•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display shows 

Add

”, “

C.1

” with “

Press

” blinking, the meter is 

ready.

•  Fill the cuvette with 10 mL of unreacted sample and 

replace the cap. Place the cuvette into the meter and 
close the meter’s cap.

•  Press the button. When the display shows “

Add

”, 

C.2

” with “

Press

” blinking the meter is zeroed.

•  Remove the cuvette from the meter and unscrew 

the cap. Using a 1 mL syringe carefully add exactly  
1 mL of 

HI772S

 Reagent to the sample. Replace 

the cap and gently invert 5 times. Place the cuvette 
back into the meter.

Note: Pay attention not to spill reagent otherwise 

full color development may be inhibited.

•  Press the button. The instrument directly displays 

the concentration of alkalinity in dKH. Alkalinity 
conversion: 

 

1 dKH = 17.86 ppm CaCO

3

 

 

= 0.358 meq/L

 

The meter automatically turns off after 10 minutes.

HI772S

TAKING MEASUREMENTS

Note: For the error messages and battery replace-

ment see the instruction manual.