Differential input voltage (AC

rms

):

1,000 V

(DC + peak AC) max.:

±1,400 V

Max. input voltage per input:

±1,400 V

Attenuation ratio:

100:1

Switchable:

1,000:1

Bandwidth:

20 / 30 MHz

Rise time:

17 / 12 ns

Input impedance:

60 MΩ II 1.5 pF

Output impedance:

50 Ω

Max. output voltage:

±1.5 V at 1 MΩ

Accuracy after 1 min:

±3 % (18 °C

 to 

30 °C)

Common mode rejection  
DC / AC 1 MHz:

 
70 dB / › 50 dB

Inputs (CAT III):

2 safety 

 connectors

HZO40

HZO41

Bandwidth:

200 MHz

800 MHz

Attenuation ratio:

10:1

10:1

Rise time (10…90 %):

1.75 ns

437 ps

Gain accuracy:

±1 %

±2 %

Max. input voltage  
per input:

 
±60 V

 
±40 V

Max. differential input 
voltage (DC + peak AC):

 
±20 V

 
±15 V

Max. common mode  
input voltage:

 
±60 V

 
±30 V

Input impedance

Between inputs:

1 MΩ II 3.5 pF

200 kΩ II 1 pF

Each input to ground:

500 kΩ II 7 pF

100 kΩ II 2 pF

Output voltage  
(into 50 Ω):

±2 V

±1.5 V

Offset (typical):

±2 mV

±5 mV

CMRR (typical):

-80 dB at 60 Hz 
-50 dB at 10 MHz

-60 dB at 60 Hz
-15 dB at 500 MHz

Battery operation:

9 V battery 6LR61 9 V battery 6LR61

Battery life (typical):

7.5 h

4.5 h

HZO50

HZO51

HZ42   2 RU:

88 mm

Dimensions (W x D):

482 x 360 mm

*

(for Series 8100, HM8143, HM7042-5, HM8001-2, HMP2020, HMP2030  
and HMF series) case height 75 mm
HZ43   3 RU:

132.5 mm

Dimensions (W x D):

482 x 360 mm

*

(for HM6050-2, HMP4030, HMP4040) case height 125 mm
HZ46   4 RU:

177 mm

Dimensions (W x D):

482 x 170 mm

*

(for HMO series 3000 and HMS series) case height 175 mm
HZC95   2 RU:

88 mm

Dimensions (W x D):

482 x 282 mm

*

(for HMC8012) case height 88 mm
HZO91   4 RU:

177 mm

Dimensions (W x D):

482 x 110 mm

*

(for HMO72X, HMO102X, HMO152X, HMO202X) case height 175 mm
HZP91   4 RU:

177 mm

Dimensions (W x D):

482 x 360 mm

*

(for HMP4030, HMP4040) case height 125 mm

*plus excess of the instrument

 

Attenuation ratio: 10:1
Bandwidth: 

400 MHz

Rise time: 

<

1.0 ns

Input impedance: 10 MΩ II 12 pF
Max. voltage: 

400 V (DC + peak AC)

Probe factor identification:  
automatically after plugging

Attenuation ratio:  1,000:1
Bandwidth: 

400 MHz

Rise time: 

<

900 ps

Input impedance: 

50 MΩ II 7.5 pF

Max. voltage: 

1,000 V

rms

Probe factor identification:  
automatically after plugging

Attenuation ratio:  10:1
Bandwidth: 

1 GHz

Rise time: 

600 ps

Input impedance: 

1 MΩ II 0.9 pF

Max. input voltage:  20 V
Input dynamic range: ±8 V
Oscilloscope input coupling: 50 Ω

Measurement range:  ±20 A

rms

 /30 A

p

 

±100 A

rms

 / 1,000 A

rms

Accuracy: 

±1 % / ±2 mA 

±1 % / ±0.1 A / ±0.5 A

Bandwidth: 

DC to 100 kHz (0.5 dB) 

DC to 20 kHz

Resolution: 

±1 mA 

±100 mA / ±500 mA

Output voltage: 

100 mV / A 

10 mV / A  / 1mV / A

Load impedance: 

› 100 kΩ II 

 100 pF 

› 100 kΩ II 

 100 pF

Max. voltage: 

300 V

rms

 (AC or DC) 

300 V

rms

 (AC or DC)

Attenuation ratio: 10:1
Bandwidth: 

500 MHz

Rise time: 

<

700 ps

Input impedance:  10 MΩ II 9.5 pF
Max. voltage: 

400 V (DC + peak AC)

Probe factor identification:  
automatically after plugging

HAMEG  OSCILLOSCOPES  ARE 

www

.SCOPE-OF-THE-ART.

com

GREAT VALUE IN 

TEST & MEASUREMENT

HZO50 | HZO51  AC/DC Current Probes

HZ355  500 MHz Passive Probe

HZO20  High Voltage Probe

HZ350  400 MHz Passive Probe

HZO30  1 GHz Active Probe

HZO40 | HZO41  Active Differential Probes

HZ42 | HZ43 | HZ46 | HZC95 | HZO91 | HZP91  19“ Rackmount Kit

HZ115  Active Differential High Voltage Probe

HAMEG Instruments GmbH  | Industriestr. 6 | D-63533 Mainhausen | Germany 
Phone: +49 (0) 61 82 80 00 | info

@

hameg.com | www.hameg.com | blog.hameg.com

Summary of Contents for HMO3000 Series

Page 1: ...Product Overview...

Page 2: ...log channel 2GSa s 1GSa s Maximum sampling rate 4GSa s 2GSa s Memory depth per channel 4MPts 1MPts Maximum memory 8MPts 2MPts Max number of logic channels 16 8 Input impedance 1M 50 1M 50 1M V div 1M...

Page 3: ...less than 8s not available for HMS1000E This unit is used to measure the voltage standing wave ratio VSWR and reflection coefficient of a device under test with an impedance of 50 Frequency range 100k...

Page 4: ...al Multimeter Three active broadband probes designed for 50 inputs of HMS series powered either by the analyzer or by batteries E field probe H field probe High impedance probe The built in preamplifi...

Page 5: ...annels 4 3 3 2 HM8115 2 Power Consumption Meter 8 kW Measurement range 1mW to 8kW Voltages from 100mV to 500V currents from 1mA to 16A Frequency range DC to 1 kHz Simultaneous display of voltage curre...

Page 6: ...ss design Optional 4 wire test fixture incl shorting plate for evaluation of leaded components Optional 4 wire transformer test cable to measure the mutual inductance and ratio HM8118 LCR Bridge HZ181...

Page 7: ...trary waveforms External amplitude modulation up to 20kHz External connectors BURST GATE SWEEP TRIGGER Arbitrary waveform generator 40MSa s 12bit 4096Pts PC software free of charge to easily create us...

Page 8: ...MC8012 case height 88mm HZO91 4 RU 177mm Dimensions W x D 482 x 110mm for HMO72X HMO102X HMO152X HMO202X case height 175mm HZP91 4 RU 177mm Dimensions W x D 482 x 360mm for HMP4030 HMP4040 case height...

Reviews: