Operation
External Calibration
The weight required for calibration/adjustment is defined in the factory settings (see “Specifications”).
Step
Key (or instruction)
Display/Data output
1.
Zero the balance
[
TARE
]
.
g
2.
Start calibration
[CAL]
PLEASE WAIT
The balance will show
“PLEASE WAIT”,
after stable it will Display
the
prompted calibration weight
(In this sample, 200g)
+
.
gCAL
4.
Apply the prompted calibration
+
.
gCAL
weight
5.Calibrate Weight
[CAL]
+
.
g
The balance will show the weighing
value and blink. After stable, the
balance will beep a sound and return
to the normal weighing mode to finish
the calibration
6. Remove the calibration weight.
.
g
11