Hanna Instruments C 221 Instruction Manual Download Page 2

POWER CONNECTION

Plug the 12VDC adapter (

HI 710005

 - 110VDC, or 

HI

710006 

- 220VDC) into the DC socket. Plug the adapter

into the outlet.
Alternatively, remove the battery cover on the back of the
meter; attach two fresh 9V batteries and replace the cover.

Note:

Insure the main line is surge protected.

Note: 

Always turn the meter off before unplugging it to

insure no data is lost.

MEASUREMENT PROCEDURE

Turn the meter on by pressing
ON/OFF.

The meter will first perform an
LCD self diagnostic test by
displaying a full set of figures.

• Then it will show a scrolling

"c221 Hanna Inst" message.

When the LCD displays “----”,
the meter is ready.

• To perform calibration, fill one

cuvet with about 4 mL of
Glycerol, up to 5  mm (0.2”)
below the rim. This is the
Glycerol Standard Reference.

• Place the Glycerol cuvet into

the holder. Then put the light
shield cap on.

Operational Guide

There is not enough light to perform a
measurement. Please check the preparation
of the Glycerol cuvet.
There is  too much light to perform a
measurement. Please check the preparation
of the Glycerol cuvet.

b) on sample reading:

Calibration was not performed. Follow the
instruction described in the measurement
procedure for calibrating the meter.

Under range. A blinking "0" indicates that
the sample absorbs less light than the
Glycerol Reference. Check the procedure
and recalibrate the instrument.

1) A flashing value of “150” (maximum
measurable value) indicates an over range
condition. The color intensity of the sample
is beyond the programmed range.
2) A flashing value lower than the maximum
measurable value (e.g. 120) indicates a
low signal-to-noise ratio condition. In this
case accuracy of the result is not guaranteed.
Repeat the reading procedure.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for your specific
application and for the environment in which they are used.
Operation of these instruments may cause  unacceptable interferences to other electronic
equipments, this requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and
the instrument safety do not use or store the instrument in  hazardous environments.

CE Declaration of Conformity

ISTC221R2 03/04

The instructions listed below should be carefully followed
during testing to ensure best accuracy.
• Use matched square cuvets, having 4 clear faces and a

10 mm path length.

• Whenever the cuvet is placed into the measurement cell,

it must be dry outside, and completely free of fingerprints,
oil or dirt. Wipe it thoroughly with 

HI 731318

 or a lint-

free cloth prior to insertion.

• It is important that the sample does not contain any

debris. This would corrupt the readings.

• Samples should be completely free of air bubbles. If

bubbles are present, tap the cuvet on the table to
remove them.

Tips for an Accurate Measurement

• Press CAL and “SIP” will

blink on the display.

• After a few seconds the

display will show “-0.0-”.
The meter is now calibrated
and ready for measurement.

• Remove the Glycerol cuvet.
• Add to a second clean cuvet

about 4 mL of honey, up
to 5 mm (0.2”) below the
rim. This is the sample.

• Place the sample cuvet into

the holder. Then put the
light shield cap on.

• Press READ and “SIP” will

blink during measurement.

• The instrument directly displays the honey color intensity

value expresses in mm Pfund, as compared to Analytical
grade Glycerol (fixed at zero Pfund).

INTERFERENCES

Interference may be caused by air bubbles or turbidity in
the sample. Scratched or dirty cuvets will also affect readings.
Always check clearness of cuvets prior to use.

ANALYSIS KIT

C219 Kit

Kit for Honey Color analysis, including
82 cuvets, 30 mL of Glycerol and two 5
mL syringe (75 tests average)

OTHER ACCESSORIES

C215-00300

5 mL graduated syringe

HI 710005

110VAC to 12VDC voltage adapter

HI 710006

220VAC to 12VDC voltage adapter

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

Accessories

BATTERY REPLACEMENT

Battery replacement must only take place in a non-hazardous
area using two 9V alkaline batteries. Remove the battery
cover on the back of the meter, attach two fresh 9V
batteries and replace the cover while paying attention to
the correct polarity.

USDA Color Standards

USDA Color Standards

USDA Color Standards

USDA Color Standards

USDA Color Standards

Color Range Pfund

Color Range Pfund

Color Range Pfund

Color Range Pfund

Color Range Pfund

Designations

Designations

Designations

Designations

Designations

Scales (mm)

Scales (mm)

Scales (mm)

Scales (mm)

Scales (mm)

Water White

8 or less

Extra White

>

 8 - 

≤ 

17

White

>

 17 - 

≤ 

34

Extra Light Amber

>

 34 - 

≤ 

50

Light Amber

>

 50 - 

≤ 

85

Amber

>

 85 - 

≤ 

114

Dark Amber

>

 114

Honey TypeHoney TypeHoney TypeHoney TypeHoney Type

AVERAGEAVERAGEAVERAGEAVERAGEAVERAGE

SDSDSDSDSD

Min. valueMin. valueMin. valueMin. valueMin. value

Max. ValueMax. ValueMax. ValueMax. ValueMax. Value

common namecommon namecommon namecommon namecommon name

 latin name latin name latin name latin name latin name

(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)

(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)

(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)

(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)(mm Pfund)

Acacia tree

Robinia pseudoacacia

15

6

11

27

Chestnut tree

Castanea sativa

92

19

62

119

Citrus spp.

Citrus spp.

14

5

11

35

Dandelion

Taraxacum officinalis

54

11

41

71

Eucalyptus

Eucalyptus spp.

58

11

41

71

Fir honeydew

98

8

83

110

Fir tree honeydew

99

16

83

130

French honeysuckle

Hedysarium coronarium

18

6

11

35

Heather

Erica arborea

96

10

83

119

Lime tree

Tilia spp.

43

17

11

71

Rhododendron

Rhododendron spp.

13

5

11

27

Strawberry tree

Arbutus unedo

70

10

55

83

Sunflower

Heliantus annus

61

6

51

71

Thyme

Thymus spp.

52

16

27

83

Table 1

Table 2

Reviews: