32
10 mL
SPECIFICATIONS
Range
0.00 to 5.00 mg/L
Resolution
0.01 mg/L
Precision
±0.03 mg/L @ 1.00 mg/L
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 575 nm
Method
Adaptation of the USEPA approved method. The
reaction between free copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity/test
HI 93702-0
Bicinchoninate
1 packet
REAGENT SETS
HI 93702-01
Reagents for 100 tests
HI 93702-03
Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Free Copper on the
secondary LCD by pressing PROGRAM
▼
and
▲
.
• Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
FREE COPPER
Free Copper