Operation Manual

FLETA 4

Clinical Centrifuge

Date of Purchase

Serial No.

Place of purchase

Summary of Contents for FLETA 4

Page 1: ...Operation Manual FLETA 4 Clinical Centrifuge Date of Purchase Serial No Place of purchase...

Page 2: ...of purchase 82 2 3472 0727 Inquiry info ihanil com Order sales ihanil com Tech support techsupport ihanil com The contents in this operating manual are subject to change for device improvement Copyri...

Page 3: ...el 9 4 2 Opening and closing the lid 10 4 3 Loading Unloading a rotor 11 4 4 Loading tubes 12 4 6 Setting RPM RCF 13 1 4 Electric safety information 4 5 Maintenance 6 Troubleshooting 6 1 General error...

Page 4: ...k of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids strong bases cesium salt or alkaline detergents 8 If the centrifuge is...

Page 5: ...1 4 Electric safety information 1 It is recommended that switchgear or circuit breakers and overcurrent protection devices be installed near the equipment 2 Use a power cord only provided with equipm...

Page 6: ...otor Display Control panel The display shows time rpm temperature and etc Power switch Power socket On off the centrifuge Emergency Release Hole If the centrifuge lid cannot be opened you can open the...

Page 7: ...ml Fixed angle Time control 100 min pulse continuous ACC DEC ramps 5 6 steps Program memory 10 Noise level 60 dB Imbalance cutoff Yes Rotor identification Manual Dimension W x D x H mm 448 5 x 531 x 2...

Page 8: ...ent due to the weight of the rotor Goodventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Ce...

Page 9: ...quirements on the label attached to the device Check the wall outlet is earth grounded surge protected and can supply enough power 1 Connect the DC plug at the power socket on the right side of the de...

Page 10: ...et the Values Start Stop Short Start Stop the centrifugation Press more than 2sec short spin 9 www ihanil com UP Down Up or down the values Move Left Right Select a parameter Lid Lid open Program Pres...

Page 11: ...d press the edge of the lid The lid can only be opened if the centrifuge is switched on Do not reach with your fingers between the housing and lid Close the lid completely before operation 4 Operation...

Page 12: ...enter hole of the rotor and turn it clockwise to fasten the rotor bolt Check the rotor to make sure it loaded firmly Unloading To remove the rotor insert the wrench into the center hole of the rotor a...

Page 13: ...ed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales Check the maximum load for each tube Put tubes into rotor hole...

Page 14: ...setting mode 00 flashes on display Sec setting mode 00 flashes on display 2 Press the Left or Right button to select Time mode 3 Press the up or down buttons until the desired run time is displayed 4...

Page 15: ...Press the up or down buttons until the desired Accel Decel step is displayed 4 Press the Setup Enter button 4 9 Start Stop Short Spin Starting a centrifugation 1 Check the all parameters are correct...

Page 16: ...ou can store 10 programs on the device 0 9 Select the program number by pressing the Up or Down button Calling up a program Press the Program button twice Select the program number by pressing the Up...

Page 17: ...o not use aggressive chemicals on the device such as alchol benzene acetone or phenol If the device is contaminated use a mild cleaning fluid to clean Make sure do not scratch the surface of equipment...

Page 18: ...5 Maintenance 5 3 Disposal In case of product is to be disposed of the local wastes laws and regulations are to be observed 17 www ihanil com...

Page 19: ...t be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Checkthe lid latch is not damaged Unusual noise and vibration Check the device whether it...

Page 20: ...heat Shut off the power supply and then turn on the power switch again to check the device E4 Motor operation error Motor operation error Check that the rotor is rotating through the RPM measuring win...

Page 21: ...device E13 Low voltage error Low Voltage Contact Hanil Scientific Inc E15 RPM Sensor error RPM Sensor error Rotor identification error Check the id disc is on the bottom of the rotor E17 Communicati o...

Page 22: ...open manually 1 Turn off the power switch and wait for rotor to stop before activating the emergency open 2 Insert the provided T wrench into the emergency release hole and turn it clockwise until the...

Page 23: ...7 Rotors A S 82 2 3452 8966 techsupport ihanil com 22 Rotor Id 4 Rotor Id 3...

Page 24: ...7 Rotors 23 www ihanil com Rotor Id 1 Rotor Id 2...

Page 25: ...memo...

Page 26: ...memo...

Page 27: ...memo...

Page 28: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of KOREA T 02 3452 8965 info ihanil com www ihanil com Doc No OMF4PDEN1801...