Quick Guide
Supra R17
Cat. No. : SU-R17
A2.0-36
A10-12
A15-12
A15c-12
A50-6
A50c-6
A50-8
A85-6
A250-6
36 x 1.5/2.0 ml / Max. 15k
12 x 10 ml / Max. 17k
12x 15 ml / Max. 17k
12 x 15 ml conical / Max. 15k
6 x 50 ml / Max. 17k
6 x 50 ml conical / Max. 15k
8 x 50 ml / Max. 15k
6 x 85 ml / Max. 15k
6 x 250 ml / Max. 10k
If the centrifuge lid cannot be opened, insert
the provided T-wrench into the emergency
release hole in the front of the device and turn it
counterclockwise. (Before activating emergency
open, switch off the device and wait for the
rotor to stop.)
Remove any dust from the ventila-
tion slits on the back side of the
device.
Empty the condensation water
regularly by removing the drain cap
on the left side of the device.
Allow a safety clearance of 30 cm
around the centrifuge during
operation.
30 cm
30 cm
30 cm
After relocating the device or loading the rotor,
balance them by using
a leveler.
If neces-
sary, level the device by adjusting the ‘Height
adjustment wheel’ on the bottom of the device.
Rev.0
Proudly made by Jiyoung Shim, Hyewon Jeong
Order & Inquiry l +82-2-3452-9996 l [email protected] l www.ihanil.com
Service l +82-2-3452-8966 l [email protected]
Connect the device to voltage souces, which
correspond to the power requirements on the
label attached to the device.
Check the wall outlet is earth-
grounded, surge- protected and
can supply enough power.
Check the Line Voltage
Grounded Outlet
Switch Off the Device
Switch off the device after use.
Unplug the power plug before cleaning or unused
for a long period of time.
2.5KVA
220~240V 50/60Hz
Rotors
Maintenance
Electrical Safety
Cautions for Use
Inspect the chamber, rotor & accessories for
foreign objects.
Wipe up the moisture in the chamber with
a dry cloth after use. Keep the lid open
when not in use.
Clean the chamber and rotor by wiping
with a cloth soaked in mild detergent and
dry completely.
Keep the Chamber Dry
Cleaning
Load tubes symmetrically into the rotor with
identical amount.
Load Tubes Symmetrically
Check the rotor,buckets and tubes are free of
scratch and crack before operation.
Check the Accessories for Abnormalities
Balancing the Centrifuge and Rotor
Before Operation
After Operation
Remove Dust
Empty the Condensation
Water
Allow a Safety Clearance
of 30 cm
Emergency Lid Open
Check the Chamber & Accessories