Fault 

Recommended Solution 

The red indicator light on 

the front side of the lift 

turns red or a loud alarm 

sound is heard when an 

individual is raised. 

The batteries are low and require charging. Refer to 

Charge the Lift

” 

on page 5. Charge the lift for at least one hour and then try to raise 

or lower the carry bar. 
If this does not correct the problem, contact your local authorized 

dealer immediately to check the lift and ensure proper continued 

operation. 

One side of the lift strap is 

starting to fray after 

continued use. 

Discontinue using immediately and replace strap. Check to be sure 

that the lift is always directly above the individual being lifted, 

especially with power traversing lifts. Refer to 

Transfer an 

Individual

 on pag

11 for correct lift positioning. If fraying still 

continues, contact your local authorized dealer immediately to 

check the lift and ensure proper continued operation. 

The lift does not pass 

through a track 

component such as a 

turntable or transistion 

gate. 

Refer to the Owner

s Manual for the specific piece of equipment in 

question. If the recommended solution does not correct the 

problem, contact your local authorized dealer immediately to check 

the lift and ensure proper continued operation. 

No Power 

Ensure the Emergency Stop Switch is set to ON. 

Lift starts beeping while 

pushing Emergency Down 

button with lift not 

coming down 

The AP-450 lift is equipped with dual sensors for strap tension and 

position, the Emergency Down Button/Switch system only works 

with load on the strap. Emergency Down Button/Switch motion 

stops immediately after the lift strap is loose. 

The lift is recommending 

preventive maintenance. 

Complete the “Semi

-

Annually” maintenance procedure on pag

18 

prior to resetting the lift counter. Instructions to reset the lift counter 

are detailed in the AP-450 Service/Spare Parts Manual. 

General Inspection and Preventive Maintenance 

Do not perform any inspection or maintenance while the lift is in use with a 

patient. 

Each Use 

(To be completed by User) 

Prior to each use the AP-450 lift and associated track, accessories, and slings must be visually 

inspected. Refer to the accessory and sling user guides for specific details regarding their 

inspection. 

Should any of these items fail the inspection do not use the lift. Contact your 
local authorized dealer for service. 

Visually check for the following: 

The lift strap shows NO signs of fraying or breaking along its entire length.

The stitching on the lift strap where it connects to the carabiner shows no signs of fraying

or breaking.

Summary of Contents for AP-300

Page 1: ...AP 450 Handicare Owner s Manual English SWL 450 lb 204 5 kg Manual 753160 Rev D 1...

Page 2: ...ack 11 Transfer an Individual 11 Hand Control Information and Diagnostics 14 Emergency Stop 14 Emergency Lowering 15 Clean and Disinfect the Lift System 15 Contraindications 15 Troubleshooting 16 Trou...

Page 3: ...ntain optimum function the AP 450 should be inspected and maintained on a regular basis See the section titled General Inspection and Maintenance See page 17 Any accessories used with the AP 450 inclu...

Page 4: ...osition mobility impaired individuals safely with dignity and in comfort while minimizing both strain and risk to the caregiver The ceiling lift also does not take up valuable floor space as most trad...

Page 5: ...n charge when not in operation and at the end of each day which will maximize the life cycle of the batteries The lift may remain connected to the charger indefinitely since the charger has a built in...

Page 6: ...d a part of the AP 450 system Use only the charger that was supplied with the lift Use of any other charger will void all warranties and may cause damage to the lift Do not position the lift where it...

Page 7: ...s on That the Hand Control connector is properly connected That the Hand Control does not display Slack Detected If there are still problems with the lift contact your local Handicare authorized deale...

Page 8: ...the Lift The lift automatically shuts off after approximately 3 minutes of non use In an emergency the lift can also be shut off by switching the Emergency Stop Switch to Off See Emergency Stop page 1...

Page 9: ...ed with a tilt sensor that activates once the lift tilts past a certain angle restricting upward movement until the lift is adjusted Tilt should be avoided as much as possible during the lifting trans...

Page 10: ...the lift press the Up button 4 To lower the lift press the Down button Mounting the Lift to the Track The AP 450 lift is designed to be easily mounted and removed from the trolley There are four poss...

Page 11: ...d See Figure 8 for the various connection types Manual Traversing Lift 1 Lower the lift to a comfortable height so you can hold it easily 2 Grab hold of the lift and move it along the track by gently...

Page 12: ...are generally attached to the corresponding hooks on the lift 10 Double check to ensure o The straps are properly attached to the hooks o The individual being lifted is properly positioned in the sli...

Page 13: ...rol are raised up away from reach of children toddlers or anyone not intended to use the product Psychological mental and emotional state assessment is key to safe use of a ceiling lift ensure that us...

Page 14: ...is for service and technical support of the lift 6 On the menu you want to review press the key The selected menu displays 7 If you select the wrong menu at the bottom of the display Select EXIT GO BA...

Page 15: ...cals or liquids should be used to clean and disinfect this lift Ensure that no liquids get inside the lift unit This lift unit is not watertight or drip proof Failure to protect the lift from liquids...

Page 16: ...Ensure the Hand Control charger is not connected to the lift the lift will not move when Hand Control charger is connected If this does not correct the problem contact your local authorized dealer imm...

Page 17: ...iately to check the lift and ensure proper continued operation No Power Ensure the Emergency Stop Switch is set to ON Lift starts beeping while pushing Emergency Down button with lift not coming down...

Page 18: ...ompleted every 6 months It is recommended that this service history be documented Should any of these items fail the inspection do not use the lift Contact your local authorized dealer for service Che...

Page 19: ...full load 450lb 204kg Annual The AP450 will operate as intended throughout its service life provided all recommended maintenance and usage recommendations are followed 1 Load test at full load 450lb 2...

Page 20: ...e frequency See General Inspection and Preventive Maintenance on page 17 The AP 450 should have visual Error status display Preventive maintenance by trained tech Semi annually or annually per usage f...

Page 21: ...21 Reference AP 450 Physical Overview...

Page 22: ...duct label located on the bottom of the lift Duty Cycle 1 Min On 9 Mins Off Max Sound Level Raising Max load 50 dB Lowering Max load 50 dB on a lift in good working order Service Life 22 500 cycles or...

Page 23: ...eal time battery percentage also shown Battery Low Icon Battery is low Unit will still function Real time battery percentage also shown Overload Icon Lift is overloaded Real time battery percentage al...

Page 24: ...rol LCD Available Functionality Up Down E Down Lift fully charged Solid Green None None Lift ready for operation Solid Green None Charging hand control Solid Orange None None Low Battery Solid Orange...

Page 25: ...25 Lift Status Lift Status LED Color Audio Signals Hand Control LCD Available Functionality Up Down E Down Tilt Detected Solid Green None None None...

Page 26: ...on is available on our website For questions about the manufacture or operation of this equipment contact your local Handicare dealer or your local Handicare business development manager About Handica...