background image

Operating Manual

Component R6

Large capacity Centrifuge

Date of Purchase

Serial No.

Place of purchase

Summary of Contents for Component R6

Page 1: ...Operating Manual Component R6 Large capacity Centrifuge Date of Purchase Serial No Place of purchase ...

Page 2: ...il ScientificInc All rights reserved Contact Us If you have and questions contact Hanil ScientificInc or place of purchase 82 2 3472 0727 Inquiry info ihanil com Order sales ihanil com Tech support techsupport ihanil com ...

Page 3: ...on panel 9 4 2 Opening and closing the lid 10 4 3 Loading tubes 10 4 4 Setting RPM RCF 11 4 5 Setting run time 11 1 4 Electric safety information 4 5 Maintenance 6 Troubleshooting 6 1 General errors 14 3 5 7 9 3 2 2 Power connetion 8 4 7 Start the centrifugation run 11 4 8 End the centrifugation 11 4 9 Saving Recall a program 12 13 5 1 Care instructions 13 5 2 Cleaning 13 14 4 6 Setting Accelation...

Page 4: ...ork of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids strong bases cesium salt or alkaline detergents 8 If the centrifuge is contaminated by toxic or radioactive samples or infectiousblood samples remove contaminantscompletely and take needful actions such as ventilation or isolation of centrifuge 9 Before operation rotor...

Page 5: ...r of electric shock 1 4 Electric safety information 1 It is recommended that switchgearor circuit breakers and overcurrent protectiondevices be installed near the equipment 2 Use a power cord only provided with equipment 3 Use sockets with a protective earth conductorand suitable power cord 4 Do not use an extension cord 5 Do not place anything on the powercable 6 Do not block vents 7 If you have ...

Page 6: ...er Optional items Rotors Power switch On off the centrifuge RPM measuring window Visual inspection for rotor stop or for a speed check using a tachometer 5 l www ihanil com Lid Chamber Where the rotor is loaded Display Control panel The display shows time rpm temperature Circuit breaker It prevents overvoltage or short circuit ...

Page 7: ...es Temperature range Temperatureoffset 20 to 40 Yes Time control 100 hr continuous Program memory 100 Noise level 65 dB Imbalance cutoff tracking Yes Yes Rotor identification Automatic Dimension W x D x H mm 816 x 906 x 965 Weight without rotor 355 kg Power requirement 5 0 kVA Power Consumption 20A Power input V Hz 220 240V 50 60 Hz Refrigerantoil Coolant Refrigerant POE ISO VG 32 404A R134A Cat N...

Page 8: ... bent due to the weight of the rotor Goodventailation For air circulation and safety maintain a 30cm clearance envelope aroundthe centrifugewhile the rotor is spinning Constant temperature humidity Centrifuge equipped with the sensitive electronic software which is fragile with humidity and temperature Must avoid direct ray or heater and be put in the ambience of controlled temp and air Avoid the ...

Page 9: ...ts on the label attached to the device Use sockets with a protective earth conductorand provided powercord 1 Connect the power cable to the power socket on the right side of the product and plug the cord into the outlet 2 Switch the centrifuge on using the power switch on the right side of the device sales ihanil com techsupport ihanil com l 8 ...

Page 10: ...mperature when centrifugation with sensitive samples Start Start centrifugation Stop Stop centrifugation Open Open centrifuge lid Clear Erase the wrong number ENTER To set values Prog Calling up a program Save Save a program Accel Set acceleration rates in 10 steps Decel Set acceleration rates in 10 steps Rotor Check the rotor information 9 l www ihanil com ...

Page 11: ...e lid 4 3 Loading tubes Only use tubes provided or approved by Hanil Scientific Inc Always use the same type of tube Tubes should be loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales Check the maximum load for each tube Put tubes into rotor holes Tubes located opposite each other...

Page 12: ...ess the Time button Press the numeric key to set centrifugationtime 100hr Press the Enter button to fix the value Press the Accel key or Decel key Press the Numeric keys to select the Acc Dec step 0 9steps The higher the setting step the faster the accelation or decelerationspeed Decel step 0 means natural brake Press the Enter button to fix the value 4 7 Start the centrifugation run Check the all...

Page 13: ...e button You can store 100 programs on the device 0 99 Select the program slot by pressing the numeric keys Recall a programs Press the Call button Press the program slot by pressing the numeric keys Press the Enter button to call up a program Press Accel button to save or decel button to cancel sales ihanil com techsupport ihanil com l 12 ...

Page 14: ...e such as alchol benzene acetone or phenol If the device is contaminated use a mild cleaning fluid to clean Make sure do not scratch the surface of equipment when cleaning it Do not use a metal sponge If the device is rusted remove it with a mild detergent and wipe it with a dry cloth Chamber If the rotor chamber is not dry wipe moisture from the chamber with a dry cloth Clean the chamber motor sh...

Page 15: ...ot be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Checkthe lid latch is not damaged Unusual noise and vibration Check the device whether it is installed on the hard and flat place Reload the rotor symmetrically Reload the tubes symmetrically Tighten the clamping of the rotor with wrench by turning clockwise sales ihanil com techsupport ...

Page 16: ... lower than the set temperature Set the temperature lower The lid is not open None Blackout Emergency lid open The lid is not closed To close the door No working of lid pin No power connection Check the power connection Close the lid On acceleration the device is vibrating Imbalance Error Rotor is loaded asymmetrically Not tighten the rotor with shaft Load the rotor symmetrically Tighten the rotor...

Page 17: ...be opened you can activted the emergency open manually 1 Wait for rotor to stop before activating the emergency open 2 Insert the provided T wrenchinto the emergency release hole and turn it clockwise until the lid is released sales ihanil com techsupport ihanil com l 16 ...

Page 18: ...7 Rotor 7 1 Swing out rotor 17 l www ihanil com ...

Page 19: ...MEMO ...

Page 20: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of KOREA T 02 3452 8965 info ihanil com www ihanil com ...