Quick Guide

Supra R17 

Cat. No. : SU-R17

A2.0-36   
A10-12  
A15-12        
A15c-12
A50-6  

A50c-6          
A50-8     
A85-6     
A250-6 

36 x 1.5/2.0 ml / Max. 15k 

12 x 10 ml / Max. 17k 

12x 15 ml  / Max. 17k 

12 x 15 ml conical / Max. 15k  

6 x 50 ml / Max. 17k  

6 x 50 ml conical / Max. 15k  

8 x 50 ml / Max. 15k 

6 x 85 ml / Max. 15k  

6 x 250 ml / Max. 10k

If the centrifuge lid cannot be opened, insert 
the provided T-wrench into the emergency 
release hole in the front of the device  and turn it 
counterclockwise. (Before activating emergency 
open, switch off the device and wait for the
rotor to stop.)

Remove any dust from the ventila-
tion slits on the back side of the 
device.

Empty the condensation water 
regularly by removing the drain cap 
on the left side of the device.

Allow a safety clearance of 30 cm 
around the centrifuge during
operation.

30 cm

30 cm

30 cm

After relocating the device or loading the rotor, 
balance them by using 

a leveler. 

If neces-

sary, level the device by adjusting the ‘Height 
adjustment wheel’ on the bottom of the device.

Rev.0

Proudly made by Jiyoung Shim, Hyewon Jeong 

Order & Inquiry l +82-2-3452-9996 l [email protected] l www.ihanil.com
Service  l +82-2-3452-8966 l [email protected]

Connect the device to voltage souces, which 
correspond to the power requirements on the 
label attached to the device.

Check the wall outlet is earth-
grounded, surge- protected and 
can supply enough power.

Check the Line Voltage

Grounded Outlet

Switch Off the Device

Switch off the device after use.
Unplug the power plug before cleaning or unused 
for a long period of time.

2.5KVA
220~240V 50/60Hz 

Rotors

Maintenance

Electrical Safety

Cautions for Use

Inspect the chamber, rotor & accessories for 
foreign objects.

Wipe up the moisture in the chamber with 
a dry cloth after use. Keep the lid open 
when not in use.

Clean the chamber and rotor by wiping 
with a cloth soaked in mild detergent and
dry completely.

Keep the Chamber Dry

Cleaning

Load tubes symmetrically into the rotor with 
identical amount.

Load Tubes Symmetrically

Check the rotor,buckets and tubes are free of 
scratch and crack before operation.

Check the Accessories for Abnormalities

Balancing the Centrifuge and Rotor 

Before Operation

After Operation

Remove Dust

Empty the Condensation 
Water

Allow a Safety Clearance 
of 30 cm

Emergency Lid Open

Check the Chamber & Accessories