MEGAPULSE
®
II USER MANUAL
9
COPYRIGHT © 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Controls and Functions
FIGURE 1
–
Megapulse
®
II
- Front Panel
The letter associated with each control or indicator in this section corresponds to the labeling in FIGURE 1.
A.
Power AC on/off switch. Two position rocker, up for ON, down for OFF. The switch is illuminated when
the power is ON.
B.
Membrane push button mode switches to select various thermal modes. The mode can only be changed
when output is off. The selected mode is indicated by the illuminated LED.
C.
Timer display, indicates the TREATMENT TIME selected, or with the treatment activated, the TREATMENT
TIME remaining.
D.
Timer selector switches, used to select the TREATMENT TIME. Maximum time selectable is 30 minutes.
E.
Output Display LED indicators to show that the output is ON or OFF.
F.
Output switch, changes the output ON and OFF as indicated by the output display LED.
G.
Pulse duration switches, select the pulse duration in microseconds as displayed in the PULSE DURATION
display. Pressing the “Up Arrow” increases the pulse duration to the next higher setting, pressing the “Down
Arrow” reduces it to the next lower setting. Does not function if 1
T, 2
T, 4
T is selected.
H.
Pulse duration display indicates the PULSE DURATION in microseconds (
sec).
I.
Pulse rate switches select the number of pulses per second as displayed in the PULSE RATE display. Pressing
the “Up Arrow” increases the number of pulses per second to the next higher setting, pressing the “Down
Arrow” reduces it to the next lower setting. If 1
T, 2
T, 4
T is selected, this function is disabled.
J.
Pulse rate display indicates the number of pulses per second (PPS).
K.
Symbol IEC 878-03-04 Non-ionizing radiation.
L.
Symbol IEC 348 Attention, Consult Service.
M.
Symbol IEC 878-02-03 Indicates Type BF Equipment.
N.
Symbol IEC 878-03-01 Indicates that high voltage is generated within the drum inductive applicator.
A
B
B
C
D
H
J
E
F
G
I
K
L
M
N
20:00
100
800
2
3/4
1/5