Trip Text
Trip
code
Description
M2 Motor Error
3C00
The control system detects that this motor has
become disconnected.
•
Check the motor, cables and the connections to
the control system.
If this does not solve the error, the power module
may be defect.
Inhibit Active
1E01
1E09
1E0A
One of the inhibit input is active and in a latched
state. The active inhibit is indicated by the last 2
digits of the trip code.
•
Turn the power off and on. This will drop out of
the Latched Mode which might clear the trip.
•
Check all cables and switches that are
connected to the active inhibit.
If this does not solve the error, the ISM may be
defect.
Joystick Calibration
Error
-
The Calibration of the joystick has not been
successful.
•
Enter OBP and re-try to calibrate.
If this does not solve the error, the joystick module
may be defect.
Memory Error
-
This is a non-specific memory error which can be
caused by any of the modules within the system.
•
Check all cables and connections.
•
Turn the power off and on.
If the error is still present and the system contains
modules not supplies by PGDT:
•
Disconnect all non-PGDT modules and turn the
power off and on.
If this has solved the error:
•
Connect each non-PGDT module in turn and
turn the power off and on each time to see
which module causes the problem.
If this does not solve the error, the power module
may be defect.
PM Memory Error
-
This is a specific error for the Power Module
•
Check all cables and connections.
•
Re-program the control system. Use the R-Net
PC programmer to program the most current
specific program for wheelchair or the
manufacturers original programming file.
If this does not solve the error, the power module
may be defect.
Troubleshooting
Service manual - Luca XL
33