Controls and Functions 

 
 
 
 

 

 
 
 
 
 
 
 
 
 
 

 

 

 
 

FIGURE 1

 – 

Megapulse

®

 II

 

- Front Panel

 

 
The letter associated with each control or indicator in this section corresponds to the labeling in FIGURE 1. 
 
A.

 

Power AC on/off switch.       Two position rocker, up for ON, down for OFF. The switch is illuminated when 
the power is ON. 

B.

 

Membrane  push  button  mode  switches  to  select  various  thermal  modes.              The  mode  can  only  be  changed 
when output is off. The selected mode is indicated by the illuminated LED. 

C.

 

Timer display, indicates the TREATMENT TIME selected, or with the treatment activated, the TREATMENT 
TIME remaining. 

D.

 

Timer selector switches, used to select the TREATMENT TIME. Maximum time selectable is 30 minutes. 

E.

 

Output Display LED indicators to show that the output is ON or OFF. 

F.

 

Output switch, changes the output ON and OFF        as indicated by the output display LED. 

G.

 

Pulse  duration  switches,  select  the  pulse  duration  in  microseconds  as  displayed  in  the  PULSE  DURATION 
display.  Pressing  the  “Up  Arrow”  increases  the  pulse  duration  to  the  next  higher  setting,  pressing  the  “Down 
Arrow” reduces it to the next lower setting. Does not function if 1

T, 2

T, 4

T is selected. 

H.

 

Pulse duration display indicates the PULSE DURATION in microseconds (

sec). 

I.

 

Pulse rate switches select the number of pulses per second as displayed in the PULSE RATE  display. Pressing 
the  “Up  Arrow”  increases  the  number  of  pulses  per  second  to  the  next  higher  setting,  pressing  the  “Down 
Arrow” reduces it to the next lower setting. If 1

T, 2

T, 4

T is selected, this function is disabled. 

J.

 

Pulse rate display indicates the number of pulses per second (PPS). 

K.

 

Symbol IEC 878-03-04 Non-ionizing radiation. 

L.

 

Symbol IEC 348 Attention, Consult Service. 

M.

 

Symbol IEC 878-02-03 Indicates Type BF Equipment. 

N.

 

Symbol IEC 878-03-01 Indicates that high voltage is generated within the drum inductive applicator. 

 

 

J

 

 L 

20:00 

 

100 

 

800 

 

 

 

3/4 

 

1/5 

Summary of Contents for ACP MEGAPULSE II

Page 1: ...by reference in its entirety This Manual may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual wit...

Page 2: ...Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR and SYNCHRONY represent the most recent worldwide advances available for therapeutic application of electromedical devices and othe...

Page 3: ...ts Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accidental c...

Page 4: ...olation 12 Cleaning Disinfecting of the Megapulse II 12 Cleaning and Low Level Disinfection 12 Intermediate Level Disinfection 13 Use of Barriers 13 MODES OF OPERATION 14 Thermal Effect Variables 14 P...

Page 5: ...healthcare professional Indications Diathermy is used therapeutically to increase the temperature of tissues It is generally accepted that heat produces the following desirable therapeutic effects Rel...

Page 6: ...ad or any implanted system that may or may not contain a lead Both the thermal and sub thermal modes of operation pose a risk of tissue destruction o If you are a physician who implants or monitors pa...

Page 7: ...y over the cranial region as little is known about the electromagnetic effects on the cerebrum to determine if it represents a serious hazard when shortwave diathermy is applied to the head Do not app...

Page 8: ...le and a mattress or couch without metal parts e g decorative buttons springs near the patient applicators or cables Do not use conductive mattresses or mattress covers During application of shortwave...

Page 9: ...time and if there is evidence of or if the patient complains of pain or overheating during treatment adjust the output downward until it is tolerated by the patient If the patient continues to complai...

Page 10: ...ether understood it is well established that this energy causes vibration of ions and molecules within the cells and tissues Depending on the intensity of the high frequency energy applied this vibrat...

Page 11: ...indicated by the output display LED G Pulse duration switches select the pulse duration in microseconds as displayed in the PULSE DURATION display Pressing the Up Arrow increases the pulse duration to...

Page 12: ...commonly used pulsed shortwave electrode Drum inductive applicators are generally referred to as monodes Drum inductive applicators are operated in the pulsed mode to avoid overheating of the coils Hi...

Page 13: ...a depth of 2 inches 5cm Thermal Dosage Pulsed shortwave diathermy provides outputs capable of producing thermal effects Average power output in watts is controlled by the peak power and the duty facto...

Page 14: ...nfecting of the Megapulse II Modality equipment shall be cleaned disinfected per facility infection control policy ACP recommends the following guidelines Proper cleaning procedures at the end of each...

Page 15: ...ed to cover any surface that may be touched during a patient treatment to help prevent cross contamination Barrier film is for single use only The film is discarded after each patient treatment The pr...

Page 16: ...e field applicators Typical values for the Megapulse II drum inductive applicator is 1 5 SAR The higher the SAR value the greater the thermal effect in the tissue III Treatment Time In general treatme...

Page 17: ...ation following stretch or mobilization making PSWD an effective modality for treatment of scar tissue and contracture Summary of Physiologic Effects 1 Vasodilatation of blood vessels 2 Increase in ce...

Page 18: ...show 800 PPS If the OUTPUT ON OFF CONTROL 6 switch is pressed with no time set on the display the buzzer will sound and the output will not be energized 3 Audible confirmation of the operation of the...

Page 19: ...t contact 5 When the drum inductive applicator has been placed in position tighten the hand wheels on the arm to prevent movement Switch treatment ON 6 The protocol LED 3 on the front panel will light...

Page 20: ...press gently on the drum applicator pin to dislodge the drum applicator from the articulating arm The drum applicator can now be applied to a patient where the articulating arm might have obstructed t...

Page 21: ...en unit is ON and treatment running drum inductive applicator is not emitting output Patient has impaired sensation over treatment area Use output tester if available to determine if unit has failed o...

Page 22: ...to May of 2004 the default pulse rate setting is 100 PPS Pulse Duration 400 200 100 65 40 20 s available Pulse Repetition Rate 50 100 200 400 600 800 PPS available Timer The digital treatment timer in...

Page 23: ...ic Representation of Pulse Trains 20 MICROSECOND Sec PULSES 800 PULSES PER SECOND PPS 65 Sec PULSES 400 PPS 200 Sec PULSES 200 PPS 400 Sec PULSES 100 PPS CONTINUOUS TRAIN OF PULSES PULSES CONTAIN 27 1...

Page 24: ...R IN WATTS W EFFECT 400 Sec 3 0W 6 0W 12 0W 24 0W 36 0W 48 0W 4o C 200 Sec 1 5W 3 0W 6 0W 12 0W 18 0W 24 0W 2o C 100 Sec 75W 1 5W 3 0W 6 0W 9 0W 12 0W 1o C 65 Sec 49W 1 0W 2 0W 3 9W 5 9W 7 8W SUBTHERM...

Page 25: ...29000 Megapulse Power Cord Hospital Grade 10ft 72763 Applicator Head Coaxial Cable 89555 Megapulse II User Manual Infection Control Supplies ITEM ITEM NO DESCRIPTION 55536 Super Sani Cloth Wipes Sing...

Page 26: ...f any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tamper...

Page 27: ...nty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessories exce...

Page 28: ......

Reviews: