with 
shipping costs prepaid. When shipping any instrument, make sure it is properly packed for 
complete protection.

Recommendations for Users

Before using Hanna Instruments products, make sure that they are entirely suitable for 
your specific application and for the environment in which they are used. Operation of 
these instruments may cause unacceptable interferences to other electronic equipment, 
thus requiring the operator to take all necessary steps to correct such interferences. Any 
variation introduced by the user to the supplied equipment may degrade the instrument’s 
EMC performance. To avoid damages or burns, do not put the instrument in microwave 
oven. For yours and the instrument safety do not use or store the instrument in hazardous 
environments.

Reviews: