If the centrifug lid cannot be opened, you can activte the emergency open manually.
1. Turn off the power switch and wait for rotor to stop before activating the emergency open.
2.
Insert the provided T-wrench into the emergency release hole and turn it clockwise
until the lid is released.
6. Troubleshooting
6.2 Error codes
21
www.ihanil.com