Hanna Instruments 96728 Instruction Manual Download Page 21

21

• Press 

READ/ /TIMER

 to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

• Press 

ZERO/CFM

 to restore the factory

calibration or press 

GLP/

 again to

abort factory calibration restore.

• The instrument briefly indicates 

“donE”

upon restoration of factory calibration prior
to returning to 

measurement mode.

BATTERY MANAGEMENT

To save battery, the instrument shuts down after 10 minutes of non-use in 

measurement mode

and after 1 hour of non-use in 

calibration mode.

If a valid measurement was displayed before
auto-shut off, the value is displayed when
the instrument is switched on. The blinking

“ZERO”

 means that a new zero has to be

performed.
One fresh battery lasts for around 750 measurements, depending on the light level.
The remaining battery capacity is evaluated at the instrument startup and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken anymore, the instrument
shows “

dEAd bAtt

” and turns off.

To restart the instrument, the battery must be replaced with a fresh one.

Summary of Contents for 96728

Page 1: ...1 Instruction Manual HI 96728 Nitrate ISM...

Page 2: ...RY MANAGEMENT 21 BATTERY REPLACEMENT 22 ACCESSORIES 22 WARRANTY 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use...

Page 3: ...ing shipment please notify your Dealer Each HI 96728 Ion Selective Meter is supplied complete with Two Sample Cuvettes and Caps 9V Battery Instrument quality certificate Instruction Manual Note Save a...

Page 4: ...m dust or dirt by a transparent cup This makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off t...

Page 5: ...se in calibration mode with last reading reminder Dimensions 192 x 104 x 69 mm 7 6 x 4 1 x 2 7 Weight 360 g 12 7 oz REQUIRED REAGENTS Code Description Quantity test HI 93728 0 Powder Reagent 1 packet...

Page 6: ...ion coefficient at wavelength c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the ot...

Page 7: ...coming value into the desired measuring unit and to display it on the LCD The measurement process is carried out in two phases first the meter is zeroed and then the actual measurement is performed Th...

Page 8: ...de press to edit the date and time OPERATING MODES Measurement mode default operation mode enables both validation and measurement Calibration mode may be entered by keeping CAL CHECK pressed for thre...

Page 9: ...ht to perform a measurement Please check the preparation of the zero cuvette No Light The instrument cannot adjust the light level Please check that the sample does not contain any debris Inverted cuv...

Page 10: ...igh The standard reading is higher than expected Cap error Appears when external light enters in the analysis cell Assure that the cuvette cap is present Cooling lamp The instrument waits for the lamp...

Page 11: ...ASUREMENT It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is...

Page 12: ...trument by removing the dust protection sleeve from the instrument cuvette holder Place the battery in the instrument as described in the BATTERY REPLACEMENT chapter Place the instrument on a flat tab...

Page 13: ...p and immediately shake vigorously for exactly 10 seconds by moving the cuvette up and down Continue to mix by inverting the cuvette gently and slowly for 50 seconds while taking care not to induce ai...

Page 14: ...erference may be caused by Ammonia and amines as urea and primary aliphatic amines Chloride above 100 mg L negative interference Chlorine above 2 mg L positive interference Copper it must be absent Ir...

Page 15: ...eroed first Place the CAL CHECKTM Standard HI 96728 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO CFM and the lamp cuvette...

Page 16: ...ent the display will show the validation standard value The reading should be within specifications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications...

Page 17: ...isplays dashes the meter is ready Press and hold CAL CHECK for three seconds to enter calibration mode The display will show CAL during calibration procedure The blinking ZERO asks for instrument zero...

Page 18: ...display depending on the measurement phase After measurement the instrument will show for three seconds the CAL CHECKTM Standard value Note If the display shows STD HIGH the standard value was too hi...

Page 19: ...ed the month number is automatically increased When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press GLP to edit the desired day...

Page 20: ...ppear on the main display and the instrument returns to measurement mode after three seconds It is possible to delete the calibration and restore factory calibration Press GLP to enter GLP mode In GLP...

Page 21: ...easurement was displayed before auto shut off the value is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 7...

Page 22: ...it clockwise to close To replace the instrument s battery follow the steps ACCESORIES REAGENT SET HI93728 01 Reagentsfor100tests HI93728 03 Reagentsfor300tests OTHER ACCESORIES HI96728 11 CAL CHECKTM...

Page 23: ...tained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If serv...

Page 24: ...24 MAN96728 01 11 PrintedinEUROPE ROMANIA Local Sales and Customer Service Office...

Reviews: